Ultrafast MR-imaging of lung ventilation using hyperpolarized helium-3.

被引:55
作者
Schreiber, WG
Weiler, N
Kauczor, HU
Markstaller, K
Eberle, B
Hast, J
Surkau, R
Grossmann, T
Deninger, A
Hanisch, G
Otten, EW
Thelen, M
机构
[1] Univ Mainz, Klin & Poliklin Radiol, D-55131 Mainz, Germany
[2] Univ Mainz, Anasthesiol Klin, D-55131 Mainz, Germany
[3] Univ Mainz, Inst Phys, D-55131 Mainz, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2000年 / 172卷 / 02期
关键词
lung ventilation; hyperpolarized gas; magnetic resonance; ultrafast imaging; helium; human study;
D O I
10.1055/s-2000-10508
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Assessment of the temporal: and spatial dynamics of hyperpolarized Helium-3 (3He) distribution in the lung with ultrafast gradient-echo magnetic-resonance imaging, Material and Methods: Coronal images of the lung were,acquired using ultrafast gradient-echo pulse sequences with TR/TE = 3.3 ms/1.3 ms (slice thickenss, 40 mm) and TR/TE = 2.0 ms/0.7 ms (without slice selection). A series of 80 or 160 projection images was obtained with 210 ms or 130 ms temporal resolution, respectively. Imaging was performed during several respiratory cycles after application of a single bolus of 300 mL hyperpolarized 3He. Measurements were performed in six healthy volunteers (spontaneous breathing). Results: Different phases of in- and expiration could be visualized.; During the course of consecutive respiratory cycles the 3He signal decreased due to dilution of 3He in residual alveolar gas and by inspired air, relaxation due to oxygen and the RF pulses; and due to Helium-3 washout. The signal of a single bolus of 3He was detected in the lung for up to four respiratory cycles. Anatomical structures were better visualized on slice selective images than on images without slice selection. Conclusion: Distribution of inspired 3He within the tracheobronchial tree and alveolar space and its washout can be visualized by ultrafast imaging of a single bolus of hyperpolarized 3He gas. This method may allow for regional analysis of lung function with temporal and spatial resolution superior to conventional methods.
引用
收藏
页码:129 / 133
页数:5
相关论文
共 14 条
[1]  
ALBERT MS, 1994, NATURE, V370, P199, DOI 10.1038/370199a0
[2]  
CHEN XJ, 1998, P INT SOC MAGN RES M
[3]   Lung air spaces:: MR imaging evaluation with hyperpolarized 3He gas [J].
de Lange, EE ;
Mugler, JP ;
Brookeman, JR ;
Knight-Scott, J ;
Truwit, JD ;
Teates, CD ;
Daniel, TM ;
Bogorad, PL ;
Cates, GD .
RADIOLOGY, 1999, 210 (03) :851-857
[4]   Quantification of regional intrapulmonary oxygen partial pressure evolution during apnea by 3He MRI [J].
Deninger, AJ ;
Eberle, B ;
Ebert, M ;
Grossmann, T ;
Heil, W ;
Kauczor, HU ;
Lauer, L ;
Markstaller, K ;
Otten, E ;
Schmiedeskamp, J ;
Schreiber, W ;
Surkau, R ;
Thelen, M ;
Weiler, N .
JOURNAL OF MAGNETIC RESONANCE, 1999, 141 (02) :207-216
[5]   Analysis of intrapulmonary O2 concentration by NIR imaging of inhaled hyperpolarized helium-3 [J].
Eberle, B ;
Weiler, N ;
Markstaller, K ;
Kauczor, HU ;
Deninger, A ;
Ebert, M ;
Grossmann, T ;
Heil, W ;
Lauer, LO ;
Roberts, TPL ;
Schreiber, WG ;
Surkau, R ;
Dick, WF ;
Otten, EW ;
Thelen, M .
JOURNAL OF APPLIED PHYSIOLOGY, 1999, 87 (06) :2043-2052
[6]   COMPUTER-TOMOGRAPHY AND FUNCTIONAL PULMONARY SCINTIGRAPHY FOR CENTRAL BRONCHIAL CARCINOMAS - A COMPARISON OF MORPHOLOGICAL AND FUNCTIONAL-CHANGES [J].
HERTER, M ;
KONIG, R .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NUKLEARMEDIZIN, 1983, 138 (06) :707-715
[7]   He-3-MRI of pulmonary ventilation: First clinical applications. [J].
Kauczor, HU ;
Ebert, M ;
Kreitner, KF ;
Grossmann, T ;
Nilgens, H ;
Hofmann, D ;
Surkau, R ;
Roberts, T ;
Heil, W ;
Otten, EW ;
Thelen, M .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 166 (03) :192-198
[8]   Normal and abnormal pulmonary ventilation: Visualization at hyperpolarized He-3 MR imaging [J].
Kauczor, HU ;
Hofmann, D ;
Kreitner, KF ;
Nilgens, H ;
Surkau, R ;
Heil, W ;
Potthast, A ;
Knopp, MV ;
Otten, EW ;
Thelen, M .
RADIOLOGY, 1996, 201 (02) :564-568
[9]   MR-IMAGING WITH HYPERPOLARIZED HE-3 GAS [J].
MIDDLETON, H ;
BLACK, RD ;
SAAM, B ;
CATES, GD ;
COFER, GP ;
GUENTHER, R ;
HAPPER, W ;
HEDLUND, LW ;
JOHNSON, GA ;
JUVAN, K ;
SWARTZ, J .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (02) :271-275
[10]   NUCLEAR-RELAXATION OF HE-3 IN THE PRESENCE OF O-2 [J].
SAAM, B ;
HAPPER, W ;
MIDDLETON, H .
PHYSICAL REVIEW A, 1995, 52 (01) :862-865