Optimization of technique factors for a silicon diode array full-field digital mammography system and comparison to screen-film mammography with matched average glandular dose

被引:32
作者
Berns, EA [1 ]
Hendrick, RE
Cutter, GR
机构
[1] Northwestern Univ, Sch Med, Lynn Sage Comprehens Breast Ctr, Chicago, IL 60611 USA
[2] Univ Nevada, Ctr Res Design & Stat Methods, Reno, NV USA
关键词
D O I
10.1118/1.1544674
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Contrast-detail experiments were performed to optimize technique factors for the detection of low-contrast lesions using a silicon diode array full-field digital mammography (FFDM) system under the conditions of a matched average glandular dose (AGD) for different techniques. Optimization was performed for compressed breast thickness from 2 to 8 cm. FFDM results were compared to screen-film mammography (SFM) at each breast thickness. Four contrast-detail (CD) images were acquired on a SFM unit with optimal techniques at 2, 4, 6, and 8 cm breast thicknesses. The AGD for each breast thickness was calculated based on half-value layer (HVL) and entrance exposure measurements on the SFM unit. A computer algorithm was developed and used to determine FFDM beam current (mAs) that matched AGD between FFDM and SFM at each thickness, while varying target, filter, and peak kilovoltage (kVp) across the full range available for the FFDM unit. CD images were then acquired on FFDM for kVp values from 23-35 for a molybdenum-molybdenum (Mo-Mo), 23-40 for a molybdenum-rhodium (Mo-Rh), and 25-49 for a rhodium-rhodium (Rh-Rh) target-filter under the constraint of matching the AGD from screen-film for each breast thickness (2, 4, 6, and 8 cm). CD images were scored independently for SFM and each FFDM technique by six readers. CD scores were analyzed to assess trends as a function of target-filter and kVp and were compared to SFM at each breast thickness. For 2 cm thick breasts, optimal FFDM CD scores occurred at the lowest possible kVp setting for each target-filter, with significant decreases in FFDM CD scores as kVp was increased under the constraint of matched AGD. For 2 cm breasts, optimal FFDM CD scores were not significantly different from SFM CD scores. For 4-8 cm breasts, optimum FFDM CD scores were superior to SFM CD scores. For 4 cm breasts, FFDM CD scores decreased as kVp increased for each target-filter combination. For 6 cm breasts, CD scores decreased slightly as kVp increased for Mo-Mo, but did not change significantly as a function of kVp for either Mo-Rh or Rh-Rh. For 8 cm breasts, Rh/Rh FFDM CD scores were superior to other target-filter combinations and increased significantly as kVp increased. These results indicate that low-contrast lesion detection was optimized for FFDM by using a softer x-ray beam for thin breasts and a harder x-ray beam for thick breasts, when AGD was kept constant for a given breast thickness. Under this constraint, optimum low-contrast lesion detection with FFDM was superior to that for SFM for all but the thinnest breasts. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:334 / 340
页数:7
相关论文
共 17 条
[1]   Performance comparison of full-field digital mammography to screen-film mammography in clinical practice [J].
Berns, EA ;
Hendrick, RE ;
Cutter, GR .
MEDICAL PHYSICS, 2002, 29 (05) :830-834
[2]   ANALYSIS OF CANCERS MISSED AT SCREENING MAMMOGRAPHY [J].
BIRD, RE ;
WALLACE, TW ;
YANKASKAS, BC .
RADIOLOGY, 1992, 184 (03) :613-617
[3]   A MULTIPARAMETER OPTIMIZATION OF DIGITAL MAMMOGRAPHY [J].
COURT, LE ;
SPELLER, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) :1841-1861
[4]   Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study [J].
Dance, DR ;
Klang, AT ;
Sandborg, M ;
Skinner, CL ;
Smith, IAC ;
Carlsson, GA .
BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (874) :1056-1067
[5]  
ELWOOD JM, 1993, J CURR CLIN TRIALS, V1, P32
[6]   OPTIMIZATION OF SPECTRAL SHAPE IN DIGITAL MAMMOGRAPHY - DEPENDENCE ON ANODE MATERIAL, BREAST THICKNESS, AND LESION TYPE [J].
FAHRIG, R ;
YAFFE, MJ .
MEDICAL PHYSICS, 1994, 21 (09) :1473-1481
[7]   REPORT OF THE INTERNATIONAL WORKSHOP ON SCREENING FOR BREAST-CANCER [J].
FLETCHER, SW ;
BLACK, W ;
HARRIS, R ;
RIMER, BK ;
SHAPIRO, S .
JOURNAL OF THE NATIONAL CANCER INSTITUTE, 1993, 85 (20) :1644-1656
[8]   Optimizing techniques in screen-film mammography [J].
Hendrick, RE ;
Berns, EA .
RADIOLOGIC CLINICS OF NORTH AMERICA, 2000, 38 (04) :701-+
[9]  
HENDRICK RE, 1999, 1999 SYLLABUS CATEGO, P79
[10]   Signal to noise ratio and radiation dose as function of photon energy in mammography. [J].
Huda, W ;
Krol, A ;
Jing, ZX ;
Boone, JM .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :355-363