Automatic quality assessment protocol for MRI equipment

被引:37
作者
Bourel, P
Gibon, D
Coste, E
Daanen, V
Rousseau, J
机构
[1] Ctr Hosp Reg & Univ Lille, Inst Med Technol, F-59037 Lille, France
[2] Oscar Lambret Oncol Ctr, Dept Radiat Therapy, Lille, France
[3] Univ Sci & Tech Lille Flandres Artois, I3D Lab, Villeneuve Dascq, France
[4] Ctr Hosp Reg & Univ Lille, MRI Dept, F-59037 Lille, France
关键词
D O I
10.1118/1.598809
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The authors have developed a protocol and software for the quality assessment of MRT equipment with a commercial test object. Automatic image analysis consists of detecting surfaces and objects, defining regions of interest, acquiring reference point coordinates and establishing gray level profiles. Signal-to-noise ratio, image uniformity, geometrical distortion, slice thickness, slice profile, and spatial resolution are checked. The results are periodically analyzed to evaluate possible drifts with time. The measurements are performed weekly on three MRI scanners made by the Siemens(TM) Company (VISION(R) 1.5T, EXPERT(R) 1.0T, and OPEN(R) 0.2T). The results obtained for the three scanners over approximately 3.5 years are presented, analyzed, and compared. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00812-3].
引用
收藏
页码:2693 / 2700
页数:8
相关论文
共 15 条
[1]  
BUSHBERG JT, 1994, ESSENTIAL PHYSICS ME, P127
[2]   Information in magnetic resonance images: Evaluation of signal, noise and contrast [J].
DeWilde, JP ;
Lunt, JA ;
Straughan, K .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (03) :259-265
[3]   MRI SCANNER VARIABILITY STUDIES USING A SEMIAUTOMATED ANALYSIS SYSTEM [J].
HYDE, RJ ;
ELLIS, JH ;
GARDNER, EA ;
ZHANG, YT ;
CARSON, PL .
MAGNETIC RESONANCE IMAGING, 1994, 12 (07) :1089-1097
[4]   MEASURING SIGNAL-TO-NOISE RATIOS IN MR IMAGING [J].
KAUFMAN, L ;
KRAMER, DM ;
CROOKS, LE ;
ORTENDAHL, DA .
RADIOLOGY, 1989, 173 (01) :265-267
[6]   The absolute signal-to-noise ratio in MRI acceptance testing [J].
McRobbie, DW .
BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (827) :1045-1048
[7]   A three-dimensional volumetric test object for geometry evaluation in magnetic resonance imaging [J].
McRobbie, DW .
MEDICAL PHYSICS, 1997, 24 (05) :737-742
[8]  
*NEMA, 1988, DET SIGN NOIS RAT SN
[9]  
*NEMA, 1989, DET IM UN DIAGN MAGN
[10]  
*NEMA, 1989, DET 2 DIM GEOM DIST