A recently developed microcantilever probe with integrated piezoresistive readout has been applied as a gas sensor. Resistors, sensitive to stress changes, are integrated on the flexible cantilevers. This makes it possible to monitor the cantilever deflection electrically and with an integrated reference cantilever background noise is subtracted directly in the measurement. A polymer coated cantilever has been exposed to vapors of various alcohols and the resulting cantilever response has been interpreted using a simple evaporation model. The model indicates that the cantilever response is a direct measure of the molecular concentration of alcohol vapor. On the basis of the model the detection limit of this cantilever-based sensor is determined to be below 10 ppm for alcohol vapor measurements. Furthermore, the time response of the cantilever can be used to distinguish between different alcohols due to a difference in the evaporation rates. (C) 2000 American Institute of Physics. [S0003-6951(00)02118-5].
引用
收藏
页码:2615 / 2617
页数:3
相关论文
共 11 条
[1]   PHOTOTHERMAL SPECTROSCOPY WITH FEMTOJOULE SENSITIVITY USING A MICROMECHANICAL DEVICE [J].
BARNES, JR ;
STEPHENSON, RJ ;
WELLAND, ME ;
GERBER, C ;
GIMZEWSKI, JK .
NATURE, 1994, 372 (6501) :79-81
[2]   Surface stress in the self-assembly of alkanethiols on gold [J].
Berger, R ;
Delamarche, E ;
Lang, HP ;
Gerber, C ;
Gimzewski, JK ;
Meyer, E ;
Guntherodt, HJ .
SCIENCE, 1997, 276 (5321) :2021-2024
[3]   Environmental sensors based on micromachined cantilevers with integrated read-out [J].
Boisen, A ;
Thaysen, J ;
Jensenius, H ;
Hansen, O .
ULTRAMICROSCOPY, 2000, 82 (1-4) :11-16
[4]   A NONDESTRUCTIVE METHOD FOR DETERMINING THE SPRING CONSTANT OF CANTILEVERS FOR SCANNING FORCE MICROSCOPY [J].
CLEVELAND, JP ;
MANNE, S ;
BOCEK, D ;
HANSMA, PK .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (02) :403-405
[5]   OBSERVATION OF A CHEMICAL-REACTION USING A MICROMECHANICAL SENSOR [J].
GIMZEWSKI, JK ;
GERBER, C ;
MEYER, E ;
SCHLITTLER, RR .
CHEMICAL PHYSICS LETTERS, 1994, 217 (5-6) :589-594
[6]   A chemical sensor based on a micromechanical cantilever array for the identification of gases and vapors [J].
Lang, HP ;
Berger, R ;
Battiston, F ;
Ramseyer, JP ;
Meyer, E ;
Andreoli, C ;
Brugger, J ;
Vettiger, P ;
Despont, M ;
Mezzacasa, T ;
Scandella, L ;
Guntherodt, HJ ;
Gerber, C ;
Gimzewski, JK .
APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 1998, 66 (Suppl 1) :S61-S64
[7]   The evaporation of small spheres. [J].
Langmuir, I .
PHYSICAL REVIEW, 1918, 12 (05) :368-370
[8]   Atomic force microscopy stress sensors for studies in liquids [J].
OShea, SJ ;
Welland, ME ;
Brunt, TA ;
Ramadan, AR ;
Rayment, T .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1996, 14 (02) :1383-1385
[9]  
THAYSEN J, 1999, TRANSDUCERS 99, P1852
[10]   DETECTION OF MERCURY-VAPOR USING RESONATING MICROCANTILEVERS [J].
THUNDAT, T ;
WACHTER, EA ;
SHARP, SL ;
WARMACK, RJ .
APPLIED PHYSICS LETTERS, 1995, 66 (13) :1695-1697