Dose Reduction to Anterior Surfaces With Organ-Based Tube-Current Modulation: Evaluation of Performance in a Phantom Study

被引:85
作者
Duan, Xinhui [1 ]
Wang, Jia [1 ]
Christner, Jodie A. [1 ]
Leng, Shuai [1 ]
Grant, Katharine L. [2 ]
McCollough, Cynthia H. [1 ]
机构
[1] Mayo Clin, Dept Radiol, CT Clin Innovat Ctr, Rochester, MN 55905 USA
[2] Siemens Med Solut, Malvern, PA USA
关键词
CT; dose reduction; tube-current modulation; BISMUTH BREAST SHIELDS; RADIOSENSITIVE ORGANS; COMPUTED-TOMOGRAPHY; CHEST CT; EXPOSURE; TISSUES;
D O I
10.2214/AJR.10.6061
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
OBJECTIVE. The purpose of this study was to evaluate in phantoms the dose reduction to anterior surfaces and image quality with organ-based tube-current modulation in head and thoracic CT. MATERIALS AND METHODS. Organ-based tube-current modulation is designed to reduce radiation dose to superficial radiosensitive organs, such as the lens of the eye, thyroid, and breast, by decreasing the tube current when the tube passes closest to these organs. Dose and image quality were evaluated in phantoms for clinical head and thorax examination protocols with and without organ-based tube-current modulation. Surface dose reduction as a function of position was measured using a 32-cm CT dose index (CTDI) phantom, an anthropomorphic adult phantom, and ion chambers. Surface dose reduction as a function of patient size was investigated using three semianthropomorphic phantoms with posteroanterior dimensions of 14, 25, and 31 cm. Image noise (the SD of CT numbers in regions of interest) was evaluated for the anthropomorphic and the semianthropomorphic phantoms. RESULTS. For equivalent scanner output (volume CTDI), the dose to the midline of the anterior surface was reduced by 27-50%, depending on the anatomic region (head or thorax) and phantom size, and the dose to the posterior surface was correspondingly increased. Image noise was not significantly different between scans with and without organ-based tubecurrent modulation (p = 0.85). CONCLUSION. Organ-based tube-current modulation can reduce the dose to the anterior surface of patients without increasing image noise by commensurately increasing the dose to the posterior surface. This technique can reduce the dose to anterior radiosensitive organs for head and thoracic CT scans.
引用
收藏
页码:689 / 695
页数:7
相关论文
共 25 条
[1]
Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation [J].
Angel, Erin ;
Yaghmai, Nazanin ;
Jude, Cecilia Matilda ;
DeMarco, John J. ;
Cagnon, Christopher H. ;
Goldin, Jonathan G. ;
McCollough, Cynthia H. ;
Primak, Andrew N. ;
Cody, Dianna D. ;
Stevens, Donna M. ;
McNitt-Gray, Michael F. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (05) :1340-1345
[2]
Monte Carlo simulations to assess the effects of tube current modulation on breast dose for multidetector CT [J].
Angel, Erin ;
Yaghmai, Nazanin ;
Jude, Cecilia Matilda ;
DeMarco, John J. ;
Cagnon, Christopher H. ;
Goldin, Jonathan G. ;
Primak, Andrew N. ;
Stevens, Donna M. ;
Cody, Dianna D. ;
McCollough, Cynthia H. ;
McNitt-Gray, Michael F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (03) :497-511
[3]
Risk of cancer from diagnostic X-rays:: estimates for the UK and 14 other countries [J].
Berrington de González, A ;
Darby, S .
LANCET, 2004, 363 (9406) :345-351
[4]
Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[5]
In-plane bismuth breast shields for pediatric CT: Effects on radiation dose and image quality using experimental and clinical data [J].
Fricke, BL ;
Donnelly, LF ;
Frush, DP ;
Yoshizumi, T ;
Varchena, V ;
Poe, SA ;
Lucaya, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 180 (02) :407-411
[6]
Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality [J].
Geleijns, J. ;
Salvado Artells, M. ;
Veldkamp, W. J. H. ;
Lopez Tortosa, M. ;
Calzado Cantera, A. .
EUROPEAN RADIOLOGY, 2006, 16 (10) :2334-2340
[7]
Organ doses to adult patients for chest CT [J].
Huda, Walter ;
Sterzik, Alexander ;
Tipnis, Sameer ;
Schoepf, U. Joseph .
MEDICAL PHYSICS, 2010, 37 (02) :842-847
[8]
Radiation Dose Savings for Adult Pulmonary Embolus 64-MDCT Using Bismuth Breast Shields, Lower Peak Kilovoltage, and Automatic Tube Current Modulation [J].
Hurwitz, Lynne M. ;
Yoshizumi, Terry T. ;
Goodman, Philip C. ;
Nelson, Rendon C. ;
Toncheva, Greta ;
Nguyen, Giao B. ;
Lowry, Carolyn ;
Anderson-Evans, Colin .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (01) :244-253
[9]
*IMV, 2008, IMV 2007 CT MARK SUM
[10]
JOSHI G, 2010, CANC RISKS ASS CHEST