Precision of computer-aided volumetry of artificial small solid pulmonary nodules in ex vivo porcine lungs

被引:25
作者
Bolte, H.
Riede, C.
Mueller-Huelsbeck, S.
Freitag-Wolf, S.
Kohl, G.
Drews, T.
Heller, M.
Bieder, J.
机构
[1] Univ Hosp Schleswig Holstein, Dept Diagnost Radiol, D-24105 Kiel, Germany
[2] Univ Hosp Schleswig Holstein, Inst Med Informat & Stat, D-24105 Kiel, Germany
[3] Siemens Med Solut, D-91301 Forchheim, Germany
关键词
D O I
10.1259/bjr/23933268
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to investigate the precision of CT-based volumetric measurements of artificial small pulmonary nodules under ex vivo conditions. We implanted 322 artificial nodules in 23 inflated ex vivo porcine lungs in a dedicated chest phantom. The lungs were examined with a multislice spiral CT (20 mAs, collimation 16 x 0.75 mm, 1 mm slice thickness, 0.7 mm increment). A commercial volumetry software package (LungCARE VA70C-W; Siemens, Erlangen, Germany) was used for volume analysis in a semi-automatic and a manual corrected mode. After imaging, the lungs were dissected to harvest the nodules for gold standard determination. The volumes of 202 solitary, solid and well-defined lesions without contact with the pleura, greater bronchi or vessels were compared with the results of volumetry. A mean nodule diameter of 8.3 mm ( +/- 2.1 mm) was achieved. The mean relative deviation from the true lesion volume was -9.2% ( +/- 10.6%) for semi-automatic and -0.3% ( +/- 6.5%) for manual corrected volumetry. The subgroup of lesions from 5 mm to <10 mm in diameter showed a mean relative deviation of -8.7% ( +/- 10.9%) for semi-automatic volumetry and -0.3% (+/- 6.9%) for manually corrected volumetry. We conclude that the presented software allowed for precise volumetry of artificial, nodules in ex vivo lung tissue. This result is comparable to the findings of previous in vitro studies.
引用
收藏
页码:414 / 421
页数:8
相关论文
共 35 条
[1]   Pulmonary nodules at chest CT: Effect of computer-aided diagnosis on radiologists' detection performance [J].
Awai, K ;
Murao, K ;
Ozawa, A ;
Komi, M ;
Hayakawa, H ;
Hori, S ;
Nishimura, Y .
RADIOLOGY, 2004, 230 (02) :347-352
[2]   Diagnostic performance of a flat-panel detector at low tube voltage in chest radiography -: A phantom study [J].
Bernhardt, TM ;
Rapp-Bernhardt, U ;
Lenzen, H ;
Röhl, FW ;
Diederich, S ;
Papke, K ;
Ludwig, K ;
Heindel, W .
INVESTIGATIVE RADIOLOGY, 2004, 39 (02) :97-103
[3]   Artificial thorax for MR imaging studies in porcine heart-lung preparations [J].
Biederer, J ;
Heller, M .
RADIOLOGY, 2003, 226 (01) :250-255
[4]   Reproducibility of computer-aided volumetry of artificial small pulmonary nodules in ex vivo porcine lungs [J].
Bolte, H ;
Riedel, C ;
Jahnke, T ;
Inan, N ;
Freitag, S ;
Kohl, G ;
Heller, M ;
Biederer, J .
INVESTIGATIVE RADIOLOGY, 2006, 41 (01) :28-35
[5]   Ex-vivo injection for implanting solid pulmonary nodules into porcine lungs for multi-slice CT [J].
Bolte, H ;
Müller-Hülsbeck, S ;
Riedel, C ;
Jahnke, T ;
Inan, N ;
Heller, M ;
Biederer, J .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (10) :1380-1384
[6]   Screening for early lung cancer with low-dose spiral CT: Prevalence in 817 asymptomatic smokers [J].
Diederich, S ;
Wormanns, D ;
Semik, M ;
Thomas, M ;
Lenzen, H ;
Roos, N ;
Heindel, W .
RADIOLOGY, 2002, 222 (03) :773-781
[7]   Detection of pulmonary nodules by multislice computed tomography: improved detection rate with reduced slice thickness [J].
Fischbach, F ;
Knollmann, F ;
Griesshaber, V ;
Freund, T ;
Akkol, E ;
Felix, R .
EUROPEAN RADIOLOGY, 2003, 13 (10) :2378-2383
[8]   Volumetric measurement of synthetic lung nodules with multi-detector row CT: Effect of various image reconstruction parameters and segmentation thresholds on measurement accuracy [J].
Goo, JM ;
Tongdee, T ;
Tongdee, R ;
Yeo, K ;
Hildebolt, CF ;
Bae, KT .
RADIOLOGY, 2005, 235 (03) :850-856
[9]   Neural networks for the analysis of small pulmonary nodules [J].
Henschke, CI ;
Yankelevitz, DF ;
Mateescu, I ;
Brettle, DW ;
Rainey, TG ;
Weingard, FS .
CLINICAL IMAGING, 1997, 21 (06) :390-399
[10]   Early Lung Cancer Action Project: overall design and findings from baseline screening [J].
Henschke, CI ;
McCauley, DI ;
Yankelevitz, DF ;
Naidich, DP ;
McGuinness, G ;
Miettinen, OS ;
Libby, DM ;
Pasmantier, MW ;
Koizumi, J ;
Altorki, NK ;
Smith, JP .
LANCET, 1999, 354 (9173) :99-105