A comparison of two methods for measuring the signal to noise ratio on MR images

被引:213
作者
Firbank, MJ [1 ]
Coulthard, A
Harrison, RM
Williams, ED
机构
[1] Univ Newcastle Upon Tyne, Royal Victoria Infirm, Dept Radiol, Newcastle Upon Tyne NE1 4LP, Tyne & Wear, England
[2] Newcastle Gen Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE4 6BE, Tyne & Wear, England
[3] Sunderland Royal Hosp, Dept Reg Med Phys, Sunderland SR4 7TP, England
关键词
D O I
10.1088/0031-9155/44/12/403
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The signal to noise ratio (SNR) is one of the important measures of the performance of a magnetic resonance imaging (MRI) system. The object of this study was to compare a single acquisition method, which estimates the noise from background pixels, with a dual acquisition method which estimates the noise from the subtraction of two sequentially acquired images. The dual acquisition method is more exact, but is slower to perform and requires image manipulation. A comparison between the two methods gave a good correlation, and a regression equation of SNRsingle = 1.1 + 0.94 SNRdual. The single acquisition method is therefore appropriate for use in a quality assurance programme, since it is quicker and simpler to perform and is a good indicator of the more exact measure.
引用
收藏
页码:N261 / N264
页数:4
相关论文
共 9 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   MEASURING SIGNAL-TO-NOISE RATIOS IN MR IMAGING [J].
KAUFMAN, L ;
KRAMER, DM ;
CROOKS, LE ;
ORTENDAHL, DA .
RADIOLOGY, 1989, 173 (01) :265-267
[4]   The absolute signal-to-noise ratio in MRI acceptance testing [J].
McRobbie, DW .
BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (827) :1045-1048
[5]   QUALITY ASSURANCE METHODS AND PHANTOMS FOR MAGNETIC-RESONANCE-IMAGING - REPORT OF AAPM NUCLEAR-MAGNETIC-RESONANCE TASK GROUP NO-1 [J].
PRICE, RR ;
AXEL, L ;
MORGAN, T ;
NEWMAN, R ;
PERMAN, W ;
SCHNEIDERS, N ;
SELIKSON, M ;
WOOD, M ;
THOMAS, SR .
MEDICAL PHYSICS, 1990, 17 (02) :287-295
[6]   Signal-to-noise ratio in MRI [J].
Redpath, TW .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (847) :704-707
[7]   Estimation of the noise in magnitude MR images [J].
Sijbers, J ;
den Dekker, AJ ;
Van Audekerke, J ;
Verhoye, M ;
Van Dyck, D .
MAGNETIC RESONANCE IMAGING, 1998, 16 (01) :87-90
[8]   SOURCES OF INTENSITY NONUNIFORMITY IN SPIN-ECHO IMAGES AT 1.5-T [J].
SIMMONS, A ;
TOFTS, PS ;
BARKER, GJ ;
ARRIDGE, SR .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (01) :121-128
[9]   CORRECTION OF INTENSITY NONUNIFORMITY IN MR IMAGES OF ANY ORIENTATION [J].
WICKS, DAG ;
BARKER, GJ ;
TOFTS, PS .
MAGNETIC RESONANCE IMAGING, 1993, 11 (02) :183-196