Magnetic resonance-guided placement of atrial septal closure device in animal model of patent foramen ovale

被引:56
作者
Buecker, A
Spuentrup, E
Grabitz, R
Freudenthal, F
Muehler, EG
Schaeffter, T
van Vaals, JJ
Günther, RW
机构
[1] Univ Technol Aachen, Clin Diagnost Radiol, D-52057 Aachen, Germany
[2] Univ Technol Aachen, Clin Pediat Cardiol, D-52057 Aachen, Germany
[3] Univ Kiel, Dept Pediat Cardiol, Kiel, Germany
[4] Philips Res Labs, Hamburg, Germany
[5] Philips Med Syst, Best, Netherlands
关键词
magnetic resonance imaging; catheterization; defects;
D O I
10.1161/01.CIR.0000023621.88708.62
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Percutaneous closure of the patent foramen ovale (PFO) is usually performed under x-ray in combination with ultrasound guidance, We tested the feasibility of applying magnetic resonance (MR) guidance for percutaneous closure of PFO in an animal model, thus avoiding the disadvantage of ionizing radiation. Methods and Results-Real-time MRI with radial or spiral k-space filling (115 frames per second) on an interventional 1.5-T high-field whole-body system was exploited to examine the feasibility of MR-guided closure of the PFO in 7 piglets weighing approximate to 14 kg. A specially designed prototype nonmagnetic closure device was introduced via the femoral vein. The short bore of the magnet and in-room monitors allowed for visualization and steering of the catheter with the loaded occluder. Catheterization of the left atrium and, finally, correct placement of the device was possible in all animals. Deployment of the device was depicted by real-time MR, and initial misplacement, which occurred in 2 animals, was easily detected and corrected. Conclusions-Real-time MR guidance of PFO closure, without the use of ionizing radiation, is feasible in an animal model.
引用
收藏
页码:511 / 515
页数:5
相关论文
共 18 条
[11]  
Konings MK, 2000, J MAGN RESON IMAGING, V12, P79, DOI 10.1002/1522-2586(200007)12:1<79::AID-JMRI9>3.3.CO
[12]  
2-K
[13]   Visualization and temporal/spatial characterization of cardiac radiofrequency ablation lesions using magnetic resonance imaging [J].
Lardo, AC ;
McVeigh, ER ;
Jumrussirikul, P ;
Berger, RD ;
Calkins, H ;
Lima, J ;
Halperin, HR .
CIRCULATION, 2000, 102 (06) :698-705
[14]   MR-guided angioplasty of renal artery stenosis in a pig model: A feasibility study [J].
Omary, RA ;
Frayne, R ;
Unal, O ;
Warner, T ;
Korosec, FR ;
Mistretta, CA ;
Strother, CM ;
Grist, TM .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (03) :373-381
[15]  
RASCHE V, 1995, MAGNET RESON MED, V34, P754, DOI 10.1002/mrm.1910340515
[16]   MR FLUOROSCOPY - TECHNICAL FEASIBILITY [J].
RIEDERER, SJ ;
TASCIYAN, T ;
FARZANEH, F ;
LEE, JN ;
WRIGHT, RC ;
HERFKENS, RJ .
MAGNETIC RESONANCE IN MEDICINE, 1988, 8 (01) :1-15
[17]   Magnetic resonance-guided coronary artery stent placement in a swine model [J].
Spuentrup, E ;
Ruebben, A ;
Schaeffter, T ;
Manning, WJ ;
Günther, RW ;
Buecker, A .
CIRCULATION, 2002, 105 (07) :874-879
[18]  
Tseng HC, 2000, J FORMOS MED ASSOC, V99, P684