Biomechanical comparison of effects of the Dynesys and Coflex dynamic stabilization systems on range of motion and loading characteristics in the lumbar spine: a finite element study

被引:42
作者
Kulduk, Ahmet [1 ]
Altun, Necdet S. [2 ]
Senkoylu, Alpaslan [2 ]
机构
[1] Gazi Mustafa Kemal State Hosp, Dept Orthoped & Traumatol, Ankara, Turkey
[2] Gazi Univ, Fac Med, Dept Orthoped & Traumatol, TR-06510 Ankara, Turkey
关键词
Finite element modeling; dynamic stabilization; spinal column; Coflex; Dynesis; interspinous devices; DEVICES; FUSION;
D O I
10.1002/rcs.1636
中图分类号
R61 [外科手术学];
学科分类号
100210 [外科学];
摘要
Background The primary purpose of dynamic stabilization is to preserve the normal range of motion (ROM) by restricting abnormal movement in the spine. Our aim was to analyze the effects of two different dynamic stabilization systems using finite element modeling (FEM). Methods Coflex and Dynesys dynamic devices were modeled and implanted at the L4-L5 segment using virtual FEM. A 400N compressive force combined with 6N flexion, extension, bending and axial rotation forces was applied to the L3-4 and L4-5 segments. ROM and disc loading forces were analyzed. Results Both systems reduced ROM and disc loading forces at the implanted lumbar segment, with the exception of the Coflex interspinous device, which increased ROM by 19% and did not change disc-loading forces in flexion. Conclusions The Coflex device prevented excessive disc loading, but increased ROM abnormally in flexion. Neither device provided satisfactory motion preservation or load sharing in other directions. Copyright (C) 2015 John Wiley & Sons, Ltd.
引用
收藏
页码:400 / 405
页数:6
相关论文
共 24 条
[1]
Interspinous process devices in the lumbar spine [J].
Bono, Christopher M. ;
Vaccaro, Alexander R. .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2007, 20 (03) :255-261
[2]
Critical analysis of trends in fusion for degenerative disc disease over the past 20 years - Influence of technique on fusion rate and clinical outcome [J].
Bono, CM ;
Lee, CK .
SPINE, 2004, 29 (04) :455-463
[3]
Cruz Mauro, 2006, Technol Health Care, V14, P421
[4]
The relationship between disc degeneration, facet joint osteoarthritis, and stability of the degenerative lumbar spine [J].
Fujiwara, A ;
Tamai, K ;
An, HS ;
Kurihashi, A ;
Lim, TH ;
Yoshida, H ;
Saotome, K .
JOURNAL OF SPINAL DISORDERS, 2000, 13 (05) :444-450
[5]
Rigid and flexible spinal stabilization devices: A biomechanical comparison [J].
Galbusera, Fabio ;
Bellini, Chiara Maria ;
Anasetti, Federica ;
Ciavarro, Cristina ;
Lovi, Alessio ;
Brayda-Bruno, Marco .
MEDICAL ENGINEERING & PHYSICS, 2011, 33 (04) :490-496
[6]
Clinical experience with the Dynesys semirigid fixation system for the lumbar spine - Surgical and patient-oriented outcome in 50 cases after an average of 2 years [J].
Grob, D ;
Benini, A ;
Junge, A ;
Mannion, AF .
SPINE, 2005, 30 (03) :324-331
[7]
Posterior motion preserving implants evaluated by means of intervertebral disc bulging and annular fiber strains [J].
Heuer, Frank ;
Schmidt, Hendrik ;
Kaefer, Wolfram ;
Graf, Nicolas ;
Wilke, Hans-Joachim .
CLINICAL BIOMECHANICS, 2012, 27 (03) :218-225
[8]
Advantages and disadvantages of nonfusion technology in spine surgery [J].
Huang, RC ;
Girardi, FP ;
Lim, MR ;
Cammisa, FP .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 2005, 36 (03) :263-+
[9]
One-year outcome evaluation after interspinous implantation for degenerative spinal stenosis with segmental instability [J].
Kong, Doo-Sik ;
Kim, Eun-Sang ;
Eoh, Whan .
JOURNAL OF KOREAN MEDICAL SCIENCE, 2007, 22 (02) :330-335
[10]
Effect of the cord pretension of the Dynesys dynamic stabilisation system on the biomechanics of the lumbar spine: a finite element analysis [J].
Liu, Chien-Lin ;
Zhong, Zheng-Cheng ;
Hsu, Hung-Wei ;
Shih, Shih-Liang ;
Wang, Shih-Tien ;
Hung, Chinghua ;
Chen, Chen-Sheng .
EUROPEAN SPINE JOURNAL, 2011, 20 (11) :1850-1858