Monte Carlo simulation of a computed tomography x-ray tube

被引:55
作者
Bazalova, Magdalena [1 ]
Verhaegen, Frank [1 ]
机构
[1] McGill Univ, Montreal Gen Hosp, Dept Med Phys, Montreal, PQ H3G 1A4, Canada
关键词
D O I
10.1088/0031-9155/52/19/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The dose delivered to patients during computed tomography ( CT) exams has increased in the past decade. With the increasing complexity of CT examinations, measurement of the dose becomes more difficult and more important. In some cases, the standard methods, such as measurement of the computed tomography dose index ( CTDI), are currently under question. One approach to determine the dose from CT exams is to use Monte Carlo ( MC) methods. Since the patient geometry can be included in the model, Monte Carlo simulations are potentially the most accurate method of determining the dose delivered to patients. In this work, we developed a MC model of a CT x-ray tube. The model was validated with half-value layer ( HVL) measurements and spectral measurements with a high resolution Schottky CdTe spectrometer. First and second HVL for beams without additional filtration calculated from the MC modelled spectra and determined from attenuation measurements differ by less than 2.5%. The differences between the first and second HVL for both filtered and non-filtered beams calculated from the MC modelled spectra and spectral measurements with the CdTe detector were less than 1.8%. The MC modelled spectra match the directly measured spectra. This works presents a first step towards an accurate MC model of a CT scanner.
引用
收藏
页码:5945 / 5955
页数:11
相关论文
共 22 条
[1]   Diagnostic x-ray spectra: A comparison of spectra generated by different computational methods with a measured spectrum [J].
Bhat, M ;
Pattison, J ;
Bibbo, G ;
Caon, M .
MEDICAL PHYSICS, 1998, 25 (01) :114-120
[2]   Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV [J].
Boone, JM ;
Seibert, JA .
MEDICAL PHYSICS, 1997, 24 (11) :1661-1670
[3]   Is it time to retire the CTDI for CT quality assurance and dose optimization? [J].
Brenner, DJ .
MEDICAL PHYSICS, 2005, 32 (10) :3225-3226
[4]   MONTE-CARLO SIMULATION STUDIES OF DETECTORS USED IN THE MEASUREMENT OF DIAGNOSTIC-X-RAY SPECTRA [J].
CHEN, CS ;
DOI, K ;
VYBORNY, C ;
CHAN, HP ;
HOLJE, G .
MEDICAL PHYSICS, 1980, 7 (06) :627-635
[5]   A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms [J].
DeMarco, JJ ;
Cagnon, CH ;
Cody, DD ;
Stevens, DM ;
McCollough, CH ;
O'Daniel, J ;
McNitt-Gray, MF .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (17) :3989-4004
[6]  
Fewell T., 1981, HDB COMPUTED TOMOGRA
[7]   AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology [J].
Ma, CM ;
Coffey, CW ;
DeWerd, LA ;
Liu, C ;
Nath, R ;
Seltzer, SM ;
Seuntjens, JP .
MEDICAL PHYSICS, 2001, 28 (06) :868-893
[8]   Compton-scattering measurement of diagnostic x-ray spectrum using high-resolution Schottky CdTe detector [J].
Maeda, K ;
Matsumoto, M ;
Taniguchi, A .
MEDICAL PHYSICS, 2005, 32 (06) :1542-1547
[9]   COMPTON SPECTROSCOPY IN THE DIAGNOSTIC-X-RAY ENERGY-RANGE .1. SPECTROMETER DESIGN [J].
MATSCHEKO, G ;
CARLSSON, GA .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (02) :185-197
[10]   COMPTON SPECTROSCOPY IN THE DIAGNOSTIC-X-RAY ENERGY-RANGE .2. EFFECTS OF SCATTERING MATERIAL AND ENERGY RESOLUTION [J].
MATSCHEKO, G ;
CARLSSON, GA ;
RIBBERFORS, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (02) :199-208