Effect of plaque volume on subsequent vessel remodeling at edges of Sirolimus-eluting stents

被引:15
作者
Asano, Tatsuhiko
Kobayashi, Yoshio [1 ]
Mintz, Gary S.
Ishio, Naoki
Fujimaki, Shigenori
Ogawa, Yasuhide
Nakayama, Takashi
Kuroda, Nakabumi
Komuro, Issei
机构
[1] Chiba Univ, Grad Sch Med, Dept Cardiovasc Sci & Med, Chiba 280, Japan
[2] Cardiovasc Res Fdn, New York, NY USA
关键词
D O I
10.1016/j.amjcard.2006.05.021
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Serial (baseline and 9-month follow-up) intravascular ultrasound analysis was performed at 5-mm reference segments immediately proximal and distal to the sirolimus-eluting stent (SES) in 33 lesions. Proximal and distal reference segments were divided into 1-mm subsegments. Between postintervention and follow-up intravascular ultrasound studies, there were significant decreases in the lumen and increases in plaque & media areas in the subsegment closest to the distal edge, with no change in external elastic membrane area. There was no significant change in external elastic membrane, lumen, and plaque & media areas within the other subsegments. At the nearest 1-mm subsegment from the proximal and distal edges, baseline plaque & media area was associated with subsequent vessel remodeling. In conclusion, a large amount of plaque at the SES edge may be a risk of negative remodeling at follow-up (stent edge restenosis). It supports the importance of "normal-to-normal" SES deployment. (c) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:1041 / 1044
页数:4
相关论文
共 10 条
[1]   Poststenting axial redistribution of atherosclerotic plaque into the reference segments and lumen reduction at the stent edge - A volumetric intravascular ultrasound study [J].
Algowhary, M ;
Matsumura, A ;
Hashimoto, Y ;
Isobe, M .
INTERNATIONAL HEART JOURNAL, 2006, 47 (02) :159-171
[2]   Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions [J].
Colombo, A ;
Drzewiecki, J ;
Banning, A ;
Grube, E ;
Hauptmann, K ;
Silber, S ;
Dudek, D ;
Fort, S ;
Schiele, F ;
Zmudka, K ;
Guagliumi, G ;
Russell, ME .
CIRCULATION, 2003, 108 (07) :788-794
[3]   Preliminary observations regarding angiographic pattern of restenosis after rapamycin-eluting stent implantation [J].
Colombo, A ;
Orlic, D ;
Stankovic, G ;
Corvaja, N ;
Spanos, V ;
Montorfano, M ;
Liistro, F ;
Carlino, M ;
Airoldi, F ;
Chieffo, A ;
Di Mario, C .
CIRCULATION, 2003, 107 (17) :2178-2180
[4]   Serial intravascular ultrasound predictors of restenosis at the margins of Palmaz-Schatz stents [J].
Hoffmann, R ;
Mintz, GS ;
Kent, KM ;
Satler, LF ;
Pichard, AD ;
Popma, JJ ;
Leon, MB .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (07) :951-&
[5]   Delivered dose and vascular response after β-radiation for in-stent restenosis -: Retrospective dosimetry and volumetric intravascular ultrasound analysis [J].
Morino, Y ;
Kaneda, H ;
Fox, T ;
Takagi, A ;
Hassan, AHM ;
Bonan, R ;
Crocker, I ;
Lansky, AJ ;
Laskey, WK ;
Suntharalingam, M ;
Bonneau, HN ;
Yock, PG ;
Honda, Y ;
Fitzgerald, PJ .
CIRCULATION, 2002, 106 (18) :2334-2339
[6]   Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery [J].
Moses, JW ;
Leon, MB ;
Popma, JJ ;
Fitzgerald, PJ ;
Holmes, DR ;
O'Shaughnessy, C ;
Caputo, RP ;
Kereiakes, DJ ;
Williams, DO ;
Teirstein, PS ;
Jaeger, JL ;
Kuntz, RE .
NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (14) :1315-1323
[7]   Predictors of edge stenosis following sirolimus-eluting stent deployment (A quantitative intravascular ultrasound analysis from the SIRIUS trial) [J].
Sakurai, R ;
Ako, J ;
Morino, Y ;
Sonoda, S ;
Kaneda, H ;
Terashima, M ;
Hassan, AHM ;
Leon, MB ;
Moses, JW ;
Popma, JJ ;
Bonneau, HN ;
Yock, PG ;
Fitzgerald, PJ ;
Honda, Y .
AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (09) :1251-1253
[8]   Vascular responses at proximal and distal edges of paclitaxel-eluting stents - Serial intravascular ultrasound analysis from the TAXUS II trial [J].
Serruys, PW ;
Degertekin, M ;
Tanabe, K ;
Russell, ME ;
Guagliumi, G ;
Webb, J ;
Hamburger, J ;
Rutsch, W ;
Kaiser, C ;
Whitbourn, R ;
Camenzind, E ;
Meredith, I ;
Reeves, F ;
Nienaber, C ;
Benit, E ;
Disco, C ;
Koglin, J ;
Colombo, A .
CIRCULATION, 2004, 109 (05) :627-633
[9]   Impact of final Stent dimensions on long-term results following sirolimus-eluting stent implantation - Serial intravascular ultrasound analysis from the SIRIUS trial [J].
Sonoda, S ;
Morino, Y ;
Ako, J ;
Terashima, M ;
Hassan, AHM ;
Bonneau, HN ;
Leon, MB ;
Moses, JW ;
Yock, PG ;
Honda, Y ;
Kuntz, RE ;
Fitzgerald, PJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (11) :1959-1963
[10]   Intravascular ultrasound assessment of lesions with target vessel failure after sirolimus-eluting stent implantation [J].
Takebayashi, H ;
Kobayashi, Y ;
Mintz, GS ;
Carlier, SG ;
Fujii, K ;
Yasuda, T ;
Moussa, I ;
Mehran, R ;
Dangas, GD ;
Collins, MB ;
Kreps, E ;
Lansky, AJ ;
Stone, GW ;
Leon, MB ;
Moses, JW .
AMERICAN JOURNAL OF CARDIOLOGY, 2005, 95 (04) :498-502