Measurement of the Detective Quantum Efficiency (DQE) of digital X-ray imaging devices according to the standard IEC 62220-1

被引:20
作者
Illers, H [1 ]
Buhr, E [1 ]
Bergmann, D [1 ]
Hoeschen, C [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
来源
MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2004年 / 5368卷
关键词
digital radiography; CR radiography; flat-panel detector; detective quantum efficiency; DQE; IEC; 62220-1; standardization;
D O I
10.1117/12.535999
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The DQEs of four digital X-ray detector systems have been measured in accordance with the new international standard IEC 62220-1: two CR detector systems of the same type, a CsI-based indirect flat panel detector and a selenium-based direct flat panel detector. A mobile measurement set-up complying with IEC 62220-1 has been realized. All equipment used was of a specific design, tested and calibrated. A standardized radiation quality (RQA5) was applied, and the air kerma at the detector entrance was varied between about 1 muGy and 20 muGy. The measurements of the two CR detector systems were performed at different sites using different X-ray generators/tubes and were in agreement within 0.02. The maximum DQE values were obtained for the lowest spatial frequency for which the DQE is required to be reported according to the IEC standard, i.e. at 0.5 mm(-1): The maximum DQE value measured was 0.21 for the CR systems, 0.42 for the indirect flat panel detector, and 0.31 for the direct Selenium-based detector. It has been demonstrated that the international standard IEC 62220-1 allows accurate and reliable measurements of the DQE to be conducted. It is now possible to objectively measure and compare DQE values of digital X-ray detector systems.
引用
收藏
页码:177 / 187
页数:11
相关论文
共 15 条
[1]  
[Anonymous], 1994, 61267 IEC
[2]   Evaluation of the imaging properties of two generations of a CCD-based system for digital chest radiography [J].
Båth, M ;
Sund, P ;
Månsson, LG .
MEDICAL PHYSICS, 2002, 29 (10) :2286-2297
[3]  
*BIPM IEC IFCC ISO, 1995, GUID EXPR UNC MEAS
[4]   Simple method for modulation transfer function determination of digital imaging detectors from edge images [J].
Buhr, E ;
Günther-Kohfahl, S ;
Neitzel, U .
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 :877-884
[5]   Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image [J].
Buhr, E ;
Günther-Kohfahl, S ;
Neitzel, U .
MEDICAL PHYSICS, 2003, 30 (09) :2323-2331
[6]  
Cunningham I. A., 2000, HDB MED IMAGING, V1, P79
[7]   DQE(F) OF 4 GENERATIONS OF COMPUTED RADIOGRAPHY ACQUISITION DEVICES [J].
DOBBINS, JT ;
ERGUN, DL ;
RUTZ, L ;
HINSHAW, DA ;
BLUME, H ;
CLARK, DC .
MEDICAL PHYSICS, 1995, 22 (10) :1581-1593
[8]   Performance of a 41X41-cm2 amorphous silicon flat panel x-ray detector for radiographic imaging applications [J].
Granfors, PR ;
Aufrichtig, R .
MEDICAL PHYSICS, 2000, 27 (06) :1324-1331
[9]   DQE of digital x-ray imaging systems: a challenge for standarization [J].
Hoeschen, D .
MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 :280-286
[10]  
IEC, 2003, 622201 IEC