A novel counterpulsation mode of rotary left ventricular assist devices can enhance myocardial perfusion

被引:43
作者
Ando, Masahiko [1 ,2 ]
Takewa, Yoshiaki [1 ]
Nishimura, Takashi [2 ]
Yamazaki, Kenji [3 ]
Kyo, Shunei [2 ]
Ono, Minoru [4 ]
Tsukiya, Tomonori [1 ]
Mizuno, Toshihide [1 ]
Taenaka, Yoshiyuki [1 ]
Tatsumi, Eisuke [1 ]
机构
[1] Natl Cerebral & Cardiovasc Ctr Res Inst, Dept Artificial Organ, Osaka 5658565, Japan
[2] Univ Tokyo, Dept Therapeut Strategy Heart Failure, Bunkyo Ku, Tokyo 1138655, Japan
[3] Tokyo Womens Med Univ, Dept Cardiovasc Surg, Shinjuku Ku, Tokyo 1628666, Japan
[4] Univ Tokyo, Dept Cardiothorac Surg, Bunkyo Ku, Tokyo 1138655, Japan
关键词
Left ventricular assist device; Myocardial perfusion; Counterpulsation; INTRAAORTIC BALLOON COUNTERPULSATION; BLOOD-FLOW; AXIAL-FLOW; PHYSIOLOGY; SUPPORT; HEART; MICROCIRCULATION; HEMODYNAMICS; BYPASS;
D O I
10.1007/s10047-011-0573-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The effect of rotary left ventricular assist devices (LVADs) on myocardial perfusion has yet to be clearly elucidated, and several studies have shown decreased coronary flow under rotary LVAD support. We have developed a novel pump controller that can change its rotational speed (RS) in synchronization with the native cardiac cycle. The aim of our study was to evaluate the effect of counterpulse mode, which increases the RS in diastole, during coronary perfusion. Experiments were performed on ten adult goats. The EVAHEART LVAD was installed by the left ventricular uptake and the descending aortic return. Ascending aortic flow, pump flow, and coronary flow of the left main trunk were monitored. Coronary flow was compared under four conditions: circuit-clamp, continuous mode (constant pump speed), counterpulse mode (increased pump speed in diastole), and copulse mode (increased pump speed in systole). There were no significant baseline changes between these groups. In counterpulse mode, coronary flow increased significantly compared with that in continuous mode. The waveform analysis clearly revealed that counterpulse mode mainly resulted in increased diastolic coronary flow. In conclusion, counterpulse mode of rotary LVADs can enhance myocardial perfusion. This novel drive mode can provide great benefits to the patients with end-stage heart failure, especially those with ischemic etiology.
引用
收藏
页码:185 / 191
页数:7
相关论文
共 17 条
[1]   Subendocardial ischemia without coronary artery disease: is elevated left ventricular end diastolic pressure the culprit? [J].
Elhabyan, AK ;
Reyes, BJ ;
Hallak, O ;
Broce, M ;
Rosencrance, JG ;
Lucas, BD ;
Fazal, H .
CURRENT MEDICAL RESEARCH AND OPINION, 2004, 20 (05) :773-777
[2]   CORONARY PHYSIOLOGY [J].
FEIGL, EO .
PHYSIOLOGICAL REVIEWS, 1983, 63 (01) :1-205
[3]   Vertical displacement of the beating heart by the octopus tissue stabilizer:: Influence on coronary flow [J].
Gründeman, PF ;
Borst, C ;
van Herwaarden, JA ;
Verlaan, CWJ ;
Jansen, EWL .
ANNALS OF THORACIC SURGERY, 1998, 65 (05) :1348-1352
[4]   Coronary microcirculation during left heart bypass with a centrifugal pump [J].
Hata, M ;
Shiono, M ;
Orime, Y ;
Nakata, K ;
Sezai, A ;
Yamada, H ;
Saito, T ;
Sezai, Y .
ARTIFICIAL ORGANS, 1996, 20 (06) :678-680
[5]  
NAKAMURA T, 1988, ARTIF ORGANS, V12, P56
[6]   Effect of pulsatile and nonpulsatile assist on heart and kidney microcirculation with cardiogenic shock [J].
Nakata, K ;
Shiono, M ;
Orime, Y ;
Hata, M ;
Sezai, A ;
Saitoh, T ;
Sezai, Y .
ARTIFICIAL ORGANS, 1996, 20 (06) :681-684
[7]  
Noda H, 1989, ASAIO Trans, V35, P445, DOI 10.1097/00002480-198907000-00088
[8]   Phasic coronary blood flow pattern during a continuous flow left ventricular assist support [J].
Ootaki, Y ;
Kamohara, K ;
Akiyama, M ;
Zahr, F ;
Kopcak, MW ;
Dessoffy, R ;
Fukamachi, K .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2005, 28 (05) :711-716
[9]   A novel external counterpulsation system for coronary artery disease and heart failure: pilot studies and initial clinical experiences [J].
Shimizu, Tsuyoshi ;
Kyo, Shunei ;
Imanaka, Kazuhito ;
Nakaoka, Kohei ;
Nishimura, Etsuji ;
Okumura, Takahiro ;
Ishii, Masaaki ;
Hisagi, Motoyuki ;
Nishimura, Takashi ;
Motomura, Noboru ;
Ono, Minoru ;
Takamoto, Shinichi .
JOURNAL OF ARTIFICIAL ORGANS, 2010, 13 (03) :161-169
[10]   IMPROVED REGIONAL MYOCARDIAL BLOOD-FLOW, LEFT-VENTRICULAR UNLOADING, AND INFARCT SALVAGE USING AN AXIAL-FLOW, TRANSVALVULAR LEFT-VENTRICULAR ASSIST DEVICE - A COMPARISON WITH INTRAAORTIC BALLOON COUNTERPULSATION AND REPERFUSION ALONE IN A CANINE INFARCTION MODEL [J].
SMALLING, RW ;
CASSIDY, DB ;
BARRETT, R ;
LACHTERMAN, B ;
FELLI, P ;
AMIRIAN, J .
CIRCULATION, 1992, 85 (03) :1152-1159