High-resolution multistation peripheral MR angiography using undersampled projection reconstruction imaging

被引:8
作者
Du, J
Korosec, FR
Thornton, FJ
Grist, TM
Mistretta, CA
机构
[1] Univ Wisconsin, Dept Radiol, Clin Sci Ctr E3 311, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
关键词
undersampled projection reconstruction; time-resolved; coil sensitivity; peripheral MRA; contrast-enhanced MRA;
D O I
10.1002/mrm.20102
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A novel protocol for three-station MR angiography (MRA) of the lower extremities is described. A time-resolved undersampled projection reconstruction (PR) acquisition was used to image the calf station during a first injection, and non-time-resolved PR acquisitions were used with the bolus-chase technique to image the abdomen and thigh stations during a second injection. The streak artifacts resulting from undersampling the PR data were reduced with the use of a spatial Fermi filter based on the sensitivity of each coil element in a peripheral vascular phased-array coil. This novel technique provided high spatial resolution and a broad range of coverage, and depicted the contrast dynamics in the most distal station of the lower extremities.
引用
收藏
页码:204 / 208
页数:5
相关论文
共 24 条
[1]   SNR improvement for multiinjection time-resolved high-resolution CE-MRA of the peripheral vasculature [J].
Du, J ;
Carroll, TJ ;
Block, WF ;
Fain, SB ;
Korosec, FR ;
Grist, TM ;
Mistretta, CA .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (05) :909-917
[2]   Time-resolved, undersampled projection reconstruction imaging for high-resolution CE-MRA of the distal runoff vessels [J].
Du, J ;
Carroll, TJ ;
Wagner, HJ ;
Vigen, K ;
Fain, SB ;
Block, WF ;
Korosec, FR ;
Grist, TM ;
Mistretta, CA .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (03) :516-522
[3]  
DU J, 2003, P 11 ANN M ISMRM TOR, P255
[4]   High-spatial-resolution multistation MR imaging of lower-extremity peripheral vasculature with segmented volume acquisition: Feasibility study [J].
Foo, TKF ;
Ho, VB ;
Hood, MN ;
Marcos, HB ;
Hess, SL ;
Choyke, PL .
RADIOLOGY, 2001, 219 (03) :835-841
[5]   Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA) [J].
Griswold, MA ;
Jakob, PM ;
Heidemann, RM ;
Nittka, M ;
Jellus, V ;
Wang, JM ;
Kiefer, B ;
Haase, A .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) :1202-1210
[6]  
Griswold MA, 2000, MAGN RESON MED, V44, P602, DOI 10.1002/1522-2594(200010)44:4<602::AID-MRM14>3.0.CO
[7]  
2-5
[8]   Optimization of gadolinium-enhanced magnetic resonance angiography using an automated bolus-detection algorithm (MR SmartPrep) - Original investigation [J].
Ho, VB ;
Foo, TKF .
INVESTIGATIVE RADIOLOGY, 1998, 33 (09) :515-523
[9]  
Jakob PM, 1998, MAGN RESON MATER PHY, V7, P42
[10]   Time-resolved contrast-enhanced 3D MR angiography [J].
Korosec, FR ;
Frayne, R ;
Grist, TM ;
Mistretta, CA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (03) :345-351