Which manufacturing characteristics are predictors of outlet strut fracture in large sixty-degree Bjork-Shiley convexo-concave mitral valves?

被引:20
作者
Kallewaard, M
Algra, A
Defauw, J
van der Graaf, Y
机构
[1] Univ Utrecht, Sch Med, Julis Ctr Patient Oriented Res, NL-3508 GA Utrecht, Netherlands
[2] St Antonius Hosp, Dept Cardiothorac Surg, Nieuwegein, Netherlands
关键词
D O I
10.1016/S0022-5223(99)70298-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Identification of predictors of outlet strut fracture is important for recipients of large (greater than or equal to 29 mm) 60-degree Bjork-Shiley convexo-concave mitral valves when it comes to decision making on prophylactic explantation. An association between the manufacturing process of Bjork-Shiley convexo-concave valves and the risk of fracture has been suggested. Objective: The aim of this study was to determine which items from the manufacturing records, in addition to known risk factors, were predictive of fracture of large 60-degree Bjork-Shiley convexo-concave mitral valves. Methods: All Dutch recipients (n = 2264) of Bjork-Shiley convexo-concave valves were followed up until fracture, death, reoperation, or end of the study (July 1, 1996). Information was abstracted from the manufacturing records of large 60-degree Bjork-Shiley convexo-concave mitral valves (n = 655) in Dutch recipients and included items that described the manufacturing process and items for which an association with strut fracture had been suggested. Manufacturing records were available for 637 valves (97%), including 25 fractured valves. Results: Multivariate analysis identified age at implantation (hazard ratio 0.95, 95% confidence interval 0.93-0.97), lot size (<175 valves versus greater than or equal to 175 valves; hazard ratio 6.6, 95% confidence interval 2.2-20.1), number of hook deflection tests performed (0 or 1 versus greater than or equal to 2; hazard ratio 4.7, 95% confidence interval 1.4-16.2), number of disks that were used (1 versus greater than or equal to 2; hazard ratio 5.9, 95% confidence interval 1.9-18.5), and lot fracture percentage (hazard ratio 1.6, 95% confidence interval 1.4-1.8) as independent predictors of fracture. Although the added predictive value of a model with these 5 variables was sizable compared with a model containing age only, it was only slightly better than a model with age, lot size, and lot fracture percentage. Conclusion: If the serial number of a large 60-degree Bjork-Shiley convexo-concave mitral valve is known, manufacturing information can add significantly to the prediction of fracture. Information on lot size and lot fracture percentage should be made available to clinicians for risk assessment of prophylactic explantation.
引用
收藏
页码:766 / 775
页数:10
相关论文
共 20 条
[1]  
CARLEY WM, 1991, WALL STREET J
[2]  
COX DR, 1972, J R STAT SOC B, V34, P187
[3]   SINGLE-LEG STRUT FRACTURES IN EXPLANTED BJORK-SHILEY VALVES [J].
DEMOL, BA ;
KALLEWAARD, M ;
MCLELLAN, RB ;
VANHERWERDEN, LA ;
DEFAUW, JJ ;
VANDERGRAAF, Y .
LANCET, 1994, 343 (8888) :9-12
[4]   STRUT FRACTURE WITH BJORK-SHILEY 70-PERCENT CONVEXO-CONCAVE VALVE - AN INTERNATIONAL MULTIINSTITUTIONAL FOLLOW-UP-STUDY [J].
ERICSSON, A ;
LINDBLOM, D ;
SEMB, G ;
HUYSMANS, HA ;
THULIN, LI ;
SCULLY, HE ;
BENNETT, JG ;
OSTERMEYER, J ;
GRUNKEMEIER, GL .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 (07) :339-346
[5]   GETTING THE BAD-NEWS ABOUT YOUR ARTIFICIAL-HEART VALVE [J].
FIELDER, JH .
HASTINGS CENTER REPORT, 1993, 23 (02) :22-28
[6]   A METHOD OF COMPARING THE AREAS UNDER RECEIVER OPERATING CHARACTERISTIC CURVES DERIVED FROM THE SAME CASES [J].
HANLEY, JA ;
MCNEIL, BJ .
RADIOLOGY, 1983, 148 (03) :839-843
[7]   THE MEANING AND USE OF THE AREA UNDER A RECEIVER OPERATING CHARACTERISTIC (ROC) CURVE [J].
HANLEY, JA ;
MCNEIL, BJ .
RADIOLOGY, 1982, 143 (01) :29-36
[8]   OUTLET STRUT FRACTURE OF THE BJORK-SHILEY 60-DEGREES CONVEXO-CONCAVE VALVE - CURRENT INFORMATION AND RECOMMENDATIONS FOR PATIENT-CARE [J].
HIRATZKA, LF ;
KOUCHOUKOS, NT ;
GRUNKEMEIER, GL ;
MILLER, DC ;
SCULLY, HE ;
WECHSLER, AS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (05) :1130-1137
[9]   Welder identity, weld date, and the risk of outlet strut fracture in Bjork-Shiley convexo-concave valves: The Dutch cohort study [J].
Kallewaard, M ;
Algra, A ;
vanderGraaf, Y .
HEART, 1996, 76 (06) :510-512
[10]   NONPARAMETRIC-ESTIMATION FROM INCOMPLETE OBSERVATIONS [J].
KAPLAN, EL ;
MEIER, P .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1958, 53 (282) :457-481