High-resolution x-ray tomography using laboratory sources

被引:17
作者
Tkachuk, Andrei [1 ]
Feser, Michael [1 ]
Cui, Hongtao [1 ]
Duewer, Fred [1 ]
Chang, Hauyee [1 ]
Yuri, Wenbing [1 ]
机构
[1] Xradia Inc, 5052 Commercial Circle, Concord, CA 94597 USA
来源
DEVELOPMENTS IN X-RAY TOMOGRAPHY V | 2006年 / 6318卷
关键词
x-ray microscopy; computed tomography; Fresnel zone plate;
D O I
10.1117/12.682383
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
X-ray computed tomography (XCT) is a powerful nondestructive 3D imaging technique, which enables the visualization of the three dimensional structure of complex, optically opaque samples. High resolution XCT using Fresnel zone plate lenses has been confined in the past to synchrotron radiation centers due to the need for a bright and intense source of x-rays. This confinement severely limits the availability and accessibility of x-ray microscopes and the wide proliferation of this methodology. We are describing a sub-50nm resolution XCT system operating at 8 keV in absorption and Zernike phase contrast mode based on a commercially available laboratory x-ray source. The system utilizes high-efficiency Fresnel zone plates with an outermost zone width of 35 nm and 700 nm structure height resulting in a current spatial resolution better than 50 nm. In addition to the technical description of the system and specifications, we present application examples in the semiconductor field.
引用
收藏
页数:8
相关论文
共 11 条
[1]  
BORN M, 1999, PRINCIPLES OPTICS, P217
[2]   Soft X-ray microscopy at a spatial resolution better than 15nm [J].
Chao, WL ;
Harteneck, BD ;
Liddle, JA ;
Anderson, EH ;
Attwood, DT .
NATURE, 2005, 435 (7046) :1210-1213
[3]   Design and performance of a laser-plasma-based compact soft x-ray microscope [J].
Johansson, GA ;
Holmberg, A ;
Hertz, HM ;
Berglund, M .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (03) :1193-1197
[4]   Compact soft x-ray transmission microscopy with sub-50 nm spatial resolution [J].
Kim, KW ;
Kwon, Y ;
Nam, KY ;
Lim, JH ;
Kim, KG ;
Chon, KS ;
Kim, BH ;
Kim, DE ;
Kim, J ;
Ahn, BN ;
Shin, HJ ;
Rah, S ;
Kim, KH ;
Chae, JS ;
Gweon, DG ;
Kang, DW ;
Kang, SH ;
Min, JY ;
Choi, KS ;
Yoon, SE ;
Kim, EN ;
Namba, Y ;
Yoon, KH .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (06) :N99-N107
[5]   X-ray microscopy in Zernike phase contrast mode at 4 keV photon energy with 60 nm resolution [J].
Neuhäusler, U ;
Schneider, G ;
Ludwig, W ;
Meyer, MA ;
Zschech, E ;
Hambach, D .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2003, 36 (10A) :A79-A82
[6]  
RUDOLPH D, 1990, AMPLITUDE PHASE CONT, P59
[7]   Cryo X-ray microscopy with high spatial resolution in amplitude and phase contrast [J].
Schneider, G .
ULTRAMICROSCOPY, 1998, 75 (02) :85-104
[8]   Hard X-ray microscopy activities at SPring-8 [J].
Suzuki, Y ;
Awaji, M ;
Takeuchi, A ;
Takano, H ;
Uesugi, K ;
Kohmura, Y ;
Kamijo, N ;
Yasumoto, M ;
Tamura, S .
JOURNAL DE PHYSIQUE IV, 2003, 104 :35-40
[9]  
WANG S, 2002, P 28 INT S TEST FAIL, P227
[10]   ZONE PLATES AND THEIR ABERRATIONS [J].
YOUNG, M .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1972, 62 (08) :972-&