The novel HemoCue® Plasma/Low Hemoglobin System accurately measures small concentrations of three different hemoglobin-based oxygen carriers in plasma:: Hemoglobin Glutamer-200 (Bovine) (Oxyglobin®), Hemoglobin Glutamer-250 (Bovine) (Hemopure®), and Hemoglobin-Raffimer (Hemolink™)

被引:18
作者
Lurie, F
Jahr, JS
Driessen, B
机构
[1] Univ Calif Los Angeles, Sch Med, Dept Anesthesiol, Los Angeles, CA 90095 USA
[2] Univ Hawaii, John A Burns Sch Med, Dept Surg, Honolulu, HI 96822 USA
[3] Charles R Drew Univ Med & Sci, Martin Luther King Jr Drew Med Ctr, Dept Anesthesiol, Los Angeles, CA 90059 USA
[4] Univ Penn, Sch Vet Med, Dept Clin Studies, Philadelphia, PA 19104 USA
关键词
D O I
10.1097/00000539-200210000-00014
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The accuracy of the HemoCue(R) Plasma/Low Hemoglobin System was validated in vitro with low levels of hemoglobin-based oxygen carriers (HBOCs). Repeated measurements were performed on 50 samples of canine plasma, each mixed with three different HBOCs at varying small concentrations (a total of 150 samples), by using plasma samples without HBOCs as controls. Two technicians performed the measurements and randomly tested each sample 10 times. The results were analyzed for correlation, and analysis of variance was used to evaluate statistical significance, with a P value of less than or equal to 0.05 considered significant. Hemoglobin concentrations determined with the bedside photometer were not significantly different from known values of hemoglobin concentration in the samples. There was no significant difference between values obtained by two independent observers for the same samples. This was true for all three tested HBOCs and for all tested concentrations. The mean bias of the measurement expressed as a percentage of sample concentration was 0.1% for hemoglobin glutamer-200 (bovine), 0.58% for hemoglobin glutamer-250 (bovine), and 0.19% for hemoglobin-raffimer. The mean error was <8% for all three HBOCs. Both intraobserver and interobserver reliabilities were high and statistically significant. The HemoCue(R) Plasma/Low Hemoglobin System is a reliable instrument for detecting and measuring small concentrations of three different HBOCs in plasma.
引用
收藏
页码:870 / 873
页数:4
相关论文
共 20 条
[1]  
*AKT LEO DIAGN, 2000, HEM PLASM LOW HEM SY
[2]  
Ali AA, 2001, ANESTH ANALG, V92, P863
[3]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[4]  
British Standards Institution, 1979, 5497 BS 1
[5]   Effects of haemoglobin-based oxygen carrier Hemoglobin glutamer-200 (bovine) on intestinal perfusion and oxygenation in a canine hypovolaemia model [J].
Driessen, B ;
Jahr, JS ;
Lurie, F ;
Griffey, SM ;
Gunther, RA .
BRITISH JOURNAL OF ANAESTHESIA, 2001, 86 (05) :683-692
[6]  
Jahr J S, 2000, Clin Lab Sci, V13, P210
[7]  
Jahr J S, 2000, Clin Lab Sci, V13, P173
[8]   The HemoCue®, a point of care B-hemoglobin photometer, measures hemoglobin concentrations accurately when mixed in vitro with canine plasma and three hemoglobin-based oxygen carriers (HBOC) [J].
Jahr, JS ;
Lurie, F ;
Driessen, B ;
Davis, JA ;
Gosselin, R ;
Gunther, RA .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2002, 49 (03) :243-248
[9]   Oxygen saturation measurements in canine blood containing hemoglobin glutamer-200 (bovine): In vitro validation of the NOVA CO-Oximeter [J].
Jahr, JS ;
Driessen, B ;
Lurie, F ;
Tang, ZP ;
Louie, RF ;
Kost, G .
VETERINARY CLINICAL PATHOLOGY, 2001, 30 (01) :39-45
[10]   A novel approach to measuring circulating blood volume: The use of a hemoglobin-based oxygen carrier in a rabbit model [J].
Jahr, JS ;
Lurie, F ;
Xi, S ;
Golkaryeh, M ;
Kuznetsova, O ;
Kullar, R ;
Driessen, B .
ANESTHESIA AND ANALGESIA, 2001, 92 (03) :609-614