Implantable drug delivery device using frequency-controlled wireless hydrogel microvalves

被引:56
作者
Rahimi, Somayyeh [1 ]
Sarraf, Elie H. [1 ]
Wong, Gregory K. [1 ]
Takahata, Kenichi [1 ]
机构
[1] Univ British Columbia, Dept Elect & Comp Engn, Vancouver, BC V6T 1Z4, Canada
基金
加拿大创新基金会; 加拿大自然科学与工程研究理事会;
关键词
Drug delivery; Controlled release; Wireless; Microvalves; Hydrogel; Polyimide; POWER TRANSFER; SYSTEM; VALVES; MEMS; FABRICATION; RELEASE;
D O I
10.1007/s10544-010-9491-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper reports a micromachined drug delivery device that is wirelessly operated using radiofrequency magnetic fields for implant applications. The controlled release from the drug reservoir of the device is achieved with the microvalves of poly(N-isopropylacrylamide) thermoresponsive hydrogel that are actuated with a wireless resonant heater, which is activated only when the field frequency is tuned to the resonant frequency of the heater circuit. The device is constructed by bonding a 1-mm-thick polyimide component with the reservoir cavity to the heater circuit that uses a planar coil with the size of 5-10 mm fabricated on polyimide film, making all the outer surfaces to be polyimide. The release holes created in a reservoir wall are opened/closed by the hydrogel microvalves that are formed inside the reservoir by in-situ photolithography that uses the reservoir wall as a photomask, providing the hydrogel structures self-aligned to the release holes. The wireless heaters exhibit fast and strong response to the field frequency, with a temperature increase of up to 20A degrees C for the heater that has the 34-MHz resonant frequency, achieving 38-% shrinkage of swelled hydrogel when the heater is excited at its resonance. An active frequency range of similar to 2 MHz is observed for the hydrogel actuation. Detailed characteristics in the fabrication and actuation of the hydrogel microvalves as well as experimental demonstrations of frequency-controlled temporal release are reported.
引用
收藏
页码:267 / 277
页数:11
相关论文
共 41 条
[1]   In vitro experiment of the pressure regulating valve for a glaucoma implant [J].
Bae, BH ;
Kee, HS ;
Kim, SH ;
Lee, Y ;
Sim, TS ;
Kim, YW ;
Park, KH .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2003, 13 (05) :613-619
[2]   A large stepwise motion electrostatic actuator for a wireless microrobot [J].
Basset, P ;
Kaiser, A ;
Bigotte, P ;
Collard, D ;
Buchaillot, L .
FIFTEENTH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 2002, :606-609
[3]   Development and Characterization of Microfluidic Devices and Systems for Magnetic Bead-Based Biochemical Detection [J].
Choi, Jin-Woo ;
Oh, Kwang W. ;
Han, Arum ;
Okulan, Nihat ;
Wijayawardhana, C. Ajith ;
Lannes, Chad ;
Bhansali, Shekhar ;
Schlueter, Kevin T. ;
Heineman, William R. ;
Halsall, H. Brain ;
Nevin, Joseph H. ;
Helmicki, Arthur J. ;
Henderson, H. Thurman ;
Ahn, Chong H. .
BIOMEDICAL MICRODEVICES, 2001, 3 (03) :191-200
[4]   A valved responsive hydrogel microdispensing device with integrated pressure source [J].
Eddington, DT ;
Beebe, DJ .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2004, 13 (04) :586-593
[5]   A low power, microvalve regulated architecture for drug delivery systems [J].
Evans, Allan Thomas ;
Park, Jong M. ;
Chiravuri, Srinivas ;
Gianchandani, Yogesh B. .
BIOMEDICAL MICRODEVICES, 2010, 12 (01) :159-168
[6]   Magnetically driven micro ball valves fabricated by multilayer adhesive film bonding [J].
Fu, C ;
Rummler, Z ;
Schomburg, W .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2003, 13 (04) :S96-S102
[7]   Oscillating magnetic field-actuated microvalves for micro- and nanofluidics [J].
Ghosh, Santaneel ;
Yang, Chao ;
Cai, Tong ;
Hu, Zhibing ;
Neogi, Arup .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2009, 42 (13)
[8]  
Glasgow IK, 1999, SENSOR MATER, V11, P269
[9]   Electronic MEMS for triggered delivery [J].
Grayson, ACR ;
Shawgo, RS ;
Li, YW ;
Cima, MJ .
ADVANCED DRUG DELIVERY REVIEWS, 2004, 56 (02) :173-184
[10]   VOLUME-PHASE TRANSITIONS OF IONIZED N-ISOPROPYLACRYLAMIDE GELS [J].
HIROTSU, S ;
HIROKAWA, Y ;
TANAKA, T .
JOURNAL OF CHEMICAL PHYSICS, 1987, 87 (02) :1392-1395