In this paper, an atmospheric pressure, helium microwave-induced plasma (MIP) ion source coupled with an orthogonal acceleration time-of-flight mass spectrometer (TOFMS) is explored for elemental analysis. Studies of the relationship between ion signals and sampling distance of the MS reveal that background signals can be suppressed dramatically without sacrificing the signal intensities of analytes when the microwave plasma plume is off the tip of the sampler orifice. This "off-cone" ion sampling mode provides a technique to obtain nearly "clean" background spectra and, thus, eliminates the spectral interference from entrainment air and the working-gas species, making it possible to sensitively determine isotopes that suffered from spectral interference in ICPMS and MIPMS (such as Ca-40, Cr-52, Mn-55, and Fe-56). On the other hand, since the high-temperature plasma is kept away from the sampler aperture, off-cone sampling places little demand on the cooling device and the lifetime of the sampler plate can be extended. The instrumental system can provide a fairly good mass resolution of 1100 (fwhm). The detection limits (3 sigma) in the tens of picograms per milliliter level for the elements studied can be achieved with a digital oscilloscope. These detection limits can be easily improved with an advanced detection system, which is currently available in commercial markets.
引用
收藏
页码:2455 / 2462
页数:8
相关论文
共 39 条
[1]   COMPARISON OF TEMPERATURE-MEASUREMENTS IN ICP AND MIP WITH AR AND HE AS PLASMA GAS [J].
ABDALLAH, MH ;
MERMET, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (05) :391-397
[2]   Determination of the rare earth elements in aqueous samples at sub-ppt levels by inductively coupled plasma mass spectrometry and flow injection ICPMS [J].
Aggarwal, JK ;
Shabani, MB ;
Palmer, MR ;
Ragnarsdottir, KV .
ANALYTICAL CHEMISTRY, 1996, 68 (24) :4418-4423
[3]   FUNDAMENTAL-STUDIES OF THE SAMPLING PROCESS IN AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER .4. REPLACEMENT OF THE INDUCTIVELY COUPLED PLASMA WITH A HELIUM MICROWAVE-INDUCED PLASMA [J].
CHAMBERS, DM ;
CARNAHAN, JW ;
JIN, Q ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (13) :1745-1765
[4]   A HELIUM INDUCTIVELY COUPLED PLASMA FOR ATOMIC EMISSION-SPECTROMETRY [J].
CHAN, SK ;
MONTASER, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1985, 40 (10-12) :1467-1472
[5]   ELEMENTAL ANALYSIS WITH AN ATMOSPHERIC-PRESSURE PLASMA (MIP, ICP) QUADRUPOLE MASS-SPECTROMETER SYSTEM [J].
DOUGLAS, DJ ;
QUAN, ESK ;
SMITH, RG .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1983, 38 (1-2) :39-48
[6]   Design and development of a highly sensitive, field portable plasma source instrument for on-line liquid stream monitoring and real-time sample analysis [J].
Duan, YX ;
Su, YX ;
Jin, Z ;
Abeln, SP .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (03) :1557-1563
[7]   TRACE ANALYSIS IN HIGH MATRIX AQUEOUS-SOLUTIONS USING HELIUM MICROWAVE-INDUCED PLASMA-MASS SPECTROMETRY [J].
FECHER, PA ;
NAGENGAST, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (09) :1021-1027
[8]   Toward the next generation of atomic mass spectrometers - Plenary lecture [J].
Hieftje, GM ;
Myers, DP ;
Li, GQ ;
Mahoney, PP ;
Burgoyne, TW ;
Ray, SJ ;
Guzowski, JP .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (03) :287-292
[9]   A NEW SENSITIVE MOLECULAR-BEAM MASS-SPECTROMETER [J].
HOGLUND, A ;
ROSENGREN, LG .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1984, 60 (SEP) :173-187
[10]  
HORLICK G, 1992, INDUCTIVELY COUPLED