Correlation between target volume and electron transport effects affecting heterogeneity corrections in stereotactic body radiotherapy for lung cancer

被引:12
作者
Akino, Yuichi [1 ]
Das, Indra J. [1 ]
Cardenes, Higinia R. [1 ]
Desrosiers, Colleen M. [1 ]
机构
[1] Indiana Univ Sch Med, Dept Radiat Oncol, Indianapolis, IN 46202 USA
基金
日本学术振兴会;
关键词
heterogeneity corrections; lung cancer; stereotactic body radiotherapy; radiation dosimetry; ANALYTICAL ANISOTROPIC ALGORITHM; PHOTON DOSE CALCULATION; MONTE-CARLO; RADIATION-THERAPY; DOSIMETRIC VERIFICATION; INHOMOGENEITY CORRECTIONS; VALIDATION; IMRT;
D O I
10.1093/jrr/rrt231
中图分类号
Q [生物科学];
学科分类号
090105 [作物生产系统与生态工程];
摘要
Recently, stereotactic body radiotherapy (SBRT) for lung cancer is conducted with heterogeneity-corrected treatment plans, as the correction greatly affects the dose delivery to the lung tumor. In this study, the correlation between the planning target volume (PTV) and the dose delivery is investigated by separation of the heterogeneity correction effects into photon attenuation and electron transport. Under Institutional Review Board exemption status, 74 patients with lung cancer who were treated with SBRT were retrospectively evaluated. All treatment plans were generated using an anisotropic analytical algorithm (AAA) of an Eclipse (Varian Medical Systems, Palo Alto, CA) treatment planning system. Two additional plans were created using the same treatment parameters (monitor units, beam angles and energy): a plan with no heterogeneity correction (NC), and a plan calculated with a pencil beam convolution algorithm (PBC). Compared with NC, AAA and PBC isocenter doses were on average 13.4% and 21.8% higher, respectively. The differences in the isocenter dose and the dose coverage for 95% of the PTV (D-95%) between PBC and AAA were correlated logarithmically (rho = 0.78 and rho = 0.46, respectively) with PTV. Although D-95% calculated with AAA was in general 2.9% larger than that for NC, patients with a small PTV showed a negative Delta D-95% for AAA due to the significant effect of electron transport. The PTV volume shows logarithmic correlation with the effects of the lateral electron transport. These findings indicate that the dosimetric metrics and prescription, especially in clinical trials, should be clearly evaluated in the context of target volume characteristics and with proper heterogeneity correction.
引用
收藏
页码:754 / 760
页数:7
相关论文
共 26 条
[1]
The effect of different lung densities on the accuracy of various radiotherapy dose calculation methods: Implications for tumour coverage [J].
Aarup, Lasse Rye ;
Nahum, Alan E. ;
Zacharatou, Christina ;
Juhler-Nottrup, Trine ;
Knoos, Tommy ;
Nystrom, Hakan ;
Specht, Lena ;
Wieslander, Elinore ;
Korreman, Stine S. .
RADIOTHERAPY AND ONCOLOGY, 2009, 91 (03) :405-414
[2]
COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[3]
Dosimetric verification of the anisotropic analytical algorithm for radiotherapy treatment planning [J].
Bragg, Christopher M. ;
Conway, John .
RADIOTHERAPY AND ONCOLOGY, 2006, 81 (03) :315-323
[4]
Dosimetric validation of Acuros® XB with Monte Carlo methods for photon dose calculations [J].
Bush, K. ;
Gagne, I. M. ;
Zavgorodni, S. ;
Ansbacher, W. ;
Beckham, W. .
MEDICAL PHYSICS, 2011, 38 (04) :2208-2221
[5]
Small fields:: Nonequilibrium radiation dosimetry [J].
Das, Indra J. ;
Ding, George X. ;
Ahnesjo, Anders .
MEDICAL PHYSICS, 2008, 35 (01) :206-215
[6]
Impact of inhomogeneity corrections on dose coverage in the treatment of lung cancer using stereotactic body radiation therapy [J].
Ding, George X. ;
Duggan, Dennis M. ;
Lu, Bo ;
Hallahan, Dennis E. ;
Cmelak, Anthony ;
Malcolm, Arnold ;
Newton, Jared ;
Deeley, Matthew ;
Coffey, Charles W. .
MEDICAL PHYSICS, 2007, 34 (07) :2985-2994
[7]
Dosimetric validation of the anisotropic analytical algorithm for photon dose calculation: fundamental characterization in water [J].
Fogliata, A ;
Nicolini, G ;
Vanetti, E ;
Clivio, A ;
Cozzi, L .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (06) :1421-1438
[8]
Recommendations for implementing stereotactic radiotherapy in peripheral stage IA non-small cell lung cancer: report from the Quality Assurance Working Party of the randomised phase III ROSEL study [J].
Hurkmans, Coen W. ;
Cuijpers, Johan P. ;
Lagerwaard, Frank J. ;
Widder, Joachim ;
van der Heide, Uulke A. ;
Schuring, Danny ;
Senan, Suresh .
RADIATION ONCOLOGY, 2009, 4
[9]
Comparison of inhomogeneity correction algorithms in small photon fields [J].
Jones, AO ;
Das, IJ .
MEDICAL PHYSICS, 2005, 32 (03) :766-776
[10]
A Monte Carlo study of IMRT beamlets in inhomogeneous media [J].
Jones, AO ;
Das, IJ ;
Jones, FL .
MEDICAL PHYSICS, 2003, 30 (03) :296-300