Diffusion-weighted MR for Differentiation of Breast Lesions at 3.0 T: How Does Selection of Diffusion Protocols Affect Diagnosis?

被引:239
作者
Bogner, Wolfgang [1 ]
Gruber, Stephan [1 ]
Pinker, Katja [1 ]
Grabner, Guenther [1 ]
Stadlbauer, Andreas [1 ]
Weber, Michael [1 ]
Moser, Ewald [1 ]
Helbich, Thomas H. [1 ]
Trattnig, Siegfried [1 ]
机构
[1] Med Univ Vienna, Vienna Gen Hosp, Dept Radiol, A-1090 Vienna, Austria
关键词
SIGNAL INTENSITY; SPIN-ECHO; HIGH-RISK; CANCER; IMAGES; TUMORS; COEFFICIENTS; REDUCTION; PERFUSION; VALUES;
D O I
10.1148/radiol.2532081718
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the diagnostic quality of diffusion-weighted (DW) imaging schemes with regard to apparent diffusion coefficient (ADC) accuracy, ADC precision, and DW imaging contrast-to-noise ratio (CNR) for different types of lesions and breast tissue. Materials and Methods: Institutional review board approval and written, informed consent were obtained. Fifty-one patients with histopathologic correlation or follow-up performed with a 3.0-T MR imager were included in this study. There were 112 regions of interest drawn in 24 malignant, 17 benign, 20 cystic, and 51 normal tissue regions. ADC maps were calculated for combinations of 10 b values (range, 0-1250 sec/mm(2)). Differences in ADC among tissue types were evaluated. The CNRs of lesions at DW imaging were compared for all b values. A repeated-measures analysis of variance was used to assess lesion differentiation. Results: ADCs calculated from b values of 50 and 850 sec/mm(2) were 0.99 X 10(-3) mm(2)/sec +/- 0.18 (standard deviation), 1.47 X 10(-3) mm(2)/sec +/- 0.21, 1.85 X 10(-3) mm(2)/sec +/- 0.22, and 2.64 X 10(-3) mm(2)/sec +/- 0.30 for malignant, benign, normal, and cystic tissues, respectively. An ADC threshold level of 1.25 X 10(-3) mm(2)/sec allowed discrimination between malignant and benign lesions with a diagnostic accuracy of 95% (P < .001). ADC calculations performed with multiple b values were not significantly more precise than those performed with only two. We found an overestimation of ADC for maximum b values of up to 1000 sec/mm(2). The best CNR for tumors was identified at 850 sec/mm(2). Conclusion: Optimum ADC determination and DW imaging quality at 3.0 T was found with a combined b value protocol of 50 and 850 sec/mm(2). This provided a high accuracy for differentiation of benign and malignant breast tumors. (C) RSNA, 2009
引用
收藏
页码:341 / 351
页数:11
相关论文
共 40 条
[21]   Quantitative diffusion-weighted MR imaging in the differential diagnosis of breast lesion [J].
Marini, C. ;
Iacconi, C. ;
Giannelli, M. ;
Cilotti, A. ;
Moretti, M. ;
Bartolozzi, C. .
EUROPEAN RADIOLOGY, 2007, 17 (10) :2646-2655
[22]   Comparison of 3.0-and 1.5-tesla diffusion-weighted imaging in the visibility of breast cancer [J].
Matsuoka, Aoi ;
Minato, Masako ;
Harada, Masafumi ;
Kubo, Hitoshi ;
Bandou, Yoshikatsu ;
Tangoku, Akira ;
Nakano, Kiichirou ;
Nishitani, Hiromu .
RADIATION MEDICINE, 2008, 26 (01) :15-20
[23]   The role of diffusion-weighted imaging and the apparent diffusion coefficient (ADC) values for breast tumors [J].
Park, Mi Jung ;
Cha, Eun Suk ;
Kang, Bong Joo ;
Ihn, Yon Kwon ;
Baik, Jun Hyun .
KOREAN JOURNAL OF RADIOLOGY, 2007, 8 (05) :390-396
[24]   Menstrual cycle variation of apparent diffusion coefficients measured in the normal breast using MRI [J].
Partridge, SC ;
McKinnon, GC ;
Henry, RG ;
Hylton, NM .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (04) :433-438
[25]   Diffusion changes precede size reduction in neoadjuvant treatment of breast cancer [J].
Pickles, Martin D. ;
Gibbs, Peter ;
Lowry, Martin ;
Turnbull, Lindsay W. .
MAGNETIC RESONANCE IMAGING, 2006, 24 (07) :843-847
[26]   Reduction of eddy-current-induced distortion in diffusion MRI using a twice-refocused spin echo [J].
Reese, TG ;
Heid, O ;
Weisskoff, RM ;
Wedeen, VJ .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) :177-182
[27]   Magnetic resonance imaging of the breast improves detection of invasive cancer, preinvasive cancer, and premalignant lesions during surveillance of women at high risk for breast cancer [J].
Riedl, Christopher C. ;
Ponhold, Lothar ;
Floery, Daniel ;
Weber, Michael ;
Kroiss, Regina ;
Wagner, Teresa ;
Fuchsjaeger, Michael ;
Helbich, Thomas H. .
CLINICAL CANCER RESEARCH, 2007, 13 (20) :6144-6152
[28]   Quantitative diffusion imaging in breast cancer: A clinical prospective study [J].
Rubesova, Erika ;
Grell, Anne-Sophie ;
De Maertelaer, Viviane ;
Metens, Thierry ;
Chao, Shih-Li ;
Lemort, Marc .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (02) :319-324
[29]   Longitudinal study of the assessment by MRI and diffusion-weighted imaging of tumor response in patients with locally advanced breast cancer undergoing neoadjuvant chemotherapy [J].
Sharma, Uma ;
Danishad, Karikanni Kalathil A. ;
Seenu, Vurthaluru ;
Jagannathan, Naranamangalam R. .
NMR IN BIOMEDICINE, 2009, 22 (01) :104-113
[30]   In vivo diffusion-weighted MRI of the breast: Potential for lesion characterization [J].
Sinha, S ;
Lucas-Quesada, FA ;
Sinha, U ;
DeBruhl, N ;
Bassett, LW .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 15 (06) :693-704