The effect of chronic intracortical microstimulation on the electrode-tissue interface

被引:44
作者
Chen, Kevin H. [1 ]
Dammann, John F. [1 ]
Boback, Jessica L. [1 ]
Tenore, Francesco V. [2 ]
Otto, Kevin J. [3 ,4 ]
Gaunt, Robert A. [5 ,6 ]
Bensmaia, Sliman J. [1 ,7 ]
机构
[1] Univ Chicago, Dept Organismal Biol & Anat, Chicago, IL 60637 USA
[2] Johns Hopkins Univ, Appl Phys Lab, Res & Exploratory Dev Dept, Laurel, MD USA
[3] Purdue Univ, Dept Biol Sci, W Lafayette, IN 47907 USA
[4] Purdue Univ, Weldon Sch Biomed Engn, W Lafayette, IN 47907 USA
[5] Univ Pittsburgh, Dept Phys Med & Rehabil, Pittsburgh, PA USA
[6] Univ Pittsburgh, Dept Bioengn, Pittsburgh, PA USA
[7] Univ Chicago, Comm Computat Neurosci, Chicago, IL 60637 USA
关键词
intracortical microstimulation; electrode-tissue interface; impedance; voltage excursion; ELECTRICAL-STIMULATION; NEURAL STIMULATION; UNIT RECORDINGS; IMPEDANCE; PULSES; TERM;
D O I
10.1088/1741-2560/11/2/026004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Somatosensation is critical for effective object manipulation, but current upper limb prostheses do not provide such feedback to the user. For individuals who require use of prosthetic limbs, this lack of feedback transforms a mundane task into one that requires extreme concentration and effort. Although vibrotactile motors and sensory substitution devices can be used to convey gross sensations, a direct neural interface is required to provide detailed and intuitive sensory feedback. The viability of intracortical microstimulation (ICMS) as a method to deliver feedback depends in part on the long-term reliability of implanted electrodes used to deliver the stimulation. The objective of the present study is to investigate the effects of chronic ICMS on the electrode-tissue interface. Approach. We stimulate the primary somatosensory cortex of three Rhesus macaques through chronically implanted electrodes for 4 h per day over a period of six months, with different electrodes subjected to different regimes of stimulation. We measure the impedance and voltage excursion as a function of time and of ICMS parameters. We also test the sensorimotor consequences of chronic ICMS by having animals grasp and manipulate small treats. Main results. We show that impedance and voltage excursion both decay with time but stabilize after 10-12 weeks. The magnitude of this decay is dependent on the amplitude of the ICMS and, to a lesser degree, the duration of individual pulse trains. Furthermore, chronic ICMS does not produce any deficits in fine motor control. Significance. The results suggest that chronic ICMS has only a minor effect on the electrode-tissue interface and may thus be a viable means to convey sensory feedback in neuroprosthetics.
引用
收藏
页数:13
相关论文
共 43 条
[21]   Neuronal loss due to prolonged controlled-current stimulation with chronically implanted microelectrodes in the cat cerebral cortex [J].
McCreery, Douglas ;
Pikov, Victor ;
Troyk, Philip R. .
JOURNAL OF NEURAL ENGINEERING, 2010, 7 (03)
[22]   In vitro comparison of sputtered iridium oxide and platinum-coated neural implantable microelectrode arrays [J].
Negi, S. ;
Bhandari, R. ;
Rieth, L. ;
Solzbacher, F. .
BIOMEDICAL MATERIALS, 2010, 5 (01) :15007
[23]   Neural electrode degradation from continuous electrical stimulation: Comparison of sputtered and activated iridium oxide [J].
Negi, Sandeep ;
Bhandari, Rajmohan ;
Rieth, Loren ;
Van Wagenen, Rick ;
Solzbacher, Florian .
JOURNAL OF NEUROSCIENCE METHODS, 2010, 186 (01) :8-17
[24]   Electrical stimulation causes rapid changes in electrode impedance of cell-covered electrodes [J].
Newbold, Carrie ;
Richardson, Rachael ;
Millard, Rodney ;
Seligman, Peter ;
Cowan, Robert ;
Shepherd, Robert .
JOURNAL OF NEURAL ENGINEERING, 2011, 8 (03)
[25]   Virtual Active Touch Using Randomly Patterned Intracortical Microstimulation [J].
O'Doherty, Joseph E. ;
Lebedev, Mikhail A. ;
Li, Zheng ;
Nicolelis, Miguel A. L. .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2012, 20 (01) :85-93
[26]   Active tactile exploration using a brain-machine-brain interface [J].
O'Doherty, Joseph E. ;
Lebedev, Mikhail A. ;
Ifft, Peter J. ;
Zhuang, Katie Z. ;
Shokur, Solaiman ;
Bleuler, Hannes ;
Nicolelis, Miguel A. L. .
NATURE, 2011, 479 (7372) :228-U106
[27]   Voltage pulses change neural interface properties and improve unit recordings with chronically implanted microelectrodes [J].
Otto, KJ ;
Johnson, MD ;
Kipke, DR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (02) :333-340
[28]   The Long-Term Effects of Modified Electrode Surfaces and Intracochlear Corticosteroids on Postoperative Impedances in Cochlear Implant Patients [J].
Paasche, Gerrit ;
Tasche, Claudia ;
Stoever, Timo ;
Lesinski-Schiedat, Anke ;
Lenarz, Thomas .
OTOLOGY & NEUROTOLOGY, 2009, 30 (05) :592-598
[29]   The functional consequences of chronic, physiologically effective intracortical microstimulation [J].
Parker, Rebecca A. ;
Davis, Tyler S. ;
House, Paul A. ;
Normann, Richard A. ;
Greger, Bradley .
BRAIN MACHINE INTERFACES: IMPLICATIONS FOR SCIENCE, CLINICAL PRACTICE AND SOCIETY, 2011, 194 :145-165
[30]  
Rousche P J, 1999, IEEE Trans Rehabil Eng, V7, P56, DOI 10.1109/86.750552