Pulmonary nodule detection with digital projection radiography: an ex-vivo study on increased latitude post-processing

被引:2
作者
Biederer, Juergen
Gottwald, Tobias
Bolte, Hendrik
Riedel, Christian
Freitag, Sandra
Van Metter, Richard
Heller, Martin
机构
[1] Univ Hosp Schleswig Holstein, Dept Diagnost Radiol, Kiel, Germany
[2] Univ Hosp Schleswig Holstein, Inst Med Informat & Stat, Kiel, Germany
[3] Eastman Kodak Co, HI World Wide Res & Dev Rochester Med Phys, Washington, DC USA
关键词
pulmonary nodule; digital radiography; phantom; image latitude;
D O I
10.1007/s00330-006-0367-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To evaluate increased image latitude post-processing of digital projection radiograms for the detection of pulmonary nodules. 20 porcine lungs were inflated inside a chest phantom, prepared with 280 solid nodules of 4-8 mm in diameter and examined with direct radiography (3.0x2.5 k detector, 125 kVp, 4 mAs). Nodule position and size were documented by CT controls and dissection. Four intact lungs served as negative controls. Image post-processing included standard tone scales and increased latitude with detail contrast enhancement (log-factors 1.0, 1.5 and 2.0). 1280 sub-images (512x512 pixel) were centred on nodules or controls, behind the diaphragm and over free parenchyma, randomized and presented to six readers. Confidence in the decision was recorded with a scale of 0-100%. Sensitivity and specificity for nodules behind the diaphragm were 0.87/0.97 at standard tone scale and 0.92/0.92 with increased latitude (log factor 2.0). The fraction of "not diagnostic" readings was reduced (from 208/1920 to 52/1920). As an indicator of increased detection confidence, the median of the ratings behind the diaphragm approached 100 and 0, respectively, and the inter-quartile width decreased (controls: p < 0.001, nodules: p=0.239) at higher image latitude. Above the diaphragm, accuracy and detection confidence remained unchanged. Here, the sensitivity for nodules was 0.94 with a specificity from 0.96 to 0.97 (all p > 0.05). Increased latitude post-processing has minimal effects on the overall accuracy, but improves the detection confidence for sub-centimeter nodules in the posterior recesses of the lung.
引用
收藏
页码:1089 / 1100
页数:12
相关论文
共 35 条
[1]   New automatic tone scale method for computed radiography [J].
Barski, LL ;
Van Metter, RL ;
Foos, DH ;
Lee, HC ;
Wang, XH .
IMAGE DISPLAY - MEDICAL IMAGING 1998, 1998, 3335 :164-178
[2]   Diagnostic performance of a flat-panel detector at low tube voltage in chest radiography -: A phantom study [J].
Bernhardt, TM ;
Rapp-Bernhardt, U ;
Lenzen, H ;
Röhl, FW ;
Diederich, S ;
Papke, K ;
Ludwig, K ;
Heindel, W .
INVESTIGATIVE RADIOLOGY, 2004, 39 (02) :97-103
[3]   Simulated pulmonary nodules implanted in a dedicated porcine chest phantom: Sensitivity of MR imaging for detection [J].
Biederer, J ;
Schoene, A ;
Freitag, S ;
Reuter, M ;
Heller, M .
RADIOLOGY, 2003, 227 (02) :475-483
[4]   Artificial thorax for MR imaging studies in porcine heart-lung preparations [J].
Biederer, J ;
Heller, M .
RADIOLOGY, 2003, 226 (01) :250-255
[5]   The ACR/NEMA proposal for a grey-scale display function standard [J].
Blume, H .
IMAGE DISPLAY: MEDICAL IMAGING 1996, 1996, 2707 :344-360
[6]   Ex-vivo injection for implanting solid pulmonary nodules into porcine lungs for multi-slice CT [J].
Bolte, H ;
Müller-Hülsbeck, S ;
Riedel, C ;
Jahnke, T ;
Inan, N ;
Heller, M ;
Biederer, J .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (10) :1380-1384
[7]   FILM-BASED CHEST RADIOGRAPHY - AMBER VS ASYMMETRIC SCREEN-FILM SYSTEMS [J].
CHOTAS, HG ;
FLOYD, CE ;
RAVIN, CE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 161 (04) :743-747
[8]   Effect of monitor luminance and ambient light on observer performance in soft-copy reading of digital chest radiographs [J].
Goo, JM ;
Choi, AY ;
Im, JG ;
Lee, HL ;
Chung, MJ ;
Han, DH ;
Park, SH ;
Kim, JH ;
Nam, SH .
RADIOLOGY, 2004, 232 (03) :762-766
[9]   Detection of simulated chest lesions by using soft-copy reading: Comparison of an amorphous silicon flat-panel-detector system and a storage-phosphor system [J].
Goo, JM ;
Im, JG ;
Lee, HJ ;
Chung, MJ ;
Seo, JB ;
Kim, HY ;
Lee, YJ ;
Kang, JW ;
Kim, JH .
RADIOLOGY, 2002, 224 (01) :242-246
[10]   IK versus 2K monitor: A clinical alternative free-response receiver operating characteristic study of observer performance using pulmonary nodules [J].
Graf, B ;
Simon, U ;
Eickmeyer, F ;
Fiedler, V .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (04) :1067-1074