Design and development of a highly sensitive, field portable plasma source instrument for on-line liquid stream monitoring and real-time sample analysis

被引:28
作者
Duan, YX
Su, YX
Jin, Z
Abeln, SP
机构
[1] Los Alamos Natl Lab, Chem Sci & Technol Div, Los Alamos, NM 87545 USA
[2] Los Alamos Natl Lab, Div Mat Sci & Technol, Los Alamos, NM 87545 USA
关键词
D O I
10.1063/1.1150501
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The development of a highly sensitive, field portable, low-powered instrument for on-site, real-time liquid waste stream monitoring is described in this article. A series of factors such as system sensitivity and portability, plasma source, sample introduction, desolvation system, power supply, and the instrument configuration, were carefully considered in the design of the portable instrument. A newly designed, miniature, modified microwave plasma source was selected as the emission source for spectroscopy measurement, and an integrated small spectrometer with a charge-coupled device detector was installed for signal processing and detection. An innovative beam collection system with optical fibers was designed and used for emission signal collection. Microwave plasma can be sustained with various gases at relatively low power, and it possesses high detection capabilities for both metal and nonmetal pollutants, making it desirable to use for on-site, real-time, liquid waste stream monitoring. An effective in situ sampling system was coupled with a high efficiency desolvation device for direct-sampling liquid samples into the plasma. A portable computer control system is used for data processing. The new, integrated instrument can be easily used for on-site, real-time monitoring in the field. The system possesses a series of advantages, including high sensitivity for metal and nonmetal elements; in situ sampling; compact structure; low cost; and ease of operation and handling. These advantages will significantly overcome the limitations of previous monitoring techniques and make great contributions to environmental restoration and monitoring. [S0034-6748(00)03103-8].
引用
收藏
页码:1557 / 1563
页数:7
相关论文
共 29 条
[1]   CAVITY FOR MICROWAVE-INDUCED PLASMAS OPERATED IN HELIUM AND ARGON AT ATMOSPHERIC-PRESSURE [J].
BEENAKKER, CIM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1976, 31 (8-9) :483-486
[2]   STABLE NITROGEN ISOTOPE ANALYSIS BY OPTICAL SPECTROSCOPY [J].
BROIDA, HP ;
CHAPMAN, MW .
ANALYTICAL CHEMISTRY, 1958, 30 (12) :2049-2055
[3]   COMPARATIVE-STUDIES OF SURFATRON AND MICROWAVE PLASMA TORCH SOURCES FOR DETERMINATION OF MERCURY BY ATOMIC-EMISSION SPECTROMETRY [J].
DUAN, YX ;
DU, XG ;
JIN, QH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (05) :629-633
[4]   ANALYTICAL PERFORMANCE OF THE MICROWAVE PLASMA TORCH IN THE DETERMINATION OF RARE-EARTH ELEMENTS WITH OPTICAL-EMISSION SPECTROMETRY [J].
DUAN, YX ;
LI, YM ;
TIAN, XD ;
ZHANG, HQ ;
JIN, QH .
ANALYTICA CHIMICA ACTA, 1994, 295 (03) :315-324
[5]   CHARACTERIZATION OF A MODIFIED, LOW-POWER ARGON MICROWAVE PLASMA TORCH (MPT) AS AN ATOMIZATION CELL FOR ATOMIC FLUORESCENCE SPECTROMETRY [J].
DUAN, YX ;
DU, XG ;
LI, YM ;
JIN, QH .
APPLIED SPECTROSCOPY, 1995, 49 (08) :1079-1085
[6]   VAPOR-GENERATION ASSISTED NEBULIZATION FOR NONMETAL DETERMINATION [J].
DUAN, YX ;
WU, M ;
JIN, WH ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (09) :971-974
[7]   VAPOR GENERATION OF NONMETALS COUPLED TO MICROWAVE PLASMA-TORCH MASS-SPECTROMETRY [J].
DUAN, YX ;
WU, M ;
JIN, QH ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (09) :1095-1108
[8]   EVALUATION OF A LOW-POWERED ARGON MICROWAVE PLASMA DISCHARGE AS AN ATOMIZER FOR THE DETERMINATION OF MERCURY BY ATOMIC FLUORESCENCE SPECTROMETRY [J].
DUAN, YX ;
KONG, XX ;
ZHANG, HQ ;
LIU, J ;
JIN, QH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (01) :7-10
[9]   DESOLVATION EFFECT ON THE ANALYTICAL PERFORMANCE OF MICROWAVE-INDUCED PLASMA-ATOMIC ABSORPTION SPECTROMETRY (MIP-AAS) [J].
DUAN, YX ;
ZHANG, HQ ;
HUO, MY ;
JIN, QH .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1994, 49 (06) :583-592
[10]  
FOSTER RW, 1995, PROCESS CONTR QUAL, V7, P79