Body composition analysis by dual X-ray absorptiometry:: in vivo and in vitro comparison of three different fan-beam instruments

被引:41
作者
Aasen, G. [1 ]
Fagertun, H. [1 ]
Halse, J. [1 ]
机构
[1] Capturo AS, Kjeller, Norway
关键词
bone mineral content (BMC); conversion factors; DXA; fat mass; lean body mass;
D O I
10.1080/00365510600898214
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background. Dual X-ray absorptiometry (DXA) is the preferred method for measuring body composition in clinical practice, but interchange between devices may pose problems with the interpretation of results. Objective. To establish conversion equations for body composition variables between three fan-beam DXA systems. Methods. Body composition was assessed in 21 subjects using Lunar Expert ( Expert), Lunar Prodigy ( Prodigy) and Hologic Delphi W ( Delphi). Weekly measurements of Hologic whole body phantom 164 were performed. Results. There were no significant differences between DXA-measured means of body weight, fat mass and lean body mass. Bland-Altman analysis revealed that Lunar Expert increasingly overestimated fat mass with increasing total mass (p < 0.001) relative to Delphi and Prodigy, while Delphi produced a constant underestimation of fat mass. Correlations between scale weights and DXA-measured body weights, and between DXA-measured body weights and the sum of fat masses, lean body masses and bone mineral contents (BMC)between the three instruments, were excellent (Rsqr 0.998-0.910; p < 0.001). Conversion factors to Prodigy for Expert and Delphi were respectively 1.003 and 1.011 for total body mass, 0.954 and 1.079 for fat mass, 1.018 and 0.967 for lean body mass and 1.049 and 1.136 for BMC ( Rsqr 0.999-0.991; p < 0.001). Standard error of estimate ( SEE) for the slopes ranged from 0.20% to 2.10%. Phantom studies revealed stable instrument function with CV% commonly < 2%, except for lean mass for Delphi (5.5%). Conclusions. Despite the significant differences in measurement of body composition between DXA fan-beam instruments, clinically relevant conversion factors can be established.
引用
收藏
页码:659 / 666
页数:8
相关论文
共 26 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Evaluation of a new body composition phantom for quality control and cross-calibration of DXA devices [J].
Diessel, E ;
Fuerst, T ;
Njeh, CF ;
Tylavsky, F ;
Cauley, J ;
Dockrell, M ;
Genant, HK .
JOURNAL OF APPLIED PHYSIOLOGY, 2000, 89 (02) :599-605
[3]   Bone mineral and body composition measurements: Cross-calibration of pencil-beam and fan-beam dual-energy X-ray absorptiometers [J].
Ellis, KJ ;
Shypailo, RJ .
JOURNAL OF BONE AND MINERAL RESEARCH, 1998, 13 (10) :1613-1618
[4]   Comparison of bone mineral density and body composition measurements in women obtained from two DXA instruments [J].
Gillette-Guyonnet, S ;
Andrieu, S ;
Nourhashemi, F ;
Cantet, C ;
Grandjean, H ;
Vellas, B .
MECHANISMS OF AGEING AND DEVELOPMENT, 2003, 124 (03) :317-321
[5]   Body composition analysis by DEXA by using dynamically changing samarium filtration [J].
Gotfredsen, A ;
Baeksgaard, L ;
Hilsted, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1997, 82 (04) :1200-1209
[6]   VALIDATION OF BODY-COMPOSITION BY DUAL ENERGY X-RAY ABSORPTIOMETRY (DEXA) [J].
HAARBO, J ;
GOTFREDSEN, A ;
HASSAGER, C ;
CHRISTIANSEN, C .
CLINICAL PHYSIOLOGY, 1991, 11 (04) :331-341
[7]  
Hansen RD, 1999, AM J CLIN NUTR, V70, P228
[8]   Change in fat-free mass assessed by bioelectrical impedance, total body potassium and dual energy X-ray absorptiometry during prolonged weight loss [J].
Hendel, HW ;
Gotfredsen, A ;
Hojgaard, L ;
Andersen, T ;
Hilsted, J .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1996, 56 (08) :671-679
[9]   DETERMINATION OF BODY-COMPOSITION - A COMPARISON OF DUAL-ENERGY X-RAY ABSORPTIOMETRY AND HYDRODENSITOMETRY [J].
JOHANSSON, AG ;
FORSLUND, A ;
SJODIN, A ;
MALLMIN, H ;
HAMBRAEUS, L ;
LJUNGHALL, S .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1993, 57 (03) :323-326
[10]   Differences in whole body measurements by DXA-scanning using two Lunar DPX-L machines [J].
Lantz, H ;
Samuelson, G ;
Bratteby, LE ;
Mallmin, H ;
Sjöström, L .
INTERNATIONAL JOURNAL OF OBESITY, 1999, 23 (07) :764-770