The influence of head frame distortions on stereotactic localization and targeting

被引:18
作者
Treuer, H [1 ]
Hunsche, S [1 ]
Hoevels, M [1 ]
Luyken, K [1 ]
Maarouf, M [1 ]
Voges, J [1 ]
Sturm, V [1 ]
机构
[1] Univ Cologne, Dept Stereotaxy & Funct Neurosurg, D-50924 Cologne, Germany
关键词
D O I
10.1088/0031-9155/49/17/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A strong attachment of a stereotactic head frame to the patient's skull may cause distortions of the head frame. The aim of this work was to identify possible distortions of the head frame, to measure the degree of distortion occurring in clinical practice and to investigate its influence on stereotactic localization and targeting. A model to describe and quantify the distortion of the Riechert-Mundinger (RM) head frame was developed. Distortions were classified as (a) bending and (b) changes from the circular ring shape. Ring shape changes were derived from stereotactic CT scans and frame bending was determined from intraoperative stereotactic x-ray images of patients with implanted I-125-seeds acting as landmarks. From the examined patient data frame bending was determined to be 0.74 mm 0.32 mm and 1.30 mm in maximum. If a CT-localizer with a top ring is used, frame bending has no influence on stereotactic CT-localization. In stereotactic x-ray localization, frame bending leads to an overestimation of the z-coordinate by 0.37 mm +/- 0.16 mm on average and by 0.65 mm in maximum. The accuracy of patient positioning in radiosurgery is not affected by frame bending. But in stereotactic surgery with an RM aiming bow trajectory displacements are expected. These displacements were estimated to be 0.36 mm +/- 0.16 mm (max. 0.74 mm) at the target point and 0.65 mm +/- 0.30 mm (max. 1.31 mm) at the entry point level. Changes from the circular ring shape are small and do not compromise the accuracy of stereotactic targeting and localization. The accuracy of CT-localization was found to be close to the resolution limit due to voxel size. Our findings for frame bending of the RM frame could be validated by statistical analysis and by comparison with an independent patient examination. The results depend on the stereotactic system and details of the localizers and instruments and also reflect our clinical practice. Therefore, a generalization is not possible. Preliminary experience with a new MR-compatible RM head frame made of ceramics shows no frame distortions as with the conventional frame made of an Al-Cu-Mg alloy.
引用
收藏
页码:3877 / 3887
页数:11
相关论文
共 17 条
[1]   An algorithm for stereotactic localization by computed tomography or magnetic resonance imaging [J].
Dai, JR ;
Zhu, YP ;
Qu, HM ;
Hu, YM .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (01) :N1-N7
[2]   Accuracy of true frameless stereotaxy: in vivo measurement and laboratory phantom studies - Technical note [J].
Dorward, NL ;
Alberti, O ;
Palmer, JD ;
Kitchen, ND ;
Thomas, DGT .
JOURNAL OF NEUROSURGERY, 1999, 90 (01) :160-168
[3]   PRECISION AND ACCURACY OF STEREOTAXIC CONVERGENT-BEAM IRRADIATIONS FROM A LINEAR-ACCELERATOR [J].
HARTMANN, GH ;
BAUERKIRPES, B ;
SERAGO, CF ;
LORENZ, WJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (02) :481-492
[4]   Stereotactic imaging for radiotherapy:: accuracy of CT, MRI, PET and SPECT [J].
Karger, CP ;
Hipp, P ;
Henze, M ;
Echner, G ;
Höss, A ;
Schad, L ;
Hartmann, GH .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (02) :211-221
[5]  
KLEIN D, 2003, THESIS U COLOGNE COL
[6]   THE APPLICATION ACCURACY OF STEREOTAXIC FRAMES [J].
MACIUNAS, RJ ;
GALLOWAY, RL ;
LATIMER, JW .
NEUROSURGERY, 1994, 35 (04) :682-694
[7]   Quality assurance in stereotactic space.: A system test for verifying the accuracy of aim in radiosurgery [J].
Mack, A ;
Czempiel, H ;
Kreiner, HJ ;
Dürr, G ;
Wowra, B .
MEDICAL PHYSICS, 2002, 29 (04) :561-568
[8]   PHOTOGRAMMETRIC ACCURACY MEASUREMENTS OF HEAD HOLDER SYSTEMS USED FOR FRACTIONATED RADIOTHERAPY [J].
MENKE, M ;
HIRSCHFELD, F ;
MACK, T ;
PASTYR, O ;
STURM, V ;
SCHLEGEL, W .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 29 (05) :1147-1155
[9]  
MUNDINGER F, 1998, TXB STEREOTACTIC FUN, P73
[10]   Frame slippage verification in stereotactic radiosurgery [J].
Otto, K ;
Fallone, BG .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (01) :199-205