A fast Look-Locker method for T 1 mapping of the head and neck region

被引:3
作者
Chikui, Toru [1 ]
Tokumori, Kenji [1 ]
Zeze, Ryosuke [2 ]
Shiraishi, Tomoko [2 ]
Ichihara, Takahiro [3 ]
Hatakenaka, Masamitsu [4 ]
Yoshiura, Kazunori [1 ]
机构
[1] Kyushu Univ, Fac Dent Sci, Dept Oral & Maxillofacial Radiol, Higashi Ku, Fukuoka 8128582, Japan
[2] Fukuoka Dent Coll, Dept Diagnost & Gen Care, Sect Image Diag, Sawara Ku, Fukuoka, Japan
[3] Fukuoka Dent Coll Med & Dent Hosp, Dept Radiol, Sawara Ku, Fukuoka, Japan
[4] Kyushu Univ, Fac Med Sci, Dept Clin Radiol, Higashi Ku, Fukuoka 8128582, Japan
关键词
T(1) map; Look-Locker sequence; Head and neck region; CONTRAST; QUANTIFICATION; OPTIMIZATION; RESOLUTION; GLANDS; NMR;
D O I
10.1007/s11282-009-0005-1
中图分类号
R78 [口腔科学];
学科分类号
100302 [口腔临床医学];
摘要
A sequence for T (1) relaxation time measurements that allows a high-resolution image to be obtained within a short acquisition time is described. Its application to the orofacial region is investigated. The sequence is based on the Look-Locker (LL) method and employs a magnetization-preparation pulse prior to data acquisition, with either a turbo-field echo (TFE) or turbo-field echo echo-planar imaging (TFEPI) sequence. We applied the multiple LL sequence by synchronizing the data acquisition with the virtual electrocardiogram. T (1) results from the LL sequence were compared with those from the inversion recovery (IR) sequence. In an in vitro study, we evaluated the T (1) maps of contrast medium at different concentrations. In an in vivo study, we evaluated the T (1) maps in seven volunteers. In the in vitro study, the correlation between the T (1) values obtained from the LL sequence and those from the IR sequence was high, with an R (2) of more than 0.99. For T (1) values between 200 and 1,500 ms, the difference between the two methods was less than 7%. In the in vivo study, a high correlation was observed between the T (1) values from the IR sequence and those from the LL sequence. The scan duration for the LL sequence was less than 3 min. Our method, based on the LL sequence, was found to be clinically useful for producing a T (1) map.
引用
收藏
页码:22 / 29
页数:8
相关论文
共 19 条
[1]
Correlation of dynamic contrast-enhanced MR imaging with histologic tumor grade: Comparison of macromolecular and small-molecular contrast media [J].
Daldrup, H ;
Shames, DM ;
Wendland, M ;
Okuhata, Y ;
Link, TM ;
Rosenau, W ;
Lu, Y ;
Brasch, RC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 171 (04) :941-949
[2]
QUANTIFICATION OF T1 VALUES BY SNAPSHOT-FLASH NMR IMAGING [J].
DEICHMANN, R ;
HAASE, A .
JOURNAL OF MAGNETIC RESONANCE, 1992, 96 (03) :608-612
[3]
Optimization of the ultrafast Look-Locker echo-planar imaging T1 mapping sequence [J].
Freeman, AJ ;
Gowland, PA ;
Mansfield, P .
MAGNETIC RESONANCE IMAGING, 1998, 16 (07) :765-772
[4]
Musculoskeletal MRI at 3.0T: Relaxation times and image contrast [J].
Gold, GE ;
Han, E ;
Stainsby, J ;
Wright, G ;
Brittain, J ;
Beaulieu, C .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 183 (02) :343-351
[5]
DYNAMIC STUDIES OF GADOLINIUM UPTAKE IN BRAIN-TUMORS USING INVERSION-RECOVERY ECHO-PLANAR IMAGING [J].
GOWLAND, P ;
MANSFIELD, P ;
BULLOCK, P ;
STEHLING, M ;
WORTHINGTON, B ;
FIRTH, J .
MAGNETIC RESONANCE IN MEDICINE, 1992, 26 (02) :241-258
[6]
PRACTICAL IMPLEMENTATION AND OPTIMIZATION OF ONE-SHOT T1 IMAGING [J].
KAY, I ;
HENKELMAN, RM .
MAGNETIC RESONANCE IN MEDICINE, 1991, 22 (02) :414-424
[7]
Three-dimensional T1 mapping for dGEMRIC at 3.0 T using the look locker method [J].
Kimelman, T ;
Vu, A ;
Storey, P ;
McKenzie, C ;
Burstein, D ;
Prasad, P .
INVESTIGATIVE RADIOLOGY, 2006, 41 (02) :198-203
[8]
CONTINUOUS MEASUREMENT OF CELL-VOLUME CHANGES IN PERFUSED RAT SALIVARY-GLANDS BY PROTON NMR [J].
LARCOMBEMCDOUALL, JB ;
SEO, Y ;
STEWARD, MC .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (02) :131-138
[9]
Look D. C., 1970, Review of Scientific Instruments, V41, P250, DOI 10.1063/1.1684482
[10]
Human myocardium: Single-breath-hold MR T1 mapping with high spatial resolution - Reproducibility study [J].
Messroghli, DR ;
Plein, S ;
Higgins, DM ;
Walters, K ;
Jones, TR ;
Ridgway, JP ;
Sivananthan, MU .
RADIOLOGY, 2006, 238 (03) :1004-1012