Reproducibility of product ion spectra acquired using a liquid chromatography/triple-quadrupole mass spectrometry (LC/MS/MS) instrument over a 4-year period, and with three other LC/MS/MS instruments, one from the same manufacturer and two from a different manufacturer, was examined. The MS/MS spectra of 30 drug substances were generated in positive electrospray ionization mode at low, medium, and high collision energies (20, 35, and 50 eV). Purity and Fit score percentages against a 400-compound LC/MS/MS spectral library were calculated using an algorithm in which fragment intensity ratios and weighting factors were included. The long-term reproducibility study was conducted using a brand A instrument; after 4 years the reproducibility of the product ion spectra was still 94%, expressed as average Purity score. The inter-laboratory study involved two parts. Firstly, two LC/MS/MS spectral libraries, created independently in separate laboratories using brand A instruments, were compared with each other. The average Fit and Purity scores of spectra from one library against the other were better than 93 and 91%. respectively, when the same collision energies were used. Secondly, for the comparison of product ion spectra between brand A and brand B instruments, fragmentation conditions were first standardized for amitriptyline as the standard analyte. The average Fit scores of brand B spectra against the brand A spectral library varied between 79 and 85% at all three collision energies. These results indicate that, after standardizing the instrumental conditions, LC/MS/MS spectral libraries of drug substances are suitable for inter-laboratory use. Copyright (C) 2004 John Wiley Sons, Ltd.
引用
收藏
页码:1039 / 1046
页数:8
相关论文
共 16 条
[1]   Poor reproducibility of in-source collisional atmospheric pressure ionization mass spectra of toxicologically relevant drugs [J].
Bogusz, MJ ;
Maier, RD ;
Krüger, KD ;
Webb, KS ;
Romeril, J ;
Miller, ML .
JOURNAL OF CHROMATOGRAPHY A, 1999, 844 (1-2) :409-418
[2]   Evaluation of protocols for reproducible electrospray in-source collisionally induced dissociation on various liquid chromatography/mass spectrometry instruments and the development of spectral libraries [J].
Bristow, AWT ;
Nichols, WF ;
Webb, KS ;
Conway, B .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (24) :2374-2386
[3]   AN APPROXIMATE METHOD FOR DETERMINING THE EMITTANCE OF ION SOURCES IN QUADRUPOLE MASS-SPECTROMETRY [J].
DAWSON, PH ;
BINGQI, Y .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1983, 54 (1-2) :159-167
[4]  
Gergov M, 2000, J MASS SPECTROM, V35, P912, DOI 10.1002/1096-9888(200007)35:7<912::AID-JMS19>3.0.CO
[5]  
2-4
[6]   Evaluation of electrospray transport CID for the generation of searchable libraries [J].
Hough, JM ;
Haney, CA ;
Voyksner, RD ;
Bereman, RD .
ANALYTICAL CHEMISTRY, 2000, 72 (10) :2265-2270
[7]   A mass spectral library based on chemical ionization and collision-induced dissociation [J].
Kienhuis, PGM ;
Geerdink, RB .
JOURNAL OF CHROMATOGRAPHY A, 2002, 974 (1-2) :161-168
[8]   Methodology for the development of a drug library based upon collision-induced fragmentation for the identification of toxicologically relevant drugs in plasma samples [J].
Lips, AGAM ;
Lameijer, W ;
Fokkens, RH ;
Nibbering, NMM .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2001, 759 (02) :191-207
[9]   In-source CID mass spectral libraries for the "general unknown" screening of drugs and toxicants [J].
Marquet, P ;
Venisse, N ;
Lacassie, É ;
Lachâtre, G .
ANALUSIS, 2000, 28 (10) :925-934A
[10]   STANDARD REFERENCE SPECTRA FOR TANDEM MASS-SPECTROMETRY QUALITY ASSURANCE, PERFORMANCE EVALUATION AND PROFICIENCY TESTING - TANDEM MASS SPECTROMETERS WHICH USE RF-ONLY MULTIPOLE COLLISION CELLS [J].
MARTINEZ, RI .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1991, 5 (05) :245-248