Ten-year NEDO BVAD development program:: Moving forward to the clinical arena

被引:4
作者
Motomura, Tadashi
Okubo, Hisashi
Oda, Takeshi
Ogawa, Daisuke
Okahisa, Toshiya
Igo, Stephen
Shinohara, Toshiyuki
Yamamoto, Yoshiro
Noguchi, Chikaya
Ishizuka, Tsukasa
Okamoto, Eiji
Nose, Yukihiko
机构
[1] Baylor Coll Med, Michael E DeBakey Dept Surg, Houston, TX 77030 USA
[2] MIWATEC Co Ltd, Tokyo, Japan
[3] Softronics Co Ltd, Saitama, Japan
[4] Hokkaido Tokai Univ, Grad Sch Sci & Engn, Sapporo, Hokkaido, Japan
关键词
D O I
10.1097/01.mat.0000225263.48393.93
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Since 1995, the Baylor Group has been developing a totally implantable NEDO BVAD system. This 10-year program was completed in March 2005, and preparation for clinical trials is underway. This article summarizes the entire 10-year NEDO program and describes the strategy for clinical trials. The project aimed to achieve: (1) dual centrifugal pumps with the ability of full biventricular support, (2) a compact system implantable into small adults, (3) a totally implantable system with transcutaneous energy transmission system (TETS), (4) a durable system with a lifetime of over 5 years, and (5) a system free of thrombus and with minimal hemolysis. The final goals are to complete preclinical system evaluations and commence the clinical trials in the near future. In vitro studies have demonstrated a pump capacity of over 8.5 I/min and an index of Hemolysis of <0.004 g/100 l. The pump-bearing life expectancy was over 5 years. To date, eight pumps endured in vivo studies of over 3 months without complications, including thromboembolic events. The in vitro endurance studies of eight pumps are longer than 1 year. There were no mechanical malfunctions or pump failure. A stepwise clinical trial is being planned: Step1, a wearable BVAD/VAD will be clinically studied; Step 2, the BVAD/VAD will be implanted intracorporeally without TETS; and, Step 3, a totally implantable system will be clinically evaluated. The NEDO BVAD system has completed preclinical testing. Clinical trial preparation is underway.
引用
收藏
页码:378 / 385
页数:8
相关论文
共 22 条
[1]
Long-term mechanical circulatory support system reliability recommendation - American Society for Artificial Internal Organs and Society of Thoracic Surgeons: Long-term mechanical circulatory support system reliability recommendation [J].
Altieri, F ;
Berson, A ;
Borovetz, H ;
Butler, K ;
Byrd, G ;
Ciarkowski, AA ;
Dunn, R ;
Frazier, OHB ;
Griffith, B ;
Hoeppner, DW ;
Jassawalla, JS ;
Kormos, RH ;
Kung, RTV ;
Lemperle, B ;
Lewis, JP ;
Pantalos, GM ;
Pennington, DG ;
Poirier, VL ;
Portner, PM ;
Rosenberg, GG ;
Shanker, R ;
Watson, JT .
ASAIO JOURNAL, 1998, 44 (01) :108-114
[2]
Real-time studies of the pivot bearings in the NEDO gyro PI-710 centrifugal blood pump [J].
Asai, T ;
Watanabe, K ;
Ito, S ;
Tsujimura, S ;
Motomura, T ;
Shinohara, T ;
Glueck, JA ;
Nosé, Y .
ARTIFICIAL ORGANS, 2004, 28 (10) :899-903
[3]
*ASTM, 2005, STAND PRACT SEL BLOO, pF1830
[4]
Outpatients on biventricular assist devices [J].
Drews, T ;
Loebe, M ;
Jurmann, M ;
Dohnol, RZ ;
Erben, M ;
Hetzer, R .
THORACIC AND CARDIOVASCULAR SURGEON, 2001, 49 (05) :296-299
[5]
Clinical experience with an implantable, intracardiac, continuous flow circulatory support device: Physiologic implications and their relationship to patient selection [J].
Frazier, OH ;
Myers, TJ ;
Westaby, S ;
Gregoric, ID .
ANNALS OF THORACIC SURGERY, 2004, 77 (01) :133-141
[6]
In vivo evaluation of the NEDO biventricular assist device with an RPM dynamic impeller suspension system [J].
Ichikawa, S ;
Linneweber, J ;
Motomura, T ;
Ishitoya, H ;
Watanabe, K ;
Ashizawa, S ;
Murai, N ;
Nishimura, I ;
Sumikura, H ;
Glueck, JA ;
Shinohara, T ;
Oestmann, DJ ;
Nosé, Y .
ASAIO JOURNAL, 2003, 49 (05) :578-582
[7]
Titania gel reduces thrombin generation [J].
Ishitoya, H ;
Kawamura, M ;
Linneweber, J ;
Motomura, T ;
Ichikawa, S ;
Nishimura, I ;
Glueck, J ;
Shinohara, T ;
Nosé, Y .
ARTIFICIAL ORGANS, 2002, 26 (11) :959-963
[8]
Antithrombogenicity evaluation of a centrifugal blood pump [J].
Nakata, K ;
Yoshikawa, M ;
Takano, T ;
Maeda, T ;
Nonaka, K ;
Linneweber, J ;
Kawahito, S ;
Glueck, J ;
Fujisawa, A ;
Makinouchi, K ;
Yokokawa, M ;
Nosé, Y .
ARTIFICIAL ORGANS, 2000, 24 (08) :667-670
[9]
Terumo implantable left ventricular assist system: Results of long-term animal study [J].
Nojiir, C ;
Kijima, T ;
Maekawa, J ;
Horiuchi, K ;
Kido, T ;
Sugiyama, T ;
Mori, T ;
Sugiura, N ;
Asada, T ;
Ozaki, T ;
Suzuki, M ;
Akamatsu, T ;
Akutsu, T .
ASAIO JOURNAL, 2000, 46 (01) :117-122
[10]
Design and development strategy for the rotary blood pump [J].