Monte Carlo simulation and patient dosimetry for a kilovoltage cone-beam CT unit

被引:57
作者
Downes, P. [1 ,3 ]
Jarvis, R. [1 ]
Radu, E. [1 ]
Kawrakow, I. [4 ]
Spezi, E. [1 ,2 ]
机构
[1] Velindre Canc Ctr, Dept Med Phys, Cardiff CF14 2TL, S Glam, Wales
[2] Cardiff Univ, Sch Engn, Inst Med Engn & Med Phys, Cardiff CF24 3AA, S Glam, Wales
[3] Cardiff Univ, Cardiff Sch Comp Sci, Cardiff CF24 3AA, S Glam, Wales
[4] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
基金
英国工程与自然科学研究理事会;
关键词
dosimetry; radiation therapy; computerised tomography; phantoms; Monte Carlo methods; patient treatment; X-RAY TUBE; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; CONCOMITANT EXPOSURES; RADIOTHERAPY; IMRT; CODE; VERIFICATION; EFFICIENCY; VOLUME;
D O I
10.1118/1.3196182
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this work is to characterize the x-ray volume imager (XVI), the cone-beam computed tomography (CBCT) unit mounted on the Elekta Synergy linac, with F1 bowtie filter and to calculate the three-dimensional dose delivered to patients using volumetric acquisition. Methods: The XVI is modeled in detail using a new Monte Carlo (MC) code, BEAMPP, under development at the National Research Council Canada. In this investigation, a new component module is developed to accurately model the unit's bowtie filter used in conjunction with the available beam collimators at the clinical energy of 120 kV. The modeling is compared against percentage depth dose (PDD) and profile measurements. Kilovoltage radiation beams' phase space files are also analyzed. The authors also describe a method for the absolute dose calibration of the MC model of the CBCT unit when used in a clinical volumetric acquisition mode. Finally, they calculate three-dimensional patient dose from CBCT image acquisition in three clinical cases of interest: Pelvis, lung, and head and neck. Results: The agreement between measurement and MC is shown to be very good: Within +/- 2% for the PDD and within +/- 3.5% inside the radiation field for all the collimators with the F1 bowtie filter. A full account of the absolute calibration method is given and dose calculation is validated against ion chamber measurements in different locations of a plastic phantom. Calculations and experiments agree within +/- 2% or better in both at the center and the periphery of the phantom, with worst agreement of 4.5% at the surface of the phantom and for one specific combination of collimator and filter. Patient dose from CBCT scan reveals that dose to tissue is between 2 and 2.5 cGy for a pelvis or a lung full acquisition. For H&N dose to tissue is 5 cGy, with the unit presets used in this work. Dose to bony structures can be two to three times higher than dose to tissue. Conclusions: The XVI CBCT unit has been fully modeled including the F1 bowtie filter. Absolute dose distribution from the unit has been successfully validated. Full MC patient dose calculation has shown that the three-dimensional dose distribution from CBCT is complex. Patient dose from CBCT exposure cannot be completely accounted for by using a numerical factor as an estimate of the dose at the center of the body. Furthermore, additional dose to bone should be taken into account when adopting any IGRT strategy and weighed vs the unquestionable benefits of the technique in order to optimize treatment. Full three-dimensional dose calculation is recommended if patient dose from CBCT is to be integrated in any adaptive planning strategy. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3196182]
引用
收藏
页码:4156 / 4167
页数:12
相关论文
共 34 条
[1]   Second cancer risk, concomitant exposures and IRMER(2000) [J].
Aird, EGA .
BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (924) :983-985
[2]   Benchmarking EGSnrc in the kilovoltage energy range against experimental measurements of charged particle backscatter coefficients [J].
Ali, E. S. M. ;
Rogers, D. W. O. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (06) :1527-1543
[3]   Development and validation of MCNP4C-based Monte Carlo simulator for fan- and cone-beam x-ray CT [J].
Ay, MR ;
Zaidi, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (20) :4863-4885
[4]   Monte Carlo simulation of a computed tomography x-ray tube [J].
Bazalova, Magdalena ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (19) :5945-5955
[5]   Induced second cancers after prostate-cancer radiotherapy: No cause for concern? [J].
Brenner, DJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (03) :637-639
[6]   Evaluation of the effect of patient dose from cone beam computed tomography on prostate IMRT using Monte Carlo simulation [J].
Chow, James C. L. ;
Leung, Michael K. K. ;
Islam, Mohammad K. ;
Norrlinger, Bernhard D. ;
Jaffray, David A. .
MEDICAL PHYSICS, 2008, 35 (01) :52-60
[7]   CERR: A computational environment for radiotherapy research [J].
Deasy, JO ;
Blanco, AI ;
Clark, VH .
MEDICAL PHYSICS, 2003, 30 (05) :979-985
[8]  
di Sopra FM, 1999, MED PHYS, V26, P2000, DOI 10.1118/1.598705
[9]   Accurate patient dosimetry of kilovoltage cone-beam CT in radiation therapy [J].
Ding, George X. ;
Duggan, Dennis M. ;
Coffey, Charles W. .
MEDICAL PHYSICS, 2008, 35 (03) :1135-1144
[10]   A study on adaptive IMRT treatment planning using kV cone-beam CT [J].
Ding, George X. ;
Duggan, Dennis M. ;
Coffey, Charles W. ;
Deeley, Matthew ;
Hallahan, Dennis E. ;
Cmelak, Anthony ;
Malcolm, Arnold .
RADIOTHERAPY AND ONCOLOGY, 2007, 85 (01) :116-125