Use of electrolytes containing multiple co-anions in the analysis of anions by capillary electrophoresis using indirect absorbance detection

被引:30
作者
Doble, P [1 ]
Haddad, PR [1 ]
机构
[1] Univ Tasmania, Dept Chem, Hobart, Tas 7001, Australia
关键词
D O I
10.1021/ac9711955
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Background electrolytes (BGEs) containing more than one W-absorbing probe co-anion were investigated as possible means to control peak symmetries and improve the sensitivity of indirect detection in the separation of a mixture of inorganic and organic anions having a range of electrophoretic mobilities. In initial experiments, chloride and propanoate, which do not absorb at the detection wavelength, were added individually to a BGE containing phthalate as the UV-absorbing probe co-anion, The response ratios (i.e., the detector response for an analyte obtained with the BGE containing the probe and added co-anion divided by the response of the BGE containing. the probe alone) were found to be dependent on the relative mobilities of the analyte, probe, and co-anion, In general, it was found that the analyte mainly displaced the BGE component to which its mobility was closest and exclusively displaced any BGE component having the same mobility, This behavior was utilized to design BGEs containing multiple probes to improve peak shapes by matching the mobilities of the BGE components with those of the analytes. A BGE comprising chromate and phthalate as probes was used to demonstrate the improvement in peak shapes when such an approach was used. This was further extended to a BGE containing three probes, namely, chromate, phthalate, and benzoate, System peaks were observed for each multiple-component BGE and for n BGE co-anions; n - 1 system peaks were induced. A simple linear function relating the mobility of the system peak for a two-co-anion BGE to the mobilities and relative concentrations of each of the co-anions was derived empirically. Finally, a series of probes was investigated to determine the optimum multiple BGE composition giving the best peak shapes and sensitivity in the separation of a mixture of 15 analytes. The best combination was a two-probe BGE consisting of chromate and 4-hydroxybenzenesulfonic acid.
引用
收藏
页码:15 / 22
页数:8
相关论文
共 17 条
[1]   Calculation of the composition of sample zones in capillary zone electrophoresis .3. Weak bases and the application of background electrolytes with two co-ions [J].
Beckers, JL .
JOURNAL OF CHROMATOGRAPHY A, 1996, 741 (02) :265-277
[2]   THEORETICAL AND EXPERIMENTAL ASPECTS OF INDIRECT DETECTION IN CAPILLARY ELECTROPHORESIS [J].
BRUIN, GJM ;
VANASTEN, AC ;
XU, XM ;
POPPE, H .
JOURNAL OF CHROMATOGRAPHY, 1992, 608 (1-2) :97-107
[3]   EFFECT OF MULTIPLE ELECTROLYTE BUFFERS ON PEAK SYMMETRY, RESOLUTION, AND SENSITIVITY IN CAPILLARY ELECTROPHORESIS [J].
BULLOCK, J ;
STRASTERS, J ;
SNIDER, J .
ANALYTICAL CHEMISTRY, 1995, 67 (18) :3246-3252
[4]   EVALUATION OF CARRIER ELECTROLYTES FOR CAPILLARY ZONE ELECTROPHORESIS OF LOW-MOLECULAR-MASS ANIONS WITH INDIRECT UV DETECTION [J].
COUSINS, SM ;
HADDAD, PR ;
BUCHBERGER, W .
JOURNAL OF CHROMATOGRAPHY A, 1994, 671 (1-2) :397-402
[5]   INDIRECT PHOTOMETRIC DETECTION IN CAPILLARY ZONE ELECTROPHORESIS [J].
FORET, F ;
FANALI, S ;
OSSICINI, L ;
BOCEK, P .
JOURNAL OF CHROMATOGRAPHY, 1989, 470 (01) :299-308
[6]  
Kuhn R., 1993, capillary electrophoresis: principles and practice, DOI DOI 10.1007/978-3-642-78058-5
[7]   DETERMINATION OF ANIONS IN WATER SAMPLES BY CAPILLARY ZONE ELECTROPHORESIS WITH INDIRECT UV DETECTION [J].
LI, K ;
LI, SFY .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1994, 17 (18) :3889-3910
[8]   OPTIMIZATION OF INDIRECT PHOTOMETRIC DETECTION OF ANIONS IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS [J].
MA, YF ;
ZHANG, RL .
JOURNAL OF CHROMATOGRAPHY, 1992, 625 (02) :341-348
[9]   System peaks in capillary zone electrophoresis .3. Practical rules for predicting the existence of system peaks in capillary zone electrophoresis of anions using indirect spectrophotometric detection [J].
Macka, M ;
Haddad, PR ;
Gebauer, P ;
Bocek, P .
ELECTROPHORESIS, 1997, 18 (11) :1998-2007
[10]   QUANTITATIVE ASPECTS OF INDIRECT UV DETECTION IN CAPILLARY ZONE ELECTROPHORESIS [J].
NIELEN, MWF .
JOURNAL OF CHROMATOGRAPHY, 1991, 588 (1-2) :321-326