Delayed fracture of a ceramic insert with modern ceramic total hip replacement

被引:36
作者
Min, Byung-Woo
Song, Kwang-Soon
Kang, Chul-Hyung
Bae, Ki-Cheol
Won, Ye-Yeon
Lee, Kwon-Yong
机构
[1] Keimyung Univ, Sch Med, Dept Orthopaed Surg, Taegu, South Korea
[2] Ajou Univ Hosp, Dept Orthopaed Surg, Suwon, South Korea
[3] Sejong Univ, Dept Mech Engn, Seoul, South Korea
关键词
total hip arthroplasty; ceramic liner fracture; impingement;
D O I
10.1016/j.arth.2005.12.012
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a case of ceramic liner fracture in a 34-year-old man that occurred 1 year 9 months after total hip arthroplasty with a BiCONTACT (Aesculap, Tuttlingen, Germany) ceramic-on-ceramic prosthesis. The ceramic liner was found within the metal shell in 4 large pieces and multiple small fragments in the peripheral portions. its condition indicated that the possible mechanism of late ceramic liner chip fracture was repeated episodes of impingement between the prosthetic neck and the edge of the ceramic liner. Squatting, kneeling, and sitting cross-legged caused impingement. These positions are more common in Asian populations than in Western populations. After the joint was thoroughly irrigated, a new modular ceramic liner and a 28-mm ceramic head were implanted. (C) 2007 Elsevier Inc. All rights reserved.
引用
收藏
页码:136 / 139
页数:4
相关论文
共 15 条
[1]
Bizot P, 2000, CLIN ORTHOP RELAT R, P134
[2]
CLARKE IC, 1992, CLIN ORTHOP RELAT R, P19
[3]
New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty [J].
D'Antonio, J ;
Capello, W ;
Manley, M ;
Bierbaum, B .
JOURNAL OF ARTHROPLASTY, 2002, 17 (04) :390-397
[4]
Fritsch EW, 1996, CLIN ORTHOP RELAT R, P129
[5]
Garino JP, 2000, CLIN ORTHOP RELAT R, P41
[6]
Fractures of ceramic bearings - History and present status [J].
Hannouche, D ;
Nich, C ;
Bizot, P ;
Meunier, A ;
Nizard, RM ;
Sedel, L .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :19-26
[7]
Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup [J].
Hasegawa, M ;
Sudo, A ;
Hirata, H ;
Uchida, A .
JOURNAL OF ARTHROPLASTY, 2003, 18 (05) :658-661
[8]
Heros R., 1998, SEMIN ARTHROPLASTY, V9, P114
[9]
MASSIVE WEAR OF A STEEL BALL HEAD BY CERAMIC FRAGMENTS IN THE POLYETHYLENE ACETABULAR CUP AFTER REVISION OF A TOTAL HIP-PROSTHESIS WITH FRACTURED CERAMIC BALL [J].
KEMPF, I ;
SEMLITSCH, M .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1990, 109 (05) :284-287
[10]
Mechanical performance of ceramic acetabular liners under impact conditions [J].
Maher, SA ;
Lipman, JD ;
Curley, LJ ;
Gilchrist, M ;
Wright, TM .
JOURNAL OF ARTHROPLASTY, 2003, 18 (07) :936-941