Prostate intrafraction motion evaluation using kV fluoroscopy during treatment delivery: A feasibility and accuracy study

被引:43
作者
Adamson, Justus [1 ,2 ]
Wu, Qiuwen [1 ]
机构
[1] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI 48073 USA
[2] Wayne State Univ, Dept Radiat Oncol, Detroit, MI 48201 USA
关键词
intrafraction motion; prostate cancer; kV fluoroscopy; IMRT; IGRT;
D O I
10.1118/1.2899998
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Margin reduction for prostate radiotherapy is limited by uncertainty in prostate localization during treatment. We investigated the feasibility and accuracy of measuring prostate intrafraction motion using kV fluoroscopy performed simultaneously with radiotherapy. Three gold coils used for target localization were implanted into the patient's prostate gland before undergoing hypofractionated online image-guided step-and-shoot intensity modulated radiation therapy (IMRT) on an Elekta Synergy linear accelerator. At each fraction, the patient was aligned using a cone-beam computed tomography (CBCT), after which the IMRT treatment delivery and fluoroscopy were performed simultaneously. In addition, a post-treatment CBCT was acquired with the patient still on the table. To measure the intrafraction motion, we developed an algorithm to register the fluoroscopy images to a reference image derived from the post-treatment CBCT, and we estimated coil motion in three-dimensional (3D) space by combining information from registrations at different gantry angles. We also detected the MV beam turning on and off using MV scatter incident in the same fluoroscopy images, and used this information to synchronize our intrafraction evaluation with the treatment delivery. In addition, we assessed the following: the method to synchronize with treatment delivery, the dose from kV imaging, the accuracy of the localization, and the error propagated into the 3D localization from motion between fluoroscopy acquisitions. With 0.16 mAs/frame and a bowtie filter implemented, the coils could be localized with the gantry at both 0 degrees and 270 degrees with the MV beam off, and at 270 degrees with the MV beam on when multiple fluoroscopy frames were averaged. The localization in two-dimensions for phantom and patient measurements was performed with submillimeter accuracy. After backprojection into 3D the patient localization error was (-0.04 +/- 0.30) mm, (0.09 +/- 0.36) mm, and (0.03 +/- 0.68) mm in the right-left (RL), anterior-posterior (AP), and superior-inferior (SI) axes, respectively. Simulations showed that while oscillating (stationary) motion cannot be effectively represented in 3D, linearly drifting (nonstationary) motion is detectable with good accuracy. These results show that measuring prostate intrafraction motion using a single kV imager during radiotherapy is feasible and can be performed with acceptable accuracy. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:1793 / 1806
页数:14
相关论文
共 37 条
[1]
Robustness and precision of an automatic marker detection algorithm for online prostate daily targeting using a standard V-EPID [J].
Aubin, S ;
Beaulieu, L ;
Pouliot, S ;
Pouliot, J ;
Roy, R ;
Girouard, LM ;
Martel-Brisson, N ;
Vigneault, E ;
Laverdière, J .
MEDICAL PHYSICS, 2003, 30 (07) :1825-1832
[2]
Measurements of intrafraction motion and interfraction and intrafraction rotation of prostate by three-dimensional analysis of daily portal imaging with radiopaque markers [J].
Aubry, JF ;
Beaulieu, L ;
Girouard, LM ;
Aubin, S ;
Tremblay, D ;
Laverdière, J ;
Vigneault, E .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01) :30-39
[3]
Accuracy of a wireless localization system for radiotherapy [J].
Balter, JM ;
Wright, JN ;
Newell, LJ ;
Friemel, B ;
Dimmer, S ;
Cheng, Y ;
Wong, J ;
Vertatschitsch, E ;
Mate, TP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (03) :933-937
[4]
Automated localization of the prostate at the time of treatment using implanted radiopaque markers: Technical feasibility [J].
Balter, JM ;
Lam, KL ;
Sandler, HM ;
Littles, JF ;
Bree, RL ;
TenHaken, RK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1281-1286
[5]
MEASUREMENT OF PROSTATE MOVEMENT OVER THE COURSE OF ROUTINE RADIOTHERAPY USING IMPLANTED MARKERS [J].
BALTER, JM ;
SANDLER, HM ;
LAM, K ;
BREE, RL ;
LICHTER, AS ;
TENHAKEN, RK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01) :113-118
[6]
Integrated radiotherapy imaging system (IRIS): design considerations of tumour tracking with linac gantry-mounted diagnostic x-ray systems with flat-panel detectors [J].
Berbeco, RI ;
Jiang, SB ;
Sharp, GC ;
Chen, GTY ;
Mostafavi, H ;
Shirato, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (02) :243-255
[7]
Britton Keith R, 2005, Radiat Med, V23, P14
[8]
Experience of ultrasound-based daily prostate localization [J].
Chandra, A ;
Dong, L ;
Huang, E ;
Kuban, DA ;
O'Neill, L ;
Rosen, I ;
Pollack, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (02) :436-447
[9]
On-line aSi portal imaging of implanted fiducial markers for the reduction of interfraction error during conformal radiotherapy of prostate carcinoma [J].
Chung, PWM ;
Haycocks, T ;
Brown, T ;
Cambridge, Z ;
Kelly, V ;
Alasti, H ;
Jaffray, DA ;
Catton, CN .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01) :329-334
[10]
Characterization of a fluoroscopic imaging system for kV and MV radiography [J].
Drake, DG ;
Jaffray, DA ;
Wong, JW .
MEDICAL PHYSICS, 2000, 27 (05) :898-905