Automated pulse pressure and stroke volume variations from radial artery: evaluation during major abdominal surgery

被引:94
作者
Derichard, A. [1 ]
Robin, E. [1 ]
Tavernier, B. [1 ]
Costecalde, M. [1 ]
Fleyfel, M. [1 ]
Onimus, J. [1 ]
Lebuffe, G. [1 ]
Chambon, J. -P. [1 ]
Vallet, B. [1 ]
机构
[1] CHU Lille, Hop Roger Salengro, Dept Vasc & Gen Surg, F-59037 Lille, France
关键词
fluids; i.v; heart; cardiac output; monitoring; arterial pressure; ventilation; mechanical; FLUID RESPONSIVENESS; CARDIAC-OUTPUT; PREDICTION; ALGORITHM; FLOTRAC(TM); MANAGEMENT; MONITOR;
D O I
10.1093/bja/aep267
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. Off-line calculation of the pulse pressure variation (PPVref) has repeatedly been shown to be a reliable predictor of fluid responsiveness in mechanically ventilated patients. This study was designed to assess the ability of two algorithms for automated calculation of PPV (PPVauto) (Intellivue MP 70) and stroke volume variation (SVVauto) (FloTrac/Vigileo) to predict fluid responsiveness during abdominal surgery. Methods. We conducted a prospective study of 56 fluid challenges given for haemodynamic instability in 11 patients undergoing major abdominal surgery. Fluid responsiveness was defined as an increase in stroke volume index (SVI) >10%. PPVref, PPVauto, SVVauto, and SVI (oesophageal Doppler) were recorded simultaneously before and after each fluid challenge. Results. PPVauto and SVVauto both correlated with PPVref [r(corr)=0.87 (P<0.0001) and 0.84 (P<0.0001), respectively; n=77]. All three indices measured before fluid challenges were higher in responder (n=32) than in non-responder (n=24) fluid challenges (P <= 0.02). The mean areas under the receiver operating characteristic curves were 0.96 (PPVref), 0.96 (PPVauto), and 0.95 (SVVauto), and the optimal threshold value for each variable was 13%, 13%, and 12%, respectively. All indices correlated with the fluid challenge-induced changes in SVI (PPVref: r(corr)=0.65; PPVauto: r(corr)=0.58; SVVauto: r(corr)=0.58, P<0.001 for all). Conclusions. PPVauto and SVVauto to predict fluid responsiveness as accurately as off-line PPVref in patients with haemodynamic instability during major abdominal surgery.
引用
收藏
页码:678 / 684
页数:7
相关论文
共 23 条
  • [1] A novel algorithm to estimate the pulse pressure variation index ΔPP
    Aboy, M
    McNames, J
    Thong, T
    Phillips, CR
    Ellenby, MS
    Goldstein, B
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (12) : 2198 - 2203
  • [2] Hemodynamic monitoring in shock and implications for management International Consensus Conference, Paris, France, 27-28 April 2006
    Antonelli, Massimo
    Levy, Mitchell
    Andrews, Peter J. D.
    Chastre, Jean
    Hudson, Leonard D.
    Manthous, Constantine
    Meduri, G. Umberto
    Moreno, Rui P.
    Putensen, Christian
    Stewart, Thomas
    Torres, Antoni
    [J]. INTENSIVE CARE MEDICINE, 2007, 33 (04) : 575 - 590
  • [3] Online monitoring of pulse pressure variation to guide fluid therapy after cardiac surgery
    Auler, Jose Otavio, Jr.
    Galas, Filomena
    Hajjar, Ludhmila
    Santos, Luciana
    Carvalho, Thiago
    Michard, Frederic
    [J]. ANESTHESIA AND ANALGESIA, 2008, 106 (04) : 1201 - 1206
  • [4] Monitoring of peri-operative fluid administration by individualized goal-directed therapy
    Bundgaard-Nielsen, M.
    Holte, K.
    Secher, N. H.
    Kehlet, H.
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2007, 51 (03) : 331 - 340
  • [5] Clinical evaluation of the FloTrac/Nigileo™ system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery
    Button, D.
    Weibel, L.
    Reuthebuch, O.
    Genoni, M.
    Zollinger, A.
    Hofer, C. K.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2007, 99 (03) : 329 - 336
  • [6] The ability of a novel algorithm for automatic estimation of the respiratory variations, in arterial pulse pressure to monitor fluid responsiveness in the operating room
    Cannesson, Maxime
    Slieker, Juliette
    Desebbe, Olivier
    Bauer, Christian
    Chiari, Pascal
    Henaine, Roland
    Lehot, Jean-Jacques
    [J]. ANESTHESIA AND ANALGESIA, 2008, 106 (04) : 1195 - 1200
  • [7] Cavallaro F, 2008, MINERVA ANESTESIOL, V74, P123
  • [8] The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults
    Dark, PM
    Singer, M
    [J]. INTENSIVE CARE MEDICINE, 2004, 30 (11) : 2060 - 2066
  • [9] Stroke volume variation obtained with FloTrac/Vigileo™ fails to predict fluid responsiveness in coronary artery bypass graft patients
    de Waal, E. E. C.
    Rex, S.
    Kruitwagen, C. L. J. J.
    Kalkman, C. J.
    Buhre, W. F.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (05) : 725 - 726
  • [10] GAO F, P 31 ANN SAS US GROU