Relationship between noise, dose, and pitch in cardiac multi-detector row CT

被引:133
作者
Primak, Andrew N. [1 ]
McCollough, Cynthia H. [1 ]
Bruesewitz, Michael R. [1 ]
Zhang, Jie [1 ]
Fletcher, Joel G. [1 ]
机构
[1] Mayo Clin, Coll Med, Dept Radiol, CT Clin Innovat Ctr, Rochester, MN 55905 USA
关键词
D O I
10.1148/rg.266065063
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In spiral computed tomography (CT), dose is always inversely proportional to pitch. However, the relationship between noise and pitch ( and hence noise and dose) depends on the scanner type ( single vs multi detector row) and reconstruction mode ( cardiac vs noncardiac). In single detector row spiral CT, noise is independent of pitch. Conversely, in noncardiac multi - detector row CT, noise depends on pitch because the spiral interpolation algorithm makes use of redundant data from different detector rows to decrease noise for pitch values less than 1 ( and increase noise for pitch values > 1). However, in cardiac spiral CT, redundant data cannot be used because such data averaging would degrade the temporal resolution. Therefore, the behavior of noise versus pitch returns to the single detector row paradigm, with noise being independent of pitch. Consequently, since faster rotation times require lower pitch values in cardiac multi - detector row CT, dose is increased without a commensurate decrease in noise. Thus, the use of faster rotation times will improve temporal resolution, not alter noise, and increase dose. For a particular application, the higher dose resulting from faster rotation speeds should be justified by the clinical benefits of the improved temporal resolution.
引用
收藏
页码:1785 / 1794
页数:10
相关论文
共 27 条
[1]   Assessment of coronary remodeling in stenotic and nonstenotic coronary atherosclerotic lesions by multidetector spiral computed tomography [J].
Achenbach, S ;
Ropers, D ;
Hofmann, U ;
MacNeill, B ;
Baum, U ;
Pohle, K ;
Brady, TJ ;
Pomerantsev, E ;
Ludwig, J ;
Flachskampf, FA ;
Wicky, S ;
Jang, IK ;
Daniel, WG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (05) :842-847
[2]   Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey [J].
Brix, G ;
Nagel, HD ;
Stamm, G ;
Veit, R ;
Lechel, U ;
Griebel, J ;
Galanski, M .
EUROPEAN RADIOLOGY, 2003, 13 (08) :1979-1991
[3]   Accuracy of 16-slice multi-detector CT to quantify the degree of coronary artery stenosis: Assessment of cross-sectional and longitudinal vessel reconstructions [J].
Cury, RC ;
Ferencik, M ;
Achenbach, S ;
Pomerantsev, E ;
Nieman, K ;
Moselewski, F ;
Abbara, S ;
Jang, IK ;
Brady, TJ ;
Hoffmann, U .
EUROPEAN JOURNAL OF RADIOLOGY, 2006, 57 (03) :345-350
[4]   FDA public health notification: Reducing radiation risk from computed tomography for pediatric and small adult patients [J].
Feigal, DW .
PEDIATRIC RADIOLOGY, 2002, 32 (04) :314-316
[5]   Image reconstruction and image quality evaluation for a 16-slice CT scanner [J].
Flohr, T ;
Stierstorfer, K ;
Bruder, H ;
Simon, J ;
Polacin, A ;
Schaller, S .
MEDICAL PHYSICS, 2003, 30 (05) :832-845
[6]   Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart [J].
Flohr, T ;
Ohnesorge, B .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2001, 25 (06) :907-923
[7]   Multi-detector row CT systems and image-reconstruction techniques [J].
Flohr, TG ;
Schaller, S ;
Stierstorfer, K ;
Bruder, H ;
Ohnesorge, BM ;
Schoepf, UJ .
RADIOLOGY, 2005, 235 (03) :756-773
[8]   First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[9]   Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot [J].
Flohr, TG ;
Stierstorfer, K ;
Ulzheimer, S ;
Bruder, H ;
Primak, AN ;
McCollough, CH .
MEDICAL PHYSICS, 2005, 32 (08) :2536-2547
[10]   Radiation exposure in spiral computed tomography -: Dose distribution and dose reduction [J].
Hidajat, N ;
Mäurer, J ;
Schröder, RJ ;
Wolf, M ;
Vogl, T ;
Felix, R .
INVESTIGATIVE RADIOLOGY, 1999, 34 (01) :51-57