Glutamine biosensors for biotechnology applications, with suppression of the endogenous glutamate signal

被引:20
作者
Madaras, MB [1 ]
Spokane, RB [1 ]
Johnson, JM [1 ]
Woodward, JR [1 ]
机构
[1] UNIV N CAROLINA,DEPT CHEM,CHAPEL HILL,NC 27599
关键词
D O I
10.1021/ac970173f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Glutamine membranes for amperometric measurements are described. The interference of the endogenous glutamate is greatly diminished by using a supplementary ''anti-glutamate'' layer consisting of immobilized glutamate oxidase and catalase on top of the glutamine-sensitive layer having co-immobilized glutaminase and glutamate oxidase. The use of polycarbonate membranes with different permeability characteristics for the control of the substrate's access to the enzyme layers is presented, as well as the effect of the density of the enzyme layer on the sensitivity of these membranes. The fabricated membranes have good operational stability (at least 5 days) and very good linearity (up to 10 mM glutamine). Using an appropriate choice of membranes and cross-linking conditions, membranes with good rejection of glutamate have been fabricated (less than 6% RSD for a 5 mM glutamine sample containing 5 mM glutamate as interferent). These membranes are suitable for monitoring of glutamine levels in mammalian cell cultures without the need of a separate measurement for glutamate.
引用
收藏
页码:3674 / 3678
页数:5
相关论文
共 23 条
[11]   MINIATURIZED THIN-FILM GLUTAMATE AND GLUTAMINE BIOSENSORS [J].
MOSER, I ;
JOBST, G ;
ASCHAUER, E ;
SVASEK, P ;
VARAHRAM, M ;
URBAN, G ;
ZANIN, VA ;
TJOUTRINA, GY ;
ZHARIKOVA, AV ;
BEREZOV, TT .
BIOSENSORS & BIOELECTRONICS, 1995, 10 (6-7) :527-532
[12]   Determination of glutamine and glutamic acid in mammalian cell cultures using tetrathiafulvalene modified enzyme electrodes [J].
Mulchandani, A ;
Bassi, AS .
BIOSENSORS & BIOELECTRONICS, 1996, 11 (03) :271-280
[13]   Interactive dual control of glucose and glutamine feeding in hybridoma cultivation [J].
Oh, GS ;
Izuishi, T ;
Inoue, T ;
Hu, WS ;
Yoshida, T .
JOURNAL OF FERMENTATION AND BIOENGINEERING, 1996, 81 (04) :329-336
[14]  
RECHNITZ GA, 1979, NATURE, V28, P466
[15]   ENZYME SENSOR-FIA-SYSTEM FOR ONLINE MONITORING OF GLUCOSE, LACTATE AND GLUTAMINE IN ANIMAL-CELL CULTURES [J].
RENNEBERG, R ;
TROTTKRIEGESKORTE, G ;
LIETZ, M ;
JAGER, V ;
PAWLOWA, M ;
KAISER, G ;
WOLLENBERGER, U ;
SCHUBERT, F ;
WAGNER, R ;
SCHMID, RD ;
SCHELLER, FW .
JOURNAL OF BIOTECHNOLOGY, 1991, 21 (1-2) :173-186
[16]  
ROMETTE JL, 1987, ANAL LETT, V20, P1069
[17]  
ROMETTE JL, 1989, Patent No. 4780191
[18]   USE OF IONOMER MEMBRANES TO ENHANCE THE SELECTIVITY OF ELECTRODE-BASED BIOSENSORS IN FLOW-INJECTION ANALYSIS [J].
ROSARIO, SA ;
CHA, GS ;
MEYERHOFF, ME ;
TROJANOWICZ, M .
ANALYTICAL CHEMISTRY, 1990, 62 (22) :2418-2424
[19]  
SCHALKHAMMER T, 1994, SENSOR ACTUAT B-CHEM, V18, P587
[20]   DETERMINATION OF GLUTAMINE IN SERUM USING AN AMPEROMETRIC ENZYME ELECTRODE [J].
VILLARTA, RL ;
PALLESCHI, G ;
SULEIMAN, A ;
GUILBAULT, GG .
ELECTROANALYSIS, 1992, 4 (01) :27-31