Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate

被引:227
作者
Moseley, Douglas J.
White, Elizabeth A.
Wiltshire, Kirsty L.
Rosewall, Tara
Sharpe, Michael B.
Siewerdsen, Jeffrey H.
Bissonnette, Jean-Pierre
Gospodarowicz, Mary
Warde, Padraig
Catton, Charles N.
Jaffray, David A.
机构
[1] Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON, Canada
[3] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2007年 / 67卷 / 03期
关键词
cone-beam CT; image-guided; prostate radiotherapy; fiducial markers; surrogates;
D O I
10.1016/j.ijrobp.2006.10.039
中图分类号
R73 [肿瘤学];
学科分类号
100214 [肿瘤学];
摘要
Purpose: The aim of this work was to assess the accuracy of kilovoltage (kV) cone-beam computed tomography (CBCT)-based setup corrections as compared with orthogonal megavoltage (MV) portal image-based corrections for patients undergoing external-beam radiotherapy of the prostate. Methods and Materials: Daily cone-beam CT volumetric images were acquired after setup for patients with three intraprostatic fiducial markers. The estimated couch shifts were compared retrospectively to patient adjustments based on two orthogonal MV portal images (the current clinical standard of care in our institution). The CBCT soft-tissue based shifts were also estimated by digitally removing the gold markers in each projection to suppress the artifacts in the reconstructed volumes. A total of 256 volumetric images for 15 patients were analyzed. Results: The Pearson coefficient of correlation for the patient position shifts using fiducial markers in MV vs. kV was (R-2 = 0.95, 0.84, 0.81) in the left-right (LR), anterior-posterior (AP), and superior-inferior (SI) directions, respectively. The correlation using soft-tissue matching was as follows: R-2 = 0.90, 0.49,0.51 in the LR, AP and SI directions. A Bland-Altman analysis showed no significant trends in the data. The percentage of shifts within a +/- 3-mm tolerance (the clinical action level) was 99.7%, 95.5%, 91.3% for fiducial marker matching and 99.5%, 70.3%, 78.4% for soft-tissue matching. Conclusions: Cone-beam CT is an accurate and precise tool for image guidance. It provides an equivalent means of patient setup correction for prostate patients with implanted gold fiducial markers. Use of the additional information provided by the visualization of soft-tissue structures is an active area of research. (c) 2007 Elsevier Inc.
引用
收藏
页码:942 / 953
页数:12
相关论文
共 37 条
[1]
Portal imaging for evaluation of daily on-line setup errors and off-line organ motion during conformal irradiation of carcinoma of the prostate [J].
Alasti, H ;
Petric, MP ;
Catton, CN ;
Warde, PR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (03) :869-884
[2]
Online ultrasound image guidance for radiotherapy of prostate cancer: Impact of image acquisition on prostate displacement [J].
Artignan, X ;
Smitsmans, MHP ;
Lebesque, JV ;
Jaffray, DA ;
van Her, M ;
Bartelink, H .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (02) :595-601
[3]
COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[4]
On-line aSi portal imaging of implanted fiducial markers for the reduction of interfraction error during conformal radiotherapy of prostate carcinoma [J].
Chung, PWM ;
Haycocks, T ;
Brown, T ;
Cambridge, Z ;
Kelly, V ;
Alasti, H ;
Jaffray, DA ;
Catton, CN .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01) :329-334
[5]
Evaluation of a contour-alignment technique for CT-guided prostate radiotherapy: An intra- and interobserver study [J].
Court, LE ;
Dong, L ;
Taylor, N ;
Ballo, M ;
Kitamura, K ;
Lee, AK ;
O'Daniel, J ;
White, RA ;
Cheung, R ;
Kuban, D .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (02) :412-418
[6]
Automatic registration of the prostate for computed-tomography-guided radiotherapy [J].
Court, LE ;
Dong, L .
MEDICAL PHYSICS, 2003, 30 (10) :2750-2757
[7]
Considerations for the implementation of target volume protocols in radiation therapy [J].
Craig, T ;
Battista, J ;
Moiseenko, V ;
Van Dyk, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (01) :241-250
[8]
Comparison of radiation side-effects of conformal and conventional radiotherapy in prostate cancer: a randomised trial [J].
Dearnaley, DP ;
Khoo, VS ;
Norman, AR ;
Meyer, L ;
Nahum, A ;
Tait, D ;
Yarnold, J ;
Horwich, A .
LANCET, 1999, 353 (9149) :267-272
[9]
Clinical feasibility study for the use of implanted gold seeds in the prostate as reliable positioning markers during megavoltage irradiation [J].
Dehnad, H ;
Nederveen, AJ ;
van der Heide, UA ;
van Moorselaar, RJA ;
Hofman, P ;
Lagendijk, JJW .
RADIOTHERAPY AND ONCOLOGY, 2003, 67 (03) :295-302
[10]
PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619