Patient-prosthesis mismatch is negligible with modern small-size aortic valve prostheses

被引:32
作者
Izzat, MB
Kadir, I
Reeves, B
Wilde, P
Bryan, AJ
Angelini, GD
机构
[1] Univ Bristol, Bristol Heart Inst, Bristol, Avon, England
[2] Univ Bristol, Res & Dev Support Unit, Bristol, Avon, England
关键词
D O I
10.1016/S0003-4975(99)00717-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Concern has been raised about residual significant gradients when small aortic prostheses are used, particularly in patients with large body surface areas. We studied the performance of six types of small aortic prostheses using dobutamine stress echocardiography. Methods. Sixty-three patients (mean age, 67 +/- 7 years) who had undergone aortic valve replacement 17 +/- 6 months previously were studied. Two bileaflet mechanical prostheses (St. Jude Medical and CarboMedics: sizes, 19 mm and 21 mm) and two biological prostheses (Medtronic Intact and St. Jude BioImplant: size, 21 mm) were evaluated. A graded infusion of dobutamine was given and Doppler studies of valve performance were carried out. Results. All prostheses except one biological valve had acceptable hemodynamic performance under stress. Using regression modeling, gradient at rest was the only variable found to predict gradient under stress (p < 0.001). Moreover, the most important predictor of gradient at rest was valve design, which accounted for 72% of the variance (p < 0.001). This relationship was independent of valve size (19 mm or 21 mm) or material tie, mechanical or biological). Body surface area accounted for 4% of the variance in gradient only. Conclusions. The main predictor of transprosthetic gradient is the inherent characteristics of each particular prosthesis, with relatively insignificant contribution from variations in body surface area. Patient-prosthesis mismatch is not a problem of clinical significance when certain modern valve prostheses are used. (C) 1999 by The Society of Thoracic Surgeons.
引用
收藏
页码:1657 / 1660
页数:4
相关论文
共 19 条
[11]  
Kadir I, 1997, J HEART VALVE DIS, V6, P123
[12]   Hemodynamic performance of the 21-mm St. Jude Bioimplant prosthesis using Dobutamine Doppler echocardiography [J].
Kadir, I ;
Izzat, MB ;
Birdi, I ;
Wilde, P ;
Reeves, B ;
Walsh, C ;
Bryan, A ;
Angelini, G .
AMERICAN JOURNAL OF CARDIOLOGY, 1998, 81 (05) :599-603
[13]  
Kadir I, 1997, ANN THORAC SURG, V63, P1128
[14]   THE RISK OF SMALL ST-JUDE AORTIC-VALVE PROSTHESES [J].
KRATZ, JM ;
SADE, RM ;
CRAWFORD, FA ;
CRUMBLEY, AJ ;
STROUD, MR .
ANNALS OF THORACIC SURGERY, 1994, 57 (05) :1114-1119
[15]   Reversibility of cardiac dysfunction after valve replacement in elderly patients with severe aortic stenosis [J].
Natsuaki, M ;
Itoh, T ;
Tomita, S ;
Naito, K .
ANNALS OF THORACIC SURGERY, 1998, 65 (06) :1634-1638
[16]  
Pibarot P, 1998, J HEART VALVE DIS, V7, P211
[17]   PROBLEM OF VALVE PROSTHESIS PATIENT MISMATCH [J].
RAHIMTOOLA, SH .
CIRCULATION, 1978, 58 (01) :20-24
[18]   St. Jude medical cardiac valves in small aortic roots: Follow-up to sixteen years [J].
Sawant, D ;
Singh, AK ;
Feng, WC ;
Bert, AA ;
Rotenberg, F .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1997, 113 (03) :499-509
[19]   CLINICAL AND HEMODYNAMIC EVALUATION OF THE 19 MM BJORK-SHILEY AORTIC-VALVE PROSTHESIS [J].
SCHAFF, HV ;
BORKON, AM ;
HUGHES, C ;
ACHUFF, S ;
DONAHOO, JS ;
GARDNER, TJ ;
WATKINS, L ;
GOTT, VL ;
MORROW, AG ;
BRAWLEY, RK .
ANNALS OF THORACIC SURGERY, 1981, 32 (01) :50-57