Model-based normalization for iterative 3D PET image reconstruction

被引:52
作者
Bai, B
Li, Q
Holdsworth, CH
Asma, E
Tai, YC
Chatziioannou, A
Leahy, RM
机构
[1] Univ So Calif, Signal & Image Proc Inst, Los Angeles, CA 90089 USA
[2] Univ Calif Los Angeles, Crump Inst Mol Imaging, Los Angeles, CA 90095 USA
[3] Washington Univ, Sch Med, St Louis, MO 63110 USA
关键词
D O I
10.1088/0031-9155/47/15/316
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We describe a method for normalization in 3D PET for use with maximum a posteriori (MAP) or other iterative model-based image reconstruction methods. This approach is an extension of previous factored normalization methods in which we include separate factors for detector sensitivity, geometric response, block effects and deadtime. Since our MAP reconstruction approach already models some of the geometric factors in the forward projection, the normalization factors must be modified to account only for effects not already included in the model. We describe a maximum likelihood approach to joint estimation of the count-rate independent normalization factors, which we apply to data from a uniform cylindrical source. We then compute block-wise and block-profile deadtime correction factors using singles and coincidence data, respectively, from a multiframe cylindrical source. We have applied this method for reconstruction of data from the Concorde microPET P4 scanner. Quantitative evaluation of this method using well-counter measurements of activity in a multicompartment phantom compares favourably with normalization based directly on cylindrical source measurements.
引用
收藏
页码:2773 / 2784
页数:12
相关论文
共 12 条
[1]   Developments in component-based normalization for 3D PET [J].
Badawi, RD ;
Marsden, PK .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02) :571-594
[2]  
BADAWI RD, 1998, THESIS U LONDON
[3]  
Casey M. E., 1995, Proceedings of the 1995 International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, P67
[4]   A NORMALIZATION TECHNIQUE FOR 3D PET DATA [J].
DEFRISE, M ;
TOWNSEND, DW ;
BAILEY, D ;
GEISSBUHLER, A ;
MICHEL, C ;
JONES, T .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :939-952
[5]   Grouped-coordinate ascent algorithms for penalized-likelihood transmission image reconstruction [J].
Fessler, JA ;
Ficaro, EP ;
Clinthorne, NH ;
Lange, K .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :166-175
[6]   A STUDY OF DATA LOSS AND MISPOSITIONING DUE TO PILEUP IN 2-D DETECTORS IN PET [J].
GERMANO, G ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :671-675
[7]   PET SYSTEM CALIBRATIONS AND CORRECTIONS FOR QUANTITATIVE AND SPATIALLY ACCURATE IMAGES [J].
HOFFMAN, EJ ;
GUERRERO, TM ;
GERMANO, G ;
DIGBY, WM ;
DAHLBOM, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :1108-1112
[8]   Performance analysis of an improved 3-D PET Monte Carlo simulation and scatter correction [J].
Holdsworth, CH ;
Levin, CS ;
Janecek, M ;
Dahlbom, M ;
Hoffman, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (01) :83-89
[9]   FAST GRADIENT-BASED METHODS FOR BAYESIAN RECONSTRUCTION OF TRANSMISSION AND EMISSION PET IMAGES [J].
MUMCUOGLU, EU ;
LEAHY, R ;
CHERRY, SR ;
ZHOU, ZY .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :687-701
[10]   DETECTOR EFFICIENCY AND COMPTON SCATTER IN FULLY 3D PET [J].
OLLINGER, JM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1168-1173