X-ray spectrum optimization of full-field digital mammography: Simulation and phantom study

被引:78
作者
Bernhardt, Philipp [1 ]
Mertelmeier, Thomas [1 ]
Hoheisel, Martin [1 ]
机构
[1] Siemens AG, D-91050 Erlangen, Germany
关键词
full-field digital mammography; spectrum optimization; signal difference-to-noise ratio; average glandular dose; TO-NOISE RATIO; IMAGE QUALITY; COMBINATIONS; REDUCTION; CONTRAST; SCATTER; SHAPE;
D O I
10.1118/1.2351951
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In contrast to conventional analog screen-film mammography new flat detectors have a high dynamic range and a linear characteristic curve. Hence, the radiographic technique can be optimized independently of the receptor exposure. It can be exclusively focused on the improvement of the image quality and the reduction of the patient dose. In this paper we measure the image quality by a physical quantity, the signal difference-to-noise ratio (SDNR), and the patient risk by the average glandular dose (AGD). Using these quantities, we compare the following different setups through simulations and phantom studies regarding the detection of microcalcifications and tumors for different breast thicknesses and breast compositions: Monochromatic radiation, three different anode/filter combinations: Molybdenum/molybdenum (Mo/Mo), molybdenum/rhodium (Mo/Rh), and tungsten/rhodium (W/Rh), different filter thicknesses, use of anti-scatter grids, and different tube voltages. For a digital mammography system based on an amorphous selenium detector it turned out that, first, the W/Rh combination is the best choice for all detection tasks studied. Second, monochromatic radiation can further reduce the AGD by a factor of up to 2.3, maintaining the image quality in comparison with a real polychromatic spectrum of an x-ray tube. And, third, the use of an anti-scatter grid is only advantageous for breast thicknesses larger than approximately 5 cm. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:4337 / 4349
页数:13
相关论文
共 29 条
[1]  
BAETZ L, 2005, P 7 INT WORKSH DIG M, P23
[2]   Spatial frequency-dependent signal-to-noise ratio as a generalized measure of image quality [J].
Bernhardt, P ;
Bätz, L ;
Rührnschope, EP ;
Hoheisel, M .
MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 :407-418
[3]   Optimization of technique factors for a silicon diode array full-field digital mammography system and comparison to screen-film mammography with matched average glandular dose [J].
Berns, EA ;
Hendrick, RE ;
Cutter, GR .
MEDICAL PHYSICS, 2003, 30 (03) :334-340
[4]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[5]   Contrast visibility of simulated microcalcifications in full field mammography systems [J].
Carton, AK ;
Bosmans, H ;
Van Ongeval, C ;
Souverijns, G ;
Rogge, F ;
Marchal, G .
MEDICAL IMAGING 2003: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2003, 5034 :412-423
[6]   The effect of the antiscatter grid on full-field digital mammography phantom images [J].
Chakraborty, DP .
JOURNAL OF DIGITAL IMAGING, 1999, 12 (01) :12-22
[7]   Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study [J].
Dance, DR ;
Klang, AT ;
Sandborg, M ;
Skinner, CL ;
Smith, IAC ;
Carlsson, GA .
BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (874) :1056-1067
[8]   MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE [J].
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1211-1219
[9]   A MODEL FOR OPTIMIZATION OF SPECTRAL SHAPE IN DIGITAL MAMMOGRAPHY [J].
FAHRIG, R ;
YAFFE, MJ .
MEDICAL PHYSICS, 1994, 21 (09) :1463-1471
[10]   X-ray imaging with amorphous selenium: Optimal spectra for digital mammography [J].
Fahrig, R ;
Rowlands, JA ;
Yaffe, MJ .
MEDICAL PHYSICS, 1996, 23 (04) :557-567