Online standard additions calibration of transient signals for inductively coupled plasma mass spectrometry

被引:19
作者
Antler, Margaret
Maxwell, E. Jane
Duford, David A.
Salin, E. D.
机构
[1] McGill Univ, Dept Chem, Montreal, PQ H3A 2K6, Canada
[2] Univ British Columbia, Dept Chem, Vancouver, BC V6T 1Z3, Canada
关键词
D O I
10.1021/ac061616v
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An online standard additions calibration method for transient signals in ICPMS is demonstrated in which a small volume of standard is injected as a spike into the sample/carrier stream, overlaying the analyte peak. This technique provides the advantages of conventional standard additions but requires only a single sample run. The method corrects for matrix effects and is suitable for transient signals in which the severity of the matrix effect changes over the analyte peak. The method uses a peak-fitting program to determine the area of the underlying peak and is shown to be effective for the determination of trace metal concentrations in both a high ionic strength matrix and in a biological matrix (urine). Eight analytes with concentrations in the range of 0.82-233.2 mu g L-1 in urine were simultaneously determined using a standard spiking solution of 75 mu g L-1 injected through a 100-mu L loop. The measured concentrations for analytes free of spectral interferences agreed with the certified values, and the precision achieved was comparable to that achieved by the certifying agency. Using a conventional cross-flow nebulizer and Scott-type spray chamber, the accuracy obtained for online standard additions calibration was within 2%, and the precision was within 5%.
引用
收藏
页码:688 / 694
页数:7
相关论文
共 37 条
[11]   INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY - A REVIEW [J].
EVANS, EH ;
GIGLIO, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) :1-18
[12]   Metallomics as integrated biometal science [J].
Haraguchi, H .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2004, 19 (01) :5-14
[13]   Isotope-dilution ICP-MS for trace element determination and speciation: from a reference method to a routine method? [J].
Heumann, KG .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 378 (02) :318-329
[14]   ELEMENTAL SPECIATION WITH LIQUID-CHROMATOGRAPHY INDUCTIVELY-COUPLED PLASMA ISOTOPE-DILUTION MASS-SPECTROMETRY [J].
HEUMANN, KG ;
ROTTMANN, L ;
VOGL, J .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (12) :1351-1355
[15]   Speciation of alkylated metals and metalloids in the environment [J].
Hirner, Alfred V. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 385 (03) :555-567
[16]   Direct multielemental analysis of human serum by ICP-MS with on-line standard addition using flow injection [J].
Huang, CY ;
Beauchemin, D .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (08) :951-952
[17]   Tandem calibration methodology: dual nebulizer sample introduction for ICP-MS [J].
Huxter, V ;
Hamier, J ;
Salin, ED .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (01) :71-75
[18]   FLOW-INJECTION SAMPLE-TO-STANDARD ADDITIONS METHOD - SPECTROPHOTOMETRIC DETERMINATION OF HYDROCHLORIC-ACID AND ORTHO-PHOSPHATE [J].
ISRAEL, Y ;
BARNES, RM .
ANALYST, 1989, 114 (07) :843-848
[19]   CONCEPTS FOR THE IDENTIFICATION AND DETERMINATION OF SELENIUM-COMPOUNDS IN THE AQUATIC ENVIRONMENT [J].
KOLBL, G .
MARINE CHEMISTRY, 1995, 48 (3-4) :185-197
[20]   CARBON-ENHANCED INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRIC DETECTION OF ARSENIC AND SELENIUM AND ITS APPLICATION TO ARSENIC SPECIATION [J].
LARSEN, EH ;
STURUP, S .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (10) :1099-1105