Bead injection ELISA for the determination of antibodies implicated in type 1 diabetes mellitus

被引:23
作者
Carroll, AD
Scampavia, L
Luo, D
Lernmark, Å
Ruzicka, J
机构
[1] Univ Washington, Dept Chem, Seattle, WA 98195 USA
[2] Univ Washington, Dept Med, Seattle, WA 98195 USA
关键词
D O I
10.1039/b302786d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work introduces a novel analytical method for the detection and study of GAD65 autoantibodies, which have been implicated in the onset of type 1 diabetes. There is a clinical need for a rapid and automated assay for GAD65 autoantibodies. Therefore, this method was designed to exploit the advantages of bead injection (BI) analysis for enzyme-linked immunosorbent assays (ELISA). BI ELISA is a microscale technique that uses enzyme labeled secondary antibodies to detect the capture of target antibodies on immobilized antigen in the flow cell of the lab-on-valve (LOV) manifold. A detection limit of 20 ng mL(-1) of GAD65 monoclonal antibody 144 compares favorably with the sensitivity and precision of a standard ELISA currently employed to detect GAD65 autoantibodies. Compared to the standard ELISA protocol, BI ELISA offers a significantly reduced assay time and complete automation of solution handling and detection.
引用
收藏
页码:1157 / 1162
页数:6
相关论文
共 25 条
[11]  
Kemeny D., 1988, ELISA OTHER SOLID PH, P1
[12]   Glutamic acid decarboxylase Gene to antigen to disease [J].
Lernmark, A .
JOURNAL OF INTERNAL MEDICINE, 1996, 240 (05) :259-277
[13]  
LUO D, 2003, IN PRESS J BIOTECHNO
[14]   Antibodies against glutamic acid decarboxylase: prevalence in neurological diseases [J].
Meinck, HM ;
Faber, L ;
Morgenthaler, N ;
Seissler, J ;
Maile, S ;
Butler, M ;
Solimena, M ;
DeCamilli, P ;
Scherbaum, WA .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 2001, 71 (01) :100-103
[15]   Highly-sensitive and specific enzyme-linked immunosorbent assays for GAD65 autoantibodies using a thioredoxin-GAD65 fusion antigen [J].
Papouchado, ML ;
Valdez, SN ;
Ermacora, MR ;
Ganan, S ;
Poskus, E .
JOURNAL OF IMMUNOLOGICAL METHODS, 1997, 207 (02) :169-178
[16]   Lab-on-valve: universal microflow analyzer based on sequential and bead injection [J].
Ruzicka, J .
ANALYST, 2000, 125 (06) :1053-1060
[17]   From flow injection to bead injection [J].
Ruzicka, J ;
Scampavia, L .
ANALYTICAL CHEMISTRY, 1999, 71 (07) :257A-263A
[18]  
RUZICKA J, 1988, FLOW INJECTION ANAL, P79
[19]   Automation of functional assays by flow injection fluorescence microscopy [J].
Scampavia, LD ;
Hodder, PS ;
Lähdesmäki, I ;
Ruzicka, J .
TRENDS IN BIOTECHNOLOGY, 1999, 17 (11) :443-447
[20]   Real-time determination of glucose consumption by live cells using a lab-on-valve system with an integrated microbioreactor [J].
Schulz, CM ;