Sensitive, high throughput HPLC-MS/MS method with on-line sample clean-up for everolimus measurement

被引:38
作者
Korecka, Magdalena [1 ]
Solari, Sandra G. [1 ]
Shaw, Leslie M. [1 ]
机构
[1] Univ Penn, Dept Pathol & Lab Med, Sch Med, Philadelphia, PA 19104 USA
关键词
everolimus; SDZ RAD 227-756; ascomycin; on-line sample clean-up; switching valve; HPLC-MS/MS;
D O I
10.1097/00007691-200608000-00002
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A new HPLC-MS/MS method for everolimus measurement was developed that includes the following features: small sample volume, short run time, fast, simple and cost-efficient sample preparation, assessment of performance of two internal standards (IS), SDZ RAD 223-756 and ascomycin and comparison of the method with an HPLC-MS/MS reference method. The authors established a multiple reaction monitoring positive ion HPLC-MS/MS method with on-line extraction and sample cleanup. This procedure includes: an API 2000 triple quadrupole mass spectrometer with turbo-ion spray, built-in Valco switching valve, an HPLC system; guard column; a Nova-Pak C18 analytical column; washing solution, methanol:30mM ammonium acetate pH 5.1 (80:20); eluting solution, methanol:30 mM ammonium acetate pH 5.1 (97:3); flow rate 0.8 mL/min; and a run time of 2.8 minutes. The first and third quadrupoles were set to detect the ammonium adduct ion and a high mass fragment of everolimus (m/z 975.5 -> 908.5), and two ISs: SDZ RAD 223-756 (m/z 989.8 -> 922.8) and ascomycin (m/z 809.5-756.5). The LLOQ was 1.0 mu g/L for everolimus using either IS. Between day precision ranged from 3.1% to 5.7% for SDZ RAD 223-756 and 6.0% to 8.6% for ascomycin using spiked blood with everolimus concentrations 2.0 to 25.0 mu g/L. Absolute recoveries using spiked samples over the range of 2.5 to 25 mu g/L averaged 77.3% (SDZ RAD 223-756) and 76.8% (ascomycin). No matrix effect on everolimus was demonstrated based on the mean observed signal detection of 98.6% (SDZ RAD 223-756) and 105% (ascomycin). Comparison of everolimus concentrations obtained using this method with two internal standards with a reference laboratory demonstrated that the mean everolimus concentration obtained with ascomycin was statistically different (lower) than results with the reference method and the method that used SDZ RAD 223-756 as the internal standard gave equivalent results compared with the reference method.
引用
收藏
页码:484 / 490
页数:7
相关论文
共 33 条
[1]   Ion suppression in mass spectrometry [J].
Annesley, TM .
CLINICAL CHEMISTRY, 2003, 49 (07) :1041-1044
[2]  
[Anonymous], 2001, BIOAN METH VAL GUID
[3]   Experience with everolimus [J].
Augustine, JJ ;
Hricik, DE .
TRANSPLANTATION PROCEEDINGS, 2004, 36 (02) :500S-503S
[4]   High-performance liquid chromatography with ultraviolet detection for therapeutic drug monitoring of everolimus [J].
Baldelli, S ;
Murgia, S ;
Merlini, S ;
Zenoni, S ;
Perico, N ;
Remuzzi, G ;
Cattaneo, D .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 816 (1-2) :99-105
[5]   The in vivo effect of rapamycin derivative SDZ RAD on lymphocyte proliferation [J].
Böhler, T ;
Waiser, J ;
Budde, K ;
Lichter, S ;
Jauho, A ;
Fritsche, L ;
Korn, A ;
Neumayer, HH .
TRANSPLANTATION PROCEEDINGS, 1998, 30 (05) :2195-2197
[6]   High-throughput semi-automated 96-well liquid/liquid extraction and liquid chromatography/mass spectrometric analysis of everolimus (RAD 001) and cyclosporin a (CsA) in whole blood [J].
Brignol, N ;
McMahon, LM ;
Luo, S ;
Tse, FLS .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (12) :898-907
[7]   Evaluation of internal standards for the measurement of Cyclosporin A by Hplc-esi-ms/ms [J].
Brown, SR ;
Taylor, PJ ;
Cooper, DP ;
Salm, P ;
Pillans, PI ;
Lynch, SV .
THERAPEUTIC DRUG MONITORING, 2005, 27 (02) :243-243
[8]   Tolerability and steady-state pharmacokinetics of everolimus in maintenance renal transplant patients [J].
Budde, K ;
Neumayer, HH ;
Lehne, G ;
Winkler, M ;
Hauser, IA ;
Lison, A ;
Fritsche, L ;
Soulillou, JP ;
Fauchald, P ;
Dantal, J .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2004, 19 (10) :2606-2614
[9]  
Crowe A, 1999, DRUG METAB DISPOS, V27, P627
[10]   Simultaneous quantification of sirolimus, everolimus, tacrolimus and cyclosporine by liquid chromatography-mass spectrometry (LC-MS) [J].
Deters, M ;
Kirchner, G ;
Resch, K ;
Kaever, V .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2002, 40 (03) :285-292