Urea rebound and delivered Kt/V determination with a continuous urea sensor

被引:25
作者
Garred, LJ
Canaud, B
Bosc, JY
Tetta, C
机构
[1] LAPEYRONIE HOSP, SERV NEPHROL, MONTPELLIER, FRANCE
[2] BELLCO SPA, MIRANDOLA, ITALY
关键词
haemodialysis; Kt/V; mathematical modelling; rebound; urea kinetic modelling; urea sensor;
D O I
10.1093/ndt/12.3.535
中图分类号
R3 [基础医学]; R4 [临床医学];
学科分类号
1001 ; 1002 ; 100602 ;
摘要
Background. The recent introduction of urea sensors for dialysis monitoring has made possible new approaches to urea kinetic modelling. In this study we show how the equilibrated postdialysis urea concentration (C-eq) and Kt/V corrected for double-pool urea kinetics (Kt/V-dp) can be accurately determined using an on-line sensor providing a continuous measure of blood water urea. A modification of the Smye constant volume double-pool theory led to the following equations for C-eq and Kt/V-dp C-eq = C-pre exp(S(ex)t) [GRAPHICS] where C-pre is the blood concentration measured at the start of dialysis, t is the length of the dialysis session (in min) and S-ex is the constant slope of the blood urea logarithm concentration decline following development of the intercompartmental urea concentration gradient in the first 30-60 min of dialysis. Methods, These equations were tested in 11 patients undergoing 165-240 min of paired filtration dialysis with continuous monitoring of blood urea concentration. C-pre was determined as the plateau concentration during a preliminary period of 15-20 min of slow isolated ultrafiltration. S-ex was accurately determined from linear regression applied to the urea sensor data from the 80-min point to the end of dialysis. Results. C-eq and Kt/V-dp determined from the above equations compared closely to values determined from 25-40 min of urea rebound monitoring with the urea sensor: 10.6 +/- 3.0 versus 10.8 +/- 2.7 mmol/l (mean +/- SD) for C-eq and 1.21 +/- 0.24 versus 1.18 +/- 0.20 for Kt/V-dp, compared to single-pool values of Kt/V = 1.34 +/- 0.23. Conclusion, This technique may be readily programmed into on-line urea monitors to provide current and extrapolated values of C-eq and Kt/V-dp from about the first hour of dialysis.
引用
收藏
页码:535 / 542
页数:8
相关论文
共 15 条
  • [1] PERCENT REDUCTION IN BLOOD UREA CONCENTRATION DURING DIALYSIS ESTIMATES KT/V IN A SIMPLE AND ACCURATE WAY
    BASILE, C
    CASINO, F
    LOPEZ, T
    [J]. AMERICAN JOURNAL OF KIDNEY DISEASES, 1990, 15 (01) : 40 - 45
  • [2] DAUGIRDAS JT, 1993, J AM SOC NEPHROL, V4, P1205
  • [3] DAUGIRDAS JT, 1989, INT J ARTIF ORGANS, V12, P411
  • [4] Simple Kt/V formulas based on urea mass balance theory
    Garred, L.J.
    Barichello, D.L.
    DiGiuseppe, B.
    McCready, W.G.
    Canaud, B.C.
    [J]. ASAIO Journal, 1994, 40 (04) : 997 - 1004
  • [5] Garred L J, 1995, Adv Ren Replace Ther, V2, P305
  • [6] GARRED LJ, 1995, ASAIO J, V41, pM804, DOI 10.1097/00002480-199507000-00126
  • [7] KT/V AND PROTEIN CATABOLIC RATE DETERMINATION FROM SERIAL UREA MEASUREMENT IN THE DIALYSATE EFFLUENT STREAM
    GARRED, LJ
    DIGIUSEPPE, B
    CHAND, W
    MCCREADY, W
    CANAUD, B
    [J]. ARTIFICIAL ORGANS, 1992, 16 (03) : 248 - 255
  • [8] Ghezzi P M, 1983, Life Support Syst, V1 Suppl 1, P271
  • [9] JINDAL KK, 1987, T AM SOC ART INT ORG, V33, P286
  • [10] ONLINE MONITORING OF THE DELIVERY OF THE HEMODIALYSIS PRESCRIPTION
    KESHAVIAH, PR
    EBBEN, JP
    EMERSON, PF
    [J]. PEDIATRIC NEPHROLOGY, 1995, 9 : S2 - S8