Implantable cardioverter defibrillator - An unusual case of inappropriate discharge during showering

被引:10
作者
Manolis, AG [1 ]
Katsivas, AG [1 ]
Vassilopoulos, CV [1 ]
Louvros, NE [1 ]
机构
[1] Hellen Red Cross Hosp, Dept Cardiol 2, Athens, Greece
关键词
implantable cardioverter defibrillators; inappropriate shocks; electromagnetic interference;
D O I
10.1023/A:1009834100540
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
The evolution provided in implantable cardioverter defibrillators (ICDs) has paralleled the advances in the therapeutic options incorporated in these systems. Contemporary devices are capable of providing recordings of electrical events surrounding all delivered and aborted device therapy. This report presents un unusual case of inappropriate discharge of an ICD, resulting from electromagnetic interference. A transvenous ICD system (Sentry Hot Can 4310 HC, Telectronics Denver, CO), was implanted in a patient with ischemic heart disease due to episodes of ventricular tachycardia refractory to antiarrhythmic treatment. One month post-implant the patient reported two consecutive shocks from the device while showering. The nonphysiological cycle length (100 ms) recorded in conjunction to the scenario of the event, raised the suspicion of electromagnetic interference through electrical current leakage in the bathroom, an hypothesis that was subsequently proved. This case report underscores that electromagnetic interference can become hazardous in common daily activities of patients with an ICD.
引用
收藏
页码:265 / 268
页数:4
相关论文
共 14 条
[1]
PACED BEATS FOLLOWING SINGLE NONSENSED COMPLEXES IN A CODEPENDENT CARDIOVERTER DEFIBRILLATOR AND BRADYCARDIA PACING SYSTEM - POTENTIAL FOR VENTRICULAR-TACHYCARDIA INDUCTION [J].
CALLANS, DJ ;
HOOK, BG ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (08) :1281-1287
[2]
UNIQUE SENSING ERRORS IN 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
CALLANS, DJ ;
HOOK, BG ;
KLEIMAN, RB ;
MITRA, RL ;
FLORES, BT ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (04) :1135-1140
[3]
ECHT DS, 1984, PACE, V7, P1345
[4]
HOOK BG, 1991, J AM COLL CARDIOL, V17, P985
[5]
IMPORTANCE OF ABORTIVE SHOCK CAPABILITY WITH ELECTROGRAM STORAGE IN CARDIOVERTER-DEFIBRILLATOR DEVICES [J].
HURWITZ, JL ;
HOOK, BG ;
FLORES, BT ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (04) :895-900
[6]
Dysfunction of implantable defibrillators caused by slot machines [J].
Madrid, A ;
Sanchez, A ;
Bosch, E ;
Fernandez, E ;
Serrano, CM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01) :212-214
[7]
ICD DATA-STORAGE - VALUE IN ARRHYTHMIA MANAGEMENT [J].
MARCHLINSKI, FE ;
GOTTLIEB, CD ;
SARTER, B ;
FINKLE, J ;
HOOK, B ;
CALLANS, D ;
SCHWARTZMAN, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :527-534
[8]
MARCHLINSKI FE, 1993, TACHYCARDIAS MECHANI, P405
[9]
MARCHLINSKI FE, 1994, CARDIAC ELECTROPHYSI
[10]
Inappropriate shocks from implanted cardioverter defibrillators caused by sensing of diaphragmatic myopotentials [J].
Peters, RW ;
Cooklin, M ;
Brockman, R ;
Shorofsky, SR ;
Gold, MR .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 1998, 2 (04) :367-370