Effect of electrostatic charge, flow, delay and multiple actuations on the in vitro delivery of salbutamol from different small volume spacers for infants

被引:110
作者
Wildhaber, JH [1 ]
Devadason, SG [1 ]
Eber, E [1 ]
Hayden, MJ [1 ]
Everard, ML [1 ]
Summers, QA [1 ]
LeSouef, PN [1 ]
机构
[1] ROYAL PERTH HOSP,DEPT RESP MED,PERTH,WA,AUSTRALIA
关键词
inhalation devices; aerosol therapy; electrostatic charge;
D O I
10.1136/thx.51.10.985
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background - A study was undertaken to determine the influences of electrostatic charge, flow, delay, and multiple actuations on the in vitro delivery of salbutamol generated by a pressurised metered dose inhaler (pMDI) from small volume spacers used in infants. Methods - Ten actuations from a salbutamol pMDI were drawn at different flow rates after either single or multiple actuations, with or without delay, through either static or reduced static spacers. An ionic detergent was used to reduce the charge of plastic spacers (Babyhaler, Babyspacer, Aerochamber, Nebuhaler). Electrostatic charge was measured using an electrometer. A multistage liquid impinger was used to determine the particle size distribution of the output of the pMDI through the spacers. Results - Electrostatic charge on the surface of plastic spacers had the greatest influence on delivery, causing a decrease in drug delivery. Reducing charge by coating the surface with ionic detergent resulted in an increase of 46.5-71.1% (p<0.001) in small (<6.8 mu m) particle delivery from small volume plastic spacers. Lower flow, delay, and multiple actuations resulted in decreased delivery from static spacers. Lower flow resulted in a decrease of 15% in small (<9.6 mu m) particle delivery. Delay and multiple actuations resulted in a decrease of 40.7% and 76.0%, respectively, in small (<6.8 mu m) particle delivery. The influences of lower flow, delay, and multiple actuations were greatly reduced or even eliminated by reducing charge. However, multiple actuations still resulted in a significant decreased delivery (p<0.05). The reduced static Nebuhaler had a higher delivery than all small volume spacers. Conclusions - Electrostatic charge has a major influence on the delivery of salbutamol from small volume spacers. Using a metal spacer or ionic detergent coating of plastic spacers resulted in no or reduced charge and hence in improved delivery. Lower flow, delay, and multiple actuations played a major part only in static spacers.
引用
收藏
页码:985 / 988
页数:4
相关论文
共 13 条
[1]   THE EFFECT OF DELAY, MULTIPLE ACTUATIONS AND SPACER STATIC CHARGE ON THE IN-VITRO DELIVERY OF BUDESONIDE FROM THE NEBUHALER [J].
BARRY, PW ;
OCALLAGHAN, C .
BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1995, 40 (01) :76-78
[2]   MULTIPLE ACTUATIONS OF SALBUTAMOL MDI INTO A SPACER DEVICE REDUCE THE AMOUNT OF DRUG RECOVERED IN THE RESPIRABLE RANGE [J].
BARRY, PW ;
OCALLAGHAN, C .
EUROPEAN RESPIRATORY JOURNAL, 1994, 7 (09) :1707-1709
[3]  
BISGAARD H, 1995, EUR RESPIR J, V8, P856
[4]   A NONELECTROSTATIC SPACER FOR AEROSOL DELIVERY [J].
BISGAARD, H ;
ANHOJ, J ;
KLUG, B ;
BERG, E .
ARCHIVES OF DISEASE IN CHILDHOOD, 1995, 73 (03) :226-230
[5]   DRUG DELIVERY FROM HOLDING CHAMBERS WITH ATTACHED FACEMASK [J].
EVERARD, ML ;
CLARK, AR ;
MILNER, AD .
ARCHIVES OF DISEASE IN CHILDHOOD, 1992, 67 (05) :580-585
[6]  
KIM CS, 1987, AM REV RESPIR DIS, V135, P157
[7]   SPACER DEVICES USED WITH METERED-DOSE INHALERS - BREAKTHROUGH OR GIMMICK [J].
KONIG, P .
CHEST, 1985, 88 (02) :276-284
[8]   SHORT-TERM EFFECT OF ALBUTEROL, DELIVERED VIA A NEW AUXILIARY DEVICE, IN WHEEZY INFANTS [J].
KRAEMER, R ;
FREY, U ;
SOMMER, CW ;
RUSSI, E .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1991, 144 (02) :347-351
[9]   A NEW BABY-SPACER DEVICE FOR AEROSOLIZED BRONCHODILATOR ADMINISTRATION IN INFANTS WITH BRONCHOPULMONARY DISEASE [J].
KRAEMER, R ;
BIRRER, P ;
MODELSKA, K ;
AEBISCHER, CC ;
SCHONI, MH .
EUROPEAN JOURNAL OF PEDIATRICS, 1992, 151 (01) :57-60
[10]   SPACING DEVICES AND METERED-DOSE INHALERS IN CHILDHOOD ASTHMA [J].
LEVISON, H ;
REILLY, PA ;
WORSLEY, GH .
JOURNAL OF PEDIATRICS, 1985, 107 (05) :662-668