A mechanical comparison of 2 posterior-stabilizing designs - Insall/Burstein 2 knee and bisurface knee

被引:18
作者
Akagi, M [1 ]
Ueo, T [1 ]
Takagi, H [1 ]
Hamanishi, C [1 ]
Nakamura, T [1 ]
机构
[1] Kinki Univ, Sch Med, Dept Orthopaed Surg, Osakasayama, Osaka 5898511, Japan
关键词
mechanical comparison; posterior-stabilizing knee; Bisurface knee; Insall-Burstein 2 (IB2) knee;
D O I
10.1054/arth.2002.32694
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A posterior-stabilizing knee prosthesis, called the Bisurface knee (Kyocera Corp, Kyoto, Japan), with a ball-and-socket joint installed in the midposterior portion of the tibiofemoral joint, has been developed to satisfy 2 conflicting demands in knee joint design: kinematics and wear resistance. To confirm if the prosthesis has achieved its design objectives, a contact area and stress study was done throughout the range of motion and compared with results obtained for the Insall-Burstein 2 knee. The posterior-stabilizing ability of the ball-and-socket joint also was assessed. This study showed that the ball-and-socket joint could provide sufficient posterior stability, earlier start of flexion, and lower contact stress in the tibia] polyethylene insert in flexion. The design of the Bisurface knee could provide a good balance between kinematics, stability, and wear resistance.
引用
收藏
页码:627 / 634
页数:8
相关论文
共 25 条
[1]
The bisurface total knee replacement: A unique design for flexion - Four-to-nine-year follow-up study [J].
Akagi, M ;
Nakamura, T ;
Matsusue, Y ;
Ueo, T ;
Nishijyo, K ;
Ohnishi, E .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (11) :1626-1633
[2]
THE INFLUENCE OF TOTAL KNEE-REPLACEMENT DESIGN ON WALKING AND STAIR-CLIMBING [J].
ANDRIACCHI, TP ;
GALANTE, JO ;
FERMIER, RW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1982, 64 (09) :1328-1335
[3]
THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[4]
Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components [J].
Bell, CJ ;
Walker, PS ;
Abeysundera, MR ;
Simmons, JMH ;
King, PM ;
Blunn, GW .
JOURNAL OF ARTHROPLASTY, 1998, 13 (03) :280-290
[5]
POLYETHYLENE WEAR IN UNICONDYLAR KNEE PROSTHESES - 106 RETRIEVED MARMOR, PCA, AND ST-GEORG TIBIAL COMPONENTS COMPARED [J].
BLUNN, GW ;
JOSHI, AB ;
LILLEY, PA ;
ENGELBRECHT, E ;
RYD, L ;
LIDGREN, L ;
HARDINGE, K ;
NIEDER, E ;
WALKER, PS .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (03) :247-255
[6]
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P232
[7]
Dennis DA, 1996, CLIN ORTHOP RELAT R, P107
[8]
DRAGANICH LF, 1987, J ORTHOPAED RES, V5, P538
[9]
THE INFLUENCE OF TIBIAL-PATELLOFEMORAL LOCATION ON FUNCTION OF THE KNEE IN PATIENTS WITH THE POSTERIOR STABILIZED CONDYLAR KNEE PROSTHESIS [J].
FIGGIE, HE ;
GOLDBERG, VM ;
HEIPLE, KG ;
MOLLER, HS ;
GORDON, NH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1035-1040
[10]