Measurement precision of body composition variables using the Lunar DPX-L densitometer

被引:137
作者
Kiebzak, GM
Leamy, LJ
Pierson, LM
Nord, RH
Zhang, ZY
机构
[1] Miller Orthopaed Clin, Res Serv, Charlotte, NC 28203 USA
[2] Univ N Carolina, Dept Biol, Charlotte, NC 28223 USA
[3] Carolinas Med Ctr, Charlotte, NC 28203 USA
[4] Lunar Corp, Madison, WI USA
[5] Univ N Carolina, Dept Math, Charlotte, NC 28223 USA
关键词
body composition; mode of analysis; precision; densitometry;
D O I
10.1385/JCD:3:1:035
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The objective of the study was to determine the precision of total and regional body composition measurements from a total-body scan using dual-energy X-ray absorptiometry (DXA). This is critical information necessary to determine the smallest change from baseline that could be detected with statistical significance when conducting longitudinal measurements of body composition variables in an individual. Twenty volunteers were scanned once each day for 4 consecutive days using a Lunar DPX-L densitometer and manufacturer-supplied software (version 1.3z). Coefficients of variation (CV, %) derived from data using the (preferred) extended research mode of analysis were 0.62, 1.89, 0.63, 2.0, 1.11, 1.10, and 1.09% for total-body bone mineral density (BMD), total percentage fat, total body tissue mass, fat mass, lean mass, bone mineral content (BMC), and total bone calcium, respectively. Regional measurements (arm, leg, trunk, pelvis, and spine) were less precise than total-body measurements, with CVs in the range of 1% to 3% (but fat mass for arms was 4.26%, trunk 3.08%, BMC 3.65%). Small but statistically significant differences in mean values for most body composition variables were found when data were compared between extended and standard modes of analysis. Inconsistent use of analysis mode in a cohort or when following a patient longitudinally may negatively affect precision. We conclude that the measurement precision of total- and regional-body composition variables was generally comparable to the precision limits typically associated with lumbar spine and proximal femur BMD data.
引用
收藏
页码:35 / 41
页数:7
相关论文
共 26 条
[1]  
[Anonymous], AM J PHYSIOL
[2]  
BONNICK SL, 1998, BONE DENSITOMETRY CL
[3]   Precision of body composition measurements by dual energy X-ray absorptiometry [J].
Braillon, P ;
Lapillonne, A ;
Giraud, S ;
Salle, BL ;
Cochat, P .
APPLIED RADIATION AND ISOTOPES, 1998, 49 (5-6) :501-502
[4]   Multicentre evaluation of KONE Optima analysis system [J].
Ferrero, C ;
Modenese, A ;
Bonini, PA ;
Ceriotti, F ;
Carobene, A ;
Franzin, M ;
Guder, WG ;
Hofmann, W ;
Puukka, M ;
Laitinen, H ;
Makarovsky, V ;
Gaidukova, N .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 1998, 36 (07) :475-484
[5]   Comparison of body composition methods: A literature analysis [J].
Fogelholm, M ;
Lichtenbelt, WV .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 1997, 51 (08) :495-503
[6]   Evaluation of i-STAT portable clinical analyzer in a hemodialysis unit [J].
Gault, MH ;
Harding, CE .
CLINICAL BIOCHEMISTRY, 1996, 29 (02) :117-124
[7]   ACCURATE ASSESSMENT OF PRECISION ERRORS - HOW TO MEASURE THE REPRODUCIBILITY OF BONE DENSITOMETRY TECHNIQUES [J].
GLUER, CC ;
BLAKE, G ;
LU, Y ;
BLUNT, BA ;
JERGAS, M ;
GENANT, HK .
OSTEOPOROSIS INTERNATIONAL, 1995, 5 (04) :262-270
[8]   VALIDATION OF BODY-COMPOSITION BY DUAL ENERGY X-RAY ABSORPTIOMETRY (DEXA) [J].
HAARBO, J ;
GOTFREDSEN, A ;
HASSAGER, C ;
CHRISTIANSEN, C .
CLINICAL PHYSIOLOGY, 1991, 11 (04) :331-341
[9]   A COMPARISON OF 3 DESK-TOP CHEMISTRY ANALYZERS FOR THE PEDIATRIC PRACTICE [J].
HICKS, JM ;
IOSEFSOHN, M ;
GREENBERG, L .
CLINICAL PEDIATRICS, 1991, 30 (04) :217-224
[10]   PRECISION AND STABILITY OF DUAL-ENERGY X-RAY ABSORPTIOMETRY MEASUREMENTS [J].
JOHNSON, J ;
DAWSONHUGHES, B .
CALCIFIED TISSUE INTERNATIONAL, 1991, 49 (03) :174-178