The performance of a CCD digital autoradiography imaging system

被引:45
作者
Ott, R [1 ]
MacDonald, J [1 ]
Wells, K [1 ]
机构
[1] Royal Marsden Hosp, Inst Canc Res, Dept Phys, Sutton, Surrey, England
关键词
D O I
10.1088/0031-9155/45/7/322
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Autoradiography is a widely used technique for imaging trace quantities of radioactivity within biological samples, conventionally using photographic film. This method produces images with high spatial resolution, but it suffers from very low sensitivity and poor dynamic range. Digital autoradiography systems with greatly improved sensitivity and linearity are commercially available, but the spatial resolution is usually much less than that achieved using him. We report here the design, construction and characterization of a novel digital autoradiography system based on scientific-grade charged coupled devices (CCDs). Images of x-ray and beta emissions from radionuclides commonly used in autoradiography show that the system can perform highspeed quantitative imaging with a spatial resolution of similar to 30 mu m. Using a frame by frame acquisition method the dynamic range is shown to be at least three orders of magnitude. The absolute detection efficiency is comparable to the best of the currently available digital systems. CCD images of(125)I and C-14 radioisotope distributions in tissue samples are superior to the equivalent film images and have been acquired in 1-10% of the time.
引用
收藏
页码:2011 / 2027
页数:17
相关论文
共 16 条
[1]  
BERTOLUCCI E, 1997, IEEE T NUCL SCI, V44, P929
[2]   A HIGH-RESOLUTION BETA- DETECTOR [J].
CHARON, Y ;
CUZON, JC ;
TRICOIRE, H ;
VALENTIN, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 273 (2-3) :748-753
[3]   HRRI - A HIGH-RESOLUTION BETA- IMAGER FOR BIOLOGICAL APPLICATIONS [J].
CHARON, Y ;
BENDALI, M ;
CUZON, JC ;
LEBLANC, M ;
MASTRIPPOLITO, R ;
TRICOIRE, H ;
VALENTIN, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 292 (01) :179-186
[4]  
ENGLERT D, 1995, CELL MOL BIOL, V41, P57
[5]   LARGE AREA REAL-TIME SYSTEM FOR SCREENING DNA LIBRARIES [J].
GORDON, JS ;
MCGANN, W ;
DALEY, K ;
MCCONCHIE, L ;
SQUILLANTE, MR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (05) :1485-1489
[6]   INFLUENCE OF ELECTRON TRAPS ON CHARGE-COLLECTION EFFICIENCY IN GAAS RADIATION DETECTORS [J].
NAVA, F ;
CANALI, C ;
CASTALDINI, A ;
CAVALLINI, A ;
DAURIA, S ;
DELPAPA, C ;
FRIGERI, C ;
ZANOTTI, L ;
CETRONIO, A ;
LANZIERI, C ;
ZICHICHI, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 349 (01) :156-159
[7]   ANALYTIC MODELING OF CHARGE DIFFUSION IN CHARGE-COUPLED-DEVICE IMAGERS [J].
HOPKINSON, GR .
OPTICAL ENGINEERING, 1987, 26 (08) :766-772
[8]   AUTORADIOGRAPHY USING STORAGE PHOSPHOR TECHNOLOGY [J].
JOHNSTON, RF ;
PICKETT, SC ;
BARKER, DL .
ELECTROPHORESIS, 1990, 11 (05) :355-360
[9]   IMAGING OF RADIONUCLIDE EMISSIONS WITH A LOW-NOISE CHARGE-COUPLED-DEVICE [J].
KARELLAS, A ;
HONG, L ;
REINHARDT, C ;
HARRIS, LJ ;
BRILL, AB .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) :979-982
[10]   CHARGE-COUPLED DEVICE DETECTOR - PERFORMANCE CONSIDERATIONS AND POTENTIAL FOR SMALL-FIELD MAMMOGRAPHIC IMAGING APPLICATIONS [J].
KARELLAS, A ;
HARRIS, LJ ;
LIU, H ;
DAVIS, MA ;
DORSI, CJ .
MEDICAL PHYSICS, 1992, 19 (04) :1015-1023