Polyethylene wear on the distal tibial insert surface in total knee arthroplasty

被引:17
作者
Gabriel, SM
Dennis, DA
Honey, MJ
Scott, RD
机构
[1] Johnson & Johnson Orthopaed Inc, Raynham, MA 02767 USA
[2] Rose Musculoskeletal Res Lab, Denver, CO 80220 USA
[3] New England Baptist Hosp, Boston, MA 02120 USA
关键词
polyethylene wear; distal surface; tibial insert; retrieval study; metal backing;
D O I
10.1016/S0968-0160(97)10025-4
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We studied polyethylene wear damage on articular (superior) and distal (inferior) surfaces of 27 tibial inserts after an average of 50 months in vivo. The damage scoring method overestimated the subtle damage on the inferior surface, but the average superior surface score was still four times the inferior surface score. Inferior surface damage correlated significantly only to implantation time and was highest for inserts having metal backings with holes. Wear damage occurs on the inferior surface of tibial inserts. In the absence of holes in the metal backing and large insert/tray relative motions, however, this wear appears insignificant. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:221 / 228
页数:8
相关论文
共 19 条
[1]
POLYETHYLENE WEAR IN UNICONDYLAR KNEE PROSTHESES - 106 RETRIEVED MARMOR, PCA, AND ST-GEORG TIBIAL COMPONENTS COMPARED [J].
BLUNN, GW ;
JOSHI, AB ;
LILLEY, PA ;
ENGELBRECHT, E ;
RYD, L ;
LIDGREN, L ;
HARDINGE, K ;
NIEDER, E ;
WALKER, PS .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (03) :247-255
[2]
CHEN PC, 1995, CLIN ORTHOP RELAT R, P44
[3]
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P232
[4]
POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES [J].
ENGH, GA ;
DWYER, KA ;
HANES, CK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :9-17
[5]
GABRIEL SM, 1995, T COMB ORS, P114
[6]
RETRIEVAL ANALYSIS OF TOTAL KNEE PROSTHESES - A METHOD AND ITS APPLICATION TO 48 TOTAL CONDYLAR PROSTHESES [J].
HOOD, RW ;
WRIGHT, TM ;
BURSTEIN, AH .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (05) :829-842
[7]
POLYETHYLENE AND METAL DEBRIS GENERATED BY NON-ARTICULATING SURFACES OF MODULAR ACETABULAR COMPONENTS [J].
HUK, OL ;
MANJULA, B ;
BETTS, F ;
RIMNAC, CM ;
LIEBERMAN, JR ;
HUO, MH ;
SALVATI, EA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (04) :568-574
[8]
POLYETHYLENE WEAR IN UNCEMENTED KNEE REPLACEMENTS [J].
JONES, SMG ;
PINDER, IM ;
MORAN, CG ;
MALCOLM, AJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :18-22
[9]
KILGUS DJ, 1991, CLIN ORTHOP RELAT R, P223
[10]